page 1 of 12‡ SIGNET 8750-1 pH/ORP Transmitter
pH/ORP
10.20 pH
25.0°C
ENTER
‡ SIGNET 8750-1 pH/ORP Transmitter ENGLISH
2. Speci ications
General
Compatible electrodes
+GF+ SIGNET 3-27XX/2720 pH/ORP Preamplifier/Electrode
Accuracy ± 0.03 pH, ± 2 mV ORP
Enclosure
• Rating NEMA 4X/IP65 front
• Case PBT
• Panel case gasket Neoprene
• Window Polyurethane coated polycarbonate
• Keypad Sealed 4-key silicone rubber
• Weight Approx. 325g (12 oz.)
Display
• Alphanumeric 2 x 16 LCD
• Contrast User selected, 5 levels
• Update rate 1 second
Electrical
• Power 12 to 24 VDC ±5%, regulated, 21 mA max.
Sensor input range
• pH 0.00 to 14.00 pH
• temp. 3K Balco, -25 to 120°C (-13 to 248°F)
• ORP -1000 to +2000 mV, isolated
(10KΩ I.D. resistance T+, T-)
Current output
• 4 to 20 mA, isolated, fully adjustable and reversible
3-8750.090-1
CAUTION!
• Remove power to unit before wiring input and
output connections.
• Follow instructions carefully to avoid personal injury.
1. Installation
ProcessPro transmitters are available in two styles panel mount and field mount. The panel mount is supplied with the necessary
hardware to install the transmitter. This manual includes complete panel mounting instructions.
Field mounting requires a separate mounting kit. The 3-8050 Universal kit enables the transmitter to be installed virtually
anywhere. Detailed instructions for field installation options are included with the 3-8050 Universal kit.
1.1 Panel Installation
1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual panel cutout, an adhesive
template is provided as an installation guide. Recommended clearance on all sides between instruments is 1 inch.
2. Place gasket on instrument, and install in panel.
3. Slide mounting bracket over back of instrument until quick-clips snap into latches on side of instrument.
4. To remove, secure instrument temporarily with tape from front or grip from rear of instrument. DO NOT RELEASE.
Press quick-clips outward and remove.
Contents
1. Installation
2. Specifications
3. Electrical Connections
4. Menu Functions
SIDE VIEW
92 mm
(3.6 in.)
97 mm
(3.8 in.)
56 mm
(2.2 in.)
41 mm
(1.6 in.)
Optional
Rear
Cover
Field Mount Panel Mount
FRONT VIEW
96 mm
(3.8 in.)
96 mm
(3.8 in.)
quick-clips
gasket panel
terminals mounting
bracket
latch
Output -
Output +
System Pwr
Loop -
System Pwr
Loop +
2
14
3
Sensr Gnd
(SHIELD)
Sensr IN
(RED)
Sensr V+
(BLACK)
7
6
5
SIDE VIEW
Field Mount &
Panel Mount Panel Mount
Installation Detail
106 mm
(4.18 in.)
82 mm
(3.23 in.)
• Max loop impedance 50 Ω max. @ 12 V
325 Ω max. @ 18 V
600 Ω max. @ 24 V
• Update rate 0.5 seconds
• Accuracy ±0.03 mA @ 25°C, 24 V
Open-collector output
• Isolated, 50 mA sink or source, 30 VDC max. pull-up voltage
• Programmable for
• Hi or Lo w/adjustable hysteresis
• Proportional pulse (400 pulses per minute maximum)
Environmental
• Operating temperature -10 to 70°C (14 to 158°F)
• Storage temperature -15 to 80°C (5 to 176°F)
• Relative humidity 0 to 95%, non-condensing
• Maximum altitude 2000 m (6562 ft)
• Insulation category II
• Pollution degree 2
Standards and Approvals
• CSA, CE, UL listed
• Immunity EN50082-2
• Emissions EN55011
• Safety EN61010
• Manufactured under ISO 9001 and ISO 14001
C-3/01 English