Realtime Technologies Ltd ECG Respiration User Guide
All rights reserved Rev 1.1
4
2. General Information
2.1. Safety Information
As a precaution it is important to note that the ECG leads are not to be applied to the subject's body
while the unit is in a USB dock or multi-charger.
2.2. Pre-Requisites
A compatible Shimmer3 ECG unit programmed with up-to-date firmware.
-The Shimmer's firmware can be updated using our Consensys application - available
for download from www.shimmersensing.com.
-In order to check if respiration detection is supported on your Shimmer3 ECG unit,
please use Consensys and, with your Shimmer unit docked, refer to the "Expansion"
board information shown on-screen. If the expansion board appears as
"ECG/EMG/Respiration", respiration mode is supported.
Five DIN snap leads.
-9-inch and 18-inch leads are shipped with ECG Development Kits.
-Replacements can be purchased from www.shimmersensing.com.
Surface ECG electrodes.
-Disposable electrodes are shipped with the ECG Development Kit.
-For replacements, see www.shimmersensing.com. Alternatively, the Covidien
Kendall Disposable Surface EMG/ECG/EKG electrodes 1" (24mm) or Covidien Kendall
Disposable Surface EMG/ECG/EKG electrodes 1 3/8" (35mm), available on www.bio-
medical.com with product codes 'BRD H124SG' and 'BRD H135SG', respectively, and
the Ambu Blue Sensor T electrodes, available from various suppliers, are all suitable
options and have been validated for use with Shimmer equipment.
2.3. ECG Unit Specification Overview
The Shimmer3 ECG unit contains two ADS1292R chips from Texas Instruments. Throughout
documentation and software applications these are referred to a 'Chip1' and 'Chip2'. The respiration
circuitry is connected to Chip2 [2].
For detailed specifications of the Shimmer3 ECG unit or for more detail on the standard electrode
placement for ECG measurement, please refer to the Shimmer3 ECG User Guide.