Koehler K18854 Guía

Operation and Instruction Manual
REV B
Koehler Instrument Company, Inc.
1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA
Toll Free: 1-800-878-9070 (US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815
http://www.koehlerinstrument.com • e-mail: info@koehlerinstrument.com
Petroleum Testing & Analysis Instrumentation • Custom Design & Manufacturing
K18854, K18864, K18855, K18865
Low Temperature Torque Apparatus

Low Temperature Torque Apparatus
Operation and Instruction Manual
K18854,55,64,65-Manual -1-

Low Temperature Torque Apparatus
Operation and Instruction Manual
K18854,55,64,65-Manual -2-
CERTIFICATE OF CONFORMANCE
Low Temperature Torque Apparatus
K188XX
This certificate verifies that part number K188XX, Low Temperature Torque
Apparatus, was manufactured in conformance with the applicable standards set
forth in this certification.
Specifications: ASTM D1478
ASTM D4693
ASTM D4950
FTM 791-334
This unit is tested before it leaves the factory, to ensure total functionality and
compliance to the above specifications and ASTM standards. Test and inspection
records are on file for verification.
Jesse Kelly
Application Engineer
Koehler Instrument Company

Low Temperature Torque Apparatus
Operation and Instruction Manual
K18854,55,64,65-Manual -3-

Low Temperature Torque Apparatus
Operation and Instruction Manual
K18854,55,64,65-Manual -4-

Low Temperature Torque Apparatus
Operation and Instruction Manual
K18854,55,64,65-Manual -5-

Low Temperature Torque Apparatus
Operation and Instruction Manual
K18854,55,64,65-Manual -6-
WEEE Directive Compliance Statement
Background
The goal of the WEEE Directive is to encourage design of environment-friendly products that increase reuse,
recycling and other forms of recovery to reduce waste streams and applies to listed Electronic and Electrical
Equipment (EEE) and Koehler's equipment falls broadly into Appendix 1A; Section 9 Monitoring and Control
Equipment: Measuring, weighing or adjusting appliances for household or as laboratory equipment.
Any associated non-embedded equipment such as Lighting (Saybolt Color) and PCs/Printers also fall under WEEE.
If provided with an order these ancillary items must be WEEE compliant. For these and other reasons (printer
cartridges are regionalized) the equipment must be supplied through a third party supplier in Europe.
The WEEE Directive applies to electrical and electronic equipment falling under the categories set out in Annex IA
provided that the equipment concerned is not part of another type of equipment that does not fall within the scope
of this Directive. Annex IB contains a list of products which fall under the categories set out in Annex IA.
http://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=OJ:L:2003:037:0024:0038:en:PDF
We do not qualify for any of the 10 exemption categories.
http://www.dpa-system.dk/en/WEEE/Products/Exemptions
Professional use
For equipment defined for ’professional use’ local authorities have no role to play. Producers and importers are
basically responsible for collection of WEEE recyclables from the professional user and for subsequent
management. A separate statement is given cataloging the items that require separation from the equipment
along with basic information on subsequent processing or recycling prior to disposal of the equipment.
http://www.dpa-system.dk/en/WEEE/Products/Private-or-professional-use
Responsibility for Registration and Annual Reporting:
Koehler will not sell directly to end users in the EU and so has no responsibility to register within each EU state and
to make annual reports. Koehler declares that this responsibility is born by the importer who is the first level of the
distribution chain and is subject to producer responsibility. We will communicate this in writing to our
distributor/importers in the EU stating they are responsible to satisfy WEEE registration and reporting requirements
in the EU states where they conduct sales activities.
It is illegal to market electrical and electronic equipment covered by producer responsibility without being
registered.
http://www.dpa-system.dk/en/WEEE/Producers/Whoissubjecttoproducerresponsibility
Product Design
Koehler's designs allow for complete disassembly to a modular level which usually allows for standard recycling. A
qualified refrigeration system technician must be consulted when disassembling and de-commissioning any
equipment with refrigeration systems.
Koehler's scientific testing equipment is robustly designed to function over a long service life and are typically
repaired many times over the course of years rather than being replaced. We believe that re-use and
refurbishment is the very best form of re-cycling.
All batteries must be readily removable not soldered in place.
Recycling instructions
In the event that replacement becomes necessary, we will include instructions, particularized to each instrument
that informs the customer of their recycling responsibilities and giving them guidance in doing this. All Koehler
equipment has been placed on the market since 13th August 2005 and so Koehler is defined as a "new WEEE
producer". As such we must provide information on refurbishment, treatment, and re-use.

Low Temperature Torque Apparatus
Operation and Instruction Manual
K18854,55,64,65-Manual -7-
Our instrument manual will include this compliance statement and indicate that any collection of materials will be
handled by their authorized distributor. In the event that the distributor is unreachable or is no longer a distributor
for Koehler Instrument, Co., other arrangements may be made including accepting the materials directly.
Recycling is free of charge. Shipping is the responsibility of the end users. Whether shipping to a distributor or to
Koehler directly, safe, properly declared, and labeled packaging and shipping expenses are the sole responsibility
of the end user.
WEEE Marking
Since Koehler products are subject to the WEEE Directive we must display the WEEE symbol shown above in
accordance with European Standard EN 50419 on the equipment. It must be indelible, at least 5mm in height, and
clearly legible. If the equipment is too small the mark must be in the product literature, guarantee certificate, or on
the packaging. Rules on marking are established in section 49 of the WEEE Order.
Koehler Instrument Company, Inc.
c/o RECYCLING
1595 Sycamore, Ave.
Bohemia, NY 11716
As a minimum the following substances, preparations and components have to be removed from any separately
collected WEEE:
- Mercury containing components, such as switches or backlighting lamps (compact fluorescent lamps, CFL),
- Batteries
- Printed circuit boards if the surface of the printed circuit board is greater than 10 square centimeters (about 4 sq
in.),
- Toner cartridges, liquid and pasty, as well as color toner,
- Chlorofluorocarbons (CFC), hydrochlorofluorocarbons (HCFC) or hydrofluorocarbons (HFC), hydrocarbons (HC)
- Liquid crystal displays (together with their casing where appropriate) of a surface greater than 100 square
centimeters and all those back-lighted with gas discharge lamps,
- External electric cables
- Components containing refractory ceramic fibers as described in Commission Directive 97/69/EC of 5 December
1997 adapting to technical progress Council Directive 67/548/EEC relating to the classification, packaging and
labeling of dangerous substances (2),
- Electrolyte capacitors containing substances of concern (height > 25 mm, diameter > 25 mm or proportionately
similar volume)
2. The following components of WEEE that is separately collected have to be treated as indicated:
- Equipment containing gases that are ozone depleting or have a global warming potential (GWP) above 15, such
as those contained in foams and refrigeration circuits: the gases must be properly extracted and properly treated.
Ozone-depleting gases must be treated in accordance with Regulation (EC) No 2037/2000 of the European
Parliament and of the Council of 29 June 2000 on substances that deplete the ozone layer (4).

Low Temperature Torque Apparatus
Operation and Instruction Manual
K18854,55,64,65-Manual -8-
Table of Contents
1Introduction..................................................................................................................9
1.1 Koehler’s Commitment to Our Customers.........................................................................................9
1.2 Recommended Resources and Publications.....................................................................................9
1.3 Instrument Specifications...................................................................................................................9
2Safety Information and Warnings.............................................................................10
3Getting Started...........................................................................................................10
3.1 Packing List......................................................................................................................................10
3.2 Unpacking........................................................................................................................................11
3.3 Setup................................................................................................................................................11
4Descriptions...............................................................................................................12
4.1 Instrument Controls..........................................................................................................................12
4.2 Temperature Controller Operation...................................................................................................12
5Operation....................................................................................................................14
5.1 Preparation for Operation ................................................................................................................14
5.2 Operating Instructions......................................................................................................................14
5.3 Calibration Procedure......................................................................................................................15
6Safety Features..........................................................................................................16
6.1 Over-Temperature Protection..........................................................................................................16
6.2 Over-Power Protection.....................................................................................................................16
7Maintenance...............................................................................................................17
7.1 Routine Maintenance.......................................................................................................................17
7.2 Replacement Parts ..........................................................................................................................17
8Wiring Diagrams........................................................................................................18
8.1 K18864 Unit Wiring (220-240V 60Hz) .............................................................................................18
8.2 K18865 Unit Wiring (220-240V 60Hz) .............................................................................................19
9Service........................................................................................................................20
10 Storage.......................................................................................................................20
11 Warranty.....................................................................................................................20
12 Returned Goods Policy.............................................................................................20
Notes..................................................................................................................................21

Low Temperature Torque Apparatus
Operation and Instruction Manual
K18854,55,64,65-Manual -9-
1 Introduction
The Koehler Low Temperature Torque
Apparatus determines the extent to which a
sample grease retards rotation of a test bearing
when subjected to low temperatures. A grease
sample is packed into an automotive wheel
bearing or a slow speed ball bearing and cold
soaked at the test temperature. The spindle
(automotive grease) or inner bearing ring (ball
bearing grease) is then rotated at 1 RPM while the
restraining force on the test rig is measured and
displayed digitally. Torque values calculated from
the force determinations measure the viscous
resistance of the grease. Models equipped with
the optional data acquisition system provide a test
record for reporting purposes.
This manual provides important information
regarding safety, technical reference, installation
requirements, operating condition specifications,
user facility resource requirements, and operating
instructions for the Low Temperature Torque
Apparatus. This manual should also be used in
conjunction with applicable published laboratory
procedures. Information on these procedures is
given in section 1.2.
1.1 Koehler’s Commitment to Our
Customers
Providing quality testing instrumentation and
technical support services for research and testing
laboratories has been our specialty for more than
50 years. At Koehler, the primary focus of our
business is providing you with the full support of
your laboratory testing needs. Our products are
backed by our staff of technically knowledgeable,
trained specialists who are experienced in both
petroleum products testing and instrument service
to better understand your requirements and
provide you with the best solutions. You can
depend on Koehler for a full range of accurate and
reliable instrumentation as well as support for your
laboratory testing programs. Please do not
hesitate to contact us at any time with your
inquiries about equipment, tests, or technical
support.
Toll Free: 1-800-878-9070 (US only)
Tel: +1 631 589 3800
Fax: +1 631 589 3815
http://www.koehlerinstrument.com
1.2 Recommended Resources and
Publications
1. American Society for Testing and Materials
(ASTM)
100 Barr Harbor Drive
West Conshohocken, Pennsylvania 19428-
2959, USA
Tel: +1 610 832 9500
Fax: +1 610 832 9555
http://www.astm.org
ASTM Publication:
ASTM D1478: Standard Test Method for
Low Temperature Torque of Ball Bearing
Grease
ASTM D4693: Standard Test Method for
Low Temperature Torque of Grease
Lubricated Wheel Bearings
ASTM D4950: Standard Classification and
Specification of Automotive Service
Greases
2. Federal Test Method (FTM)
FTM Publication:
FTM 791-334: Grease Torque Test
1.3 Instrument Specifications
Models:
K18854, K18855
K18864, K18865
Electrical
Requirements:
220-240V 50Hz 20 A
220-240V 60Hz 20 A
Temperature
Range:
Ambient to -54°C (-65°F)
(Models K18854 and
K18864 only)
Optional
Temperature
Range:
Ambient to -73°C (-100°F)
(Models K18855 and
K18865 only)
Temperature
Control Stability:
±1°F (±0.5°C)
Refrigeration:
Air Cooled Mechanical
Cascade Hermetic System
Cabinet:
Floor-mount, polished
stainless steel exterior,
rides on swivel casters
Este manual sirve para los siguientes modelos
3
Tabla de contenidos
Otros manuales de Equipo de laboratorio de Koehler
Manuales populares de Equipo de laboratorio de otras marcas

Agilent Technologies
Agilent Technologies 5800 ICP-OES Manual de usuario

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Manual de usuario

NI
NI PXI-5422 Manual

Collomix
Collomix Aqix Manual de usuario

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Manual de usuario

Ocean Insight
Ocean Insight FLAME-NIR+ Manual de usuario