InTemp CX600 Dry Ice/CX700 Cryogenic Logger Manual
www.prosensor.com3
c. InTempConnect users:Log in with your InTempConnect
user credenals. Make sure to check the box that says “I
am an InTempConnect user” when signing in.
InTemp app only users: If you will not be using
InTempConnect, create a user account and log in when
prompted. DO NOT check the box that says “I am an
InTempConnect user” when signing in.
3. Congure the logger.Note that InTempConnect users
require privileges for configuring the logger.
Important: CX602 and CX702 loggers cannot be
restarted once logging begins. Do not connue with
these steps unl you are ready to use these loggers.
InTempConnect users: Configuring the logger requires
privileges. Administrators or those with the required
privileges can also set up custom profiles and trip
informaon fields. This should be done before compleng
these steps. If you plan to use the logger with the
InTempVerify app, then you must create a logger profile
with InTempVerify enabled. See
www.intempconnect.com/help for details.
InTemp App only users: The logger includes preset profiles.
To set up a custom profile, tap the Sengs icon and tap
CX600 or CX700 Logger before compleng these steps.
a. Press the buon on the logger to wake it up.
b. Tap the Devices icon in the app. Find the logger in the list
and tap it to connect to it. If you are working with
mulple loggers, press the buon again to bring the
logger to the top of the list.
If you are having trouble connecng:
•Make sure the logger is within range of your mobile
device. The range for successful wireless
communicaon is approximately 30.5 m (100 ) with
full line-of-sight.
•If your device can connect to the logger intermiently
or loses its connecon, move closer to the logger,
within sight if possible.
•Change the orientaon of your phone or tablet to
ensure the antenna in your device is pointed toward
the logger. Obstacles between the antenna in the
device and the logger may result in intermient
connecons.
•If the logger appears in the list, but you cannot connect
to it, close the app, power down the mobile device,
and then turn it back on. This forces the previous
Bluetooth connecon to close.
c.Once connected, tap Configure. Swipe le and right to
select a logger profile. Type a name or label for the
logger. Tap Start to load the selected profile to the
logger. InTempConnect users: If trip informaon fields
were set up, you will be prompted to enter addional
informaon. Tap Start in the upper right corner when
done.
4. Deploy and start the logger.
Important: Reminder, CX601 and CX602 loggers cannot
be restarted once logging begins. Do not connue with
these steps unl you are ready to use these loggers.
a. Deploy the logger to the locaon where you will be
monitoring the temperature.
b. Press the buon on the logger when you want logging to
begin (or if you chose a custom profile, logging will begin
based on the sengs in the profile).
If the logger was configured with alarm sengs, an alarm will
trip when the temperature reading is outside the range
specified in the logger profile. The logger alarm LED will blink
every 4 seconds, an alarm icon appears in the app, and an
Alarm Out of Range event is logged. You can review alarm
informaon in the logger report (see Downloading the Logger).
InTempConnect users can also receive noficaons when an
alarm is tripped. See www.intempconnect.com/help for more
details on configuring the logger and monitoring alarms.
Passkey Protection
The logger is protected by an encrypted passkey automacally
generated by the InTemp app for InTempConnect users and
oponally available if you are using the InTemp app only. The
passkey uses a proprietary encrypon algorithm that changes
with every connecon.
InTempConnect Users
Only InTempConnect users belonging to the same
InTempConnect account can connect to a logger once it is
configured. When an InTempConnect user first configures a
logger, it is locked with an encrypted passkey that is
automacally generated by the InTemp app. Aer the logger is
configured, only acve users associated with that account will
be able to connect to it. If a user belongs to a different account,
that user will not be able to connect to the logger with the
InTemp app, which will display an invalid passkey message.
Administrators or users with the required privileges can also
view the passkey from the device configuraon page in
InTempConnect and share them if needed. See
www.intempconnect.com/help for more details. Note: This
does not apply to InTempVerify. if the logger was configured
with a logger profile in which InTempVerify was enabled, then
anyone can download the logger with the InTempVerify app.
InTemp App Only Users
If you are using the InTemp app only (not logging in as an
InTempConnect user), you can create an encrypted passkey for
the logger that will be required if another phone or tablet
aempts to connect to it. This is recommended to ensure that
a deployed logger is not mistakenly stopped or purposely
altered by others.
To set a passkey:
1. Press the buon on the logger to wake it up.
2. Tap the Devices icon and connect to the logger.