MAINTENANCE
ROUTINE MAINTENANCE
The DDU-2200 AED is designed to be very low maintenance. Simple maintenance tasks are
recommended to be performed regularly to ensure its readiness (see sample maintenance
table below). Different maintenance intervals may be appropriate depending on the environ-
ment where the AED is deployed, and ultimately the maintenance program is at the discretion
of the emergency response program’s medical director.
Daily Monthly After Each Use Action
• • •
Check that the Active Status Indicator is flashing green
• •
Check the condition of the unit and accessories
•
Run manually-initiated self-test
•
Replace pads
•
Check pad and battery pack expiration dates
•
Check the DDC card, if one was installed
Note: If the unit has been dropped, mishandled, or abused, a manually-initiated self-test
should be performed.
If the unit still requires attention after a manually-initiated self-test has been performed, refer
to “Troubleshooting” on page 22 or call Defibtech for service. For contact information, refer
to the “Contacts” section on page 34.
Check the condition of the unit and accessories
SAMPLE
Check the condition of the unit and accessories
Run manually-initiated self-test
SAMPLE
Run manually-initiated self-test
Replace pads
SAMPLE
Replace pads
Check pad and battery pack expiration dates
SAMPLE
Check pad and battery pack expiration dates
Check the DDC card, if one was installed
SAMPLE
Check the DDC card, if one was installed
MAINTENANCE MODE (AED Main Menu)
Maintenance Mode permits the user to perform
maintenance-related actions, such as viewing
unit information, initiating unit self-tests, changing
unit parameters, downloading rescue data, and
upgrading software.
To enter Maintenance Mode, press
the bottom softkey button to the right
of the Tool Icon (shown at left) on the
AED Status screen (for instructions on
how to access the AED Status screen, see previous
page). The AED Main Menu will now be displayed,
as shown at right. If the AED is needed to perform
a rescue while in Maintenance Mode, navigate to
and select the “Rescue now” menu option.
AED Main Menu screen
(Maintenance Mode)
Rescue now
AED status
AED maintenance
AED options
Rescue options
Help topics
Turn AED off
AED Main Menu
MAINTENANCE MODE (AED Maintenance)
The AED Maintenance screen allows the user to select
such options as AED tests, software upgrades, data
backups, and data card functions.
For detailed information about each of the functions
that can be accessed from this screen,
refer to the
DDU-2200 User Manual, which can be viewed at or
downloaded from www.defibtech.com.
MAINTENANCE MODE (Help Topics)
The Help Topics screen provides a list of available videos
that detail various aspects of using and maintaining the
DDU-2200 AED.
To exit Maintenance Mode, turn the unit off by pressing
the ON/OFF button or navigate to the AED Status Screen
by selecting the “Go to main menu” option and then
navigate to and select the “Turn AED off” option.
MAINTENANCE MODE (AED Options)
The AED Options screen allows the user to manually
configure AED options such as time, date, volume, and
audio recording.
For detailed information about each of the functions
that can be accessed from this screen,
refer to the
DDU-2200 User Manual, which can be viewed at or
downloaded from www.defibtech.com.
MAINTENANCE MODE (Rescue Options)
The Rescue Options screen allows the user to manually
configure rescue options such as rescue protocol and
CPR breathing.
For detailed information about the functions that can
be accessed from this screen,
refer to the DDU-2200
User Manual at www.defibtech.com.
Rescue now
Perform AED test
Upgrade AED
Transfer data to card
Format data card
Run application from card
Go to main menu
AED Maintenance
Rescue now
Preparing the patient
Analyzing and shocking
Performing CPR
Replacing the battery
Replacing the pads
Checking the AED status
Replacing the data card
Go to main menu
Help Topics
Rescue now
System time hh:mm:ss
System date yyyy mm dd
Volume level High
Audio recording Disabled
Status broadcast Disabled
Go to main menu
AED Options
Rescue now
CPR breathing Disabled
Rescue protocol AHA
Settings
1 / 2 / 30 / 50 / 100
Go to main menu
Rescue Options