2. Indicator showing “0000” ZERO SPAN out of range
Incorrect weighing result - void damages by external
environment force such as free-drop to the ground, collision by
external objects, etc.
Proper re-calibration procedure required to correct the setting of
weighing accuracy.
Interference due to RF disturbance, ground vibration…etc.
Unstable platform feet adjustments according to bubble level
indication
Incorrect position or other external objects within weighing area
The weighing-scale is not put in a solid & firm ground area, such
as carpet floor or lawn.
3. onnection failure for data transmission to P or printer
Wrong connection wires or faulty wires for transmission between
the digital indicator & load cells.
Wrong indicator models
Wrong internal wiring or wire broken
In case of the following defective mode occurs, it is suggested to contact
your nearest Authorized Dealer for further technician service & repair:
1. POWER switch-on failure :
Push-button faulty
Short circuit wires - Wire broken
Safety fuse burnt out
Wire connection problem
Main power adaptor faulty – Parts Replacement
2. L D display faulty
Possible hardware defects include: Uneven brightness in the LCD
5
WEIGHING OPERATION
Before reading detailed instructions on how to use all the weighing
functions that are built into your scale, please read the following
important guidelines:
Always be sure that the display shows `Zero` before use, if it does
not then please press the ZERO key.
The device is designed to detect when a stable weight is achieved,
the indicator will `bleep` twice to indicate a stable weight value, your
reading should be taken at this point.
WARRANTY-LIABILITY
If a fault or defect is present on receipt of the unit which is within
CHARDER MEDICAL’s scope of responsibility, CHARDER shall
have the right to either repair the fault or supply a replacement unit.
Replaced parts shall be the property of CHARDER. Should the fault
repairs or replacement delivery not be successful, the statutory
provisions shall be valid. The period of warranty shall be two years,
beginning on the date of purchase. Please retain your receipt as
proof of purchase. Should your scale require servicing, please
contact your dealer or CHARDER MEDICAL Customer Service.
No responsibility shall be accepted for damage caused through any
of the following reasons: Unsuitable or improper storage or use,
incorrect installation or commissioning by the owner or third parties,
natural wear, changes or modifications, incorrect or negligent
handling, overuse, chemical, electrochemical or electrical
interference or humidity, unless this is attributable to negligence on
the part of CHARDER MEDICAL.
If operating, climatic or any other influences lead to a major change
in conditions or material quality, the treaty for perfect unit