Step 4 - Weighing
Step 5 - Calibration/Single Point
A. To get the scale ready to weigh a patient, rst press the TARE
button once to make the indicator display “0”.
B. Place the patient on the platform.
C. Once the “kg, lb, or pcs” character stops ashing, you can begin
to read it.
A. Under the condition of weighing, long press the UNIT button until
the window displays “------“.
B. Short press the UNIT button or HOLD button to switch and make
the window display CAL.
C. Short press the TARE button to make sure you enter the
calibration mode.
D. After entering calibration mode, the window will display “LOAD”
and “0.0 kg” alternately.
a. This means the weight “0.0 kg” should be loaded on the
platform.
E. After making sure that there is nothing on the platform, short
press the T button to make the scale begins with “0.0 kg”
calibration.
a. “0.0 kg” will icker rapidly.
F. Keep the platform stable and without other interference factors
that will inuence the weighing performance.
a. If the weighing keeps stable exceeding 2 seconds, the scale
will nish “0.0 kg” calibration automatically and will begin the
next calibration point.
G. If it doesn’t keep stable, the “0.0 kg” will keep ickering while
waiting for stable weight.
a. Repeat steps D. and E., calibrate the half capacity range point
and the full capacity range point orderly, save the calibrated
parameters automatically and return to the weighing mode.