GETTING STARTED
1. INSTALLATION –SET-UP
•
Remove the red rubber cap from the optical sensorconnector
(SMA)
•
Remove the protective cap from the male plug on the optical
fiber and insert it in the SMA connector of the NomaSense
O
2
P6000. The safety nut has to be screwed on.
•
Insert the temperature probe by matching the two red marks
of the connector and the plug.
2. STARTING THE DEVICE
•
Hold the button on the control panel to turn on
the device, after some automatic tests you reach
themeasurement screen directly.
•
The navigation bar at the bottom shows the functions of
therespective control buttons.
•
The status bar at the top shows the time, the name of
thecurrent selected sensor, the name of the current
user, a symbol indicating if logging is activated or not
and the battery status.
3. CREATE A NEW SENSOR WITH A QR CODE
•
Activate the QR code reader by pressing the (Scan)
buttonwhile on the measurement screen.
•
Place the respective sensor QR code (delivered with your
oxygen sensor) about 10 cm from the NomaSense O
2
P6000 ; the target cross has to be aimed at the QR code.
Reducing or increasing the distance between QR code and
QR code reader can speed up the scanning process.
•
A notification window opens on the screen, when
thescanning process is done.
•
A keyboard screen opens. Use the navigation arrows to type in the new Sensor Name.
•
When you have finished typing go to the Done button and press .
•
The new sensor shows in the sensor list. Press once more to select the new sensorand the display
will switch to the measurement screen automatically.