TPS ProChem R Manual de usuario

ProChem_R
TPS Pty Ltd
ABN 30 009 773 371
Phone: (07) 32058027
International: 61 732058027
Email: [email protected]
Web: www.tps.com.au

Congratulations !
Your new proCHEM-R controller is the latest in simple, reliable process
control instrumentation. With correct operation and maintenance, your
proCHEM-R will give you many years of reliable service.
The proCHEM-R is a breeze to operate. This manual has been
designed to help you get started, and also contains some handy
application tips. If at any stage you require assistance, please contact
either your local TPS representative or the TPS factory in Brisbane.
The manual is divided into the following sections:
1. Table of Contents
Each major section of the handbook is clearly listed. Sub-sections
have also been included to enable you to find the information you
need at a glance.
1. Introduction
The introduction has a diagram and explanation of the display and
controls of the proCHEM-R. It also contains a full listing of all of the
items that you should have received with the unit. Please take the
time to read this section, as it explains some of items that are
mentioned in subsequent sections.
1. Main Section
The main section of the handbook provides complete details of the
proCHEM-R, including operating modes, calibration, troubleshooting,
specifications, and warranty terms.
1. Appendices
Appendices containing background information and application notes
are provided at the back of this manual.
2

Contents
1. Introduction 5
1.1 proCHEM-R Illustration 5
1.2 Unpacking Information 8
1.3 Specifications 9
2. Operating Modes 10
3. Installation and Set-up 11
3.1 Connection and Configuration Diagram 11
3.2 Mounting the Enclosure 12
3.3 Mounting the Sensor 13
3.4 Terminal Connections 14
3.5 Setting the Control Limit 16
3.6 Selecting Current or Voltage Output 19
3.7 Selecting Standard or Custom 4 to 20 mA Output 20
3.8 Calibrating the 4 to 20mA Output 21
3.9 CMRR (RV7) Control 21
3.10 Factory Configuration (J3) Jumpers 22
4. Calibration 23
4.1 Calibration Procedure 23
4.2 Calibration Notes 25
5. Process Monitoring 26
6. Automatic Control 26
7. Fuse Replacement 26
8. Troubleshooting 29
8.1 Instrument Function Troubleshooting 29
8.2 Redox Troubleshooting 29
9. Warranty 30
10. Appendices 33
10.1 Power Terminal connections when 12V DC option is fitted 33
10.2 Relay Output Wiring Examples for mains powered units 34
3

1. Introduction
1.1 proCHEM-R Illustration
5

Set Upper Limit Control
Adjust this control to the Millivolt value at which the High LED and
relay output are activated. Factory-set to activate when the reading is
above the Upper Limit. See section 3.5.
High LED
The High LED provides visual indication of when the Upper Limit has
been exceeded. Factory-set to light up when the reading is above the
Upper Limit. See section 3.5.
Set Lower Limit Control
Adjust this control to the Millivolt value at which the Low LED and
relay output are activated. Factory-set to activate when the reading is
below the Lower Limit. See section 3.5.
Low LED
The Low LED provides visual indication of when the Lower Limit has
been exceeded. Factory-set to light up when the reading is below the
Lower Limit. See section 3.5.
Zero Calibrate Control
Adjust this control to calibrate the proCHEM-R in a Redox standard.
See section 4.
Span Calibrate Control
This control is not required for the proCHEM-R.
Function Switch
The Function Switch is used to select the mode of operation. See
section 2.
Display Window
The proCHEM-R has a large, easy to read LCD display. The units of
measurement are clearly shown in the window beneath the display.
Terminal Cover
6

The terminal cover provides easy access to all of the connections,
relay output fuses, configuration jumpers and user-adjustable
trimmers. The cover is water resistant to IP65.
7

1.2 Unpacking Information
Before using your new proCHEM-R, please check that the following
accessories have been included:
Part No
1. proCHEM-R Process Controller 111122
1. ZoBell Redox Standard, 229mV, 200mL ea Part A & Part B
121309
1. proCHEM-R Handbook 130050
Options that may have been ordered with your proCHEM-R:
Dip mounted Redox Sensors, and submersible assembly…
1. Submersible, Non-flow Redox Sensor, 5m 111260
1. Intermediate Junction Redox Sensor, 5m 121267
1. Stainless Steel solution earth rod, 5m 121360
1. Submersible protective housing for Redox & earth rod 111303
Screw-in Redox Sensor, and adaptor for 40mm PVC pipe…
1. Screw-in, Non-flow Redox Sensor, 5m 111262
2. Stainless Steel solution earth rod, 5m (not screw-in) 121360
1. ¼” BSP thread adaptor for 40mm PVC pipe 111305
Instrument options…
1. 4-20mA loop-Rowered, remote LCD 130080
2. 12V DC Power Option 130072
8

1.3 Specifications
Range 0 to 1000 mV
Resolution 1 mV
Accuracy ±1 mV
Linearity ±1 mV
Repeatability ±1 mV
Ambient Drift <0.02% Span per 1OC change
Long term drift<0.1% per year
Input Impedance >3 x 1012 Ω
Zero Range ±100 mV
Common Mode Rejection 120 dB
Enclosure Polycarbonate, waterproof to IP65
Display 12.7 mm LCD (optional remote 4-20mA loop
powered display available)
Analogue Outputs 4 to 20mA (500 Ohms max. loop resistance )
0 to 1 V DC (min. load 1000 Ω)
0 to 10 V DC optional (min. load 1000 Ω)
Control Outputs 2 x Clean contact changeover relays,
rated to 2A at 240V AC
Isolation Galvanic isolation of sensor input
Power 240V AC, 50/60Hz (120V AC optional)
Dimensions 215 x 190 x 115 mm
Mass Instrument only : Approx 900g
Full Kit : Approx 1.5kg
Operating Environment: Temperature : 0 to 45 oC
Humidity : 0 to 95 % R.H.
9

2. Operating Modes
The function switch is used to select the required mode. The four
operating modes available are…
1. Set Upper Limit : Switch to Set Upper
Limit when setting the Millivolt value at which
the High LED and relay output are activated.
The relay output is not operational in this mode.
Factory-set to activate when the reading is
above the Upper Limit. Can be user set to be a
activate when the reading is below the Upper
Limit. See section 3.5.
2. Set Lower Limit : Switch to Set Lower
Limit when setting the Millivolt value at which
the Low LED and relay output are activated.
The relay output is not operational in this mode.
Factory-set to activate when the reading is
below the Lower Limit. Can be user set to be a
activate when the reading is above the Lower
Limit. See section 3.5.
3. Calibrate : Switch to Calibrate mode to
place the proCHEM-R into standby (eg. when
calibrating). This mode disables the two output
relays to avoid chemicals being added when
placing the sensor into calibration buffers.
4. Automatic Control : Switch to
Automatic Control to enable the Alarm LED’s
and relay outputs. These will be activated when
the Millivolt reading is outside the Upper and
Lower limit values. See section 3.5.
10