Tol Watertechniek HD 200 Manual de usuario

Manual
Purusaqua
ORP controller

© Tol Watertechniek Page 2 10-1-2021
Dear Customer
We are honoured that you have chosen for the products of Tol WaterTechniek (hereinafter referred to
as TWT). Read this manual carefully so you will use the product correctly and so a correct functioning
is guaranteed. If after reading this manual or while using our product you have any questions, please
contact your dealer or TWT. We will be happy to help
!

© Tol Watertechniek Page 3 10-1-2021
Table of contents
Table of contents ........................................................................................................................ 3
1 EG declaration.................................................................................................................... 4
2 Safety.................................................................................................................................. 6
2.1 Intended scope............................................................................................................ 6
2.2 Risks in the use of ozone equipment.......................................................................... 6
3 Requirements regarding the installation environment........................................................ 7
4 ORP controller.................................................................................................................... 8
4.1 Explanation of the controls ........................................................................................ 8
4.2 Installing the ORP controller...................................................................................... 9
4.3 Checking the right connection of the ORP controller ................................................ 9
4.4 Setpoint and hysteresis ............................................................................................... 9
4.4.1 Adjusting the setpoint....................................................................................... 10
4.4.2 Setting the hysteresis........................................................................................ 10
4.5 Calibrating the controller ......................................................................................... 11
4.6 Maintenance ............................................................................................................. 12
4.7 Storing the electrode................................................................................................. 12
4.8 Problem solving........................................................................................................ 12
4.9 Technical data .......................................................................................................... 13
5 Miscellaneous peripherals ................................................................................................ 14
5.1 Flow-trough assembly .............................................................................................. 14
6 Warranty determination.................................................................................................... 15
7 Disclaimer ........................................................................................................................ 16
8 Contact information.......................................................................................................... 17

© Tol Watertechniek Page 4 10-1-2021
1 EG declaration

© Tol Watertechniek Page 5 10-1-2021

© Tol Watertechniek Page 6 10-1-2021
2 Safety
2.1 Intended scope
The products described in this manual are intended to be used for the treatment of slightly
contaminated water. Water of pools, ponds, aquariums and light industry or similar for example.
The use of these products in a different situation than recommended may result in personal injury and
possible premature wear of the equipment.
2.2 Risks in the use of measuring equipment
Despite the fact that our products are designed in such a way that the risks are minimalized there is
always a risk of dangerous situations. During the use of our measuring equipment various parts are
connected to the mains voltage. Never open the enclosure of our measuring equipment if the power
lead is still connected to the mains.

© Tol Watertechniek Page 7 10-1-2021
3 Requirements regarding the installation environment
Measuring equipment should be placed in a dry and cool environment protected against direct sunlight
and rain.
It is of great importance that the cable inlet of the electrode remains above water at all times.
Therefore it is preferred to use a flow trough assembly to mount the electrode . Electrodes are not
suited in systems operating under high pressure. The use is only permitted in places with a pressure
below 0.3 bar.
Defects caused by neglecting these installation requirements will not be covered by warranty!

© Tol Watertechniek Page 8 10-1-2021
4 ORP controller
4.1 Explanation of the controls
Display (green screen)
On the display the measured values are shown such as the current ORP value, setpoint and the status
of the output. In the shown photo the "value screen" is shown.
Menu button
If the menu button is pressed while the value screen is shown the setting menu will be shown. In this
menu the controller can be calibrated and the parameters of the output can be set. If the menu button
is pressed a second time the display will show the value screen again.
↑Scroll button "up"
The up button is used to navigate through the menu and to adjust settings. The setpoint and
calibration parameters can be adjusted with this button.
↓Scroll button "down"
The down button is used to navigate through the menu and to adjust settings. The setpoint and
calibration parameters can be adjusted with this button.
Ok button
The Ok button is used to select a menu item or to confirm a setting.

© Tol Watertechniek Page 9 10-1-2021
4.2 Installing the ORP controller
Uninstall the lid by turning the 4 screws in the lid one quarter. De screws will pop up en are raised
above the lid. Now you can fold the lid open.
Mount the ORP controller on a flat surface in a dry environment en screw the lid
back in place. Only mount the ORP controller in a dry and well ventilated
environment protected from direct sunlight. Only connect the ORP controller to a
wall socket with a safety earth connection!
It is advised to use a flow trough assembly to mount the electrode (available by
your dealer).You can place the electrode in a sieve or vortex as well but be
careful that the wire connection on the electrode remains above the water level
at all time.
4.3 Checking the right connection of the ORP controller
There are a few simple tests which you could perform to test if the controller is connected right.
•Place your electrode in a glass of tap water, the controller should show a value between 150
and 300mV.
•Gently wipe your finger over the electrode tip, the shown value should change quite rapidly.
4.4 Setpoint and hysteresis
The ORP controller has an adjustable relay output. This can be used to control a dosing pump or a
ozone generator.
The value at which this relay should be controlled is referred to as "setpoint". You also can adjust the
hysteresis. The hysteresis prevents that the relay will flicker around its setpoint, below an example.
Assume that the setpoint is set to 275mV and the hysteresis is set to 5mV. As soon as the measured
value becomes below 270mV (setpoint - hysteresis) the relay will be activated. The measured value
need to be raised to 280mV (setpoint + hysteresis) to get deactivate.
At a hysteresis of 0mV the relay would be activated at a measured value below 275mV and
deactivated at a value higher as 275mV. Thus the relay would flicker around the setpoint of 275mV,
this is prevented by the hysteresis.

© Tol Watertechniek Page 10 10-1-2021
4.4.1 Adjusting the setpoint
We advice you to adjust the setpoint according to the table shown on this page. The
hysteresis is preset at 5mV in the factory, for most applications this is a good setting.
To adjust the setpoint press the menu button once, the screen below is shown:
Next press the "ok" to confirm that you want to adjust the setpoint. The following screen
appears which ask you to adjust the setpoint:
You can adjust the setpoint to the desired value by pressing the ↑and ↓buttons. If the setpoint is
adjusted to the desired value you must confirm the setting by pressing the "ok" button. The following
screen appears which ask you to set the hysteresis:
4.4.2 Setting the hysteresis
The hysteresis is set by pressing the ↑and ↓buttons. If the hysteresis is adjusted to the desired value
you need to confirm the setting by pressing the "ok" button. The setpoint and hysteresis are now set
and the display of the ORP controller now will show the menu again. By pressing the ↑and ↓buttons
you can scroll through the menu. By pressing the "menu" button the ORP controller will return to the
value screen.
pH
Max. mV
6
330
6,1
330
6,2
320
6,3
320
6,4
310
6,5
310
6,6
300
6,7
290
6,8
290
6,9
280
7
270
7,1
270
7,2
260
7,3
260
7,4
250
7,5
250
7,6
240
7,7
230
7,8
230
7,9
220
8
210