5.8.4 Ifitis necessary to interrupttheoperation,for exampleto replace acylinder,thePAUSE feature
can beused.See section 5.10 for detailed instructions.
5.8.5 When displayed valuereaches programmed amount,four beeps will beheard,thevalvecloses,
"
Auto
"
enunciator will disappear from thedisplay and theamountofrefrigerantcharged will
remain displayed.
5.8.6 Leavethemanifold valve(s) open for few seconds in order to draw therefrigerantfrom the
supply hoseinto thesystem.Closemanifold valves.
5.9 Recovery
5.9.1 Open themanifold valves or switch on recovery unit,to fill thehosebetween the
manifold and the9055.Then open therefrigerantcylinder valve.
5.9.2 Initiateflow by pressing theGO Key.Thevalvewill open,theunitwill beeptwice and refrigerant
will begin to flow into thecylinder.The
"
Auto
"
enunciator will appear on thedisplay and thedis-
play will countup as therefrigerantis added.
5.9.3 Ifitis necessary to interrupttheoperation,for exampleto replace thecylinder,thePAUSE
featurecan beused.See section 5.10 for detailed instructions.
5.9.4 When displayed valuereaches programmed amount,four beeps will beheard,thevalvecloses,
"
Auto
"
enunciator will disappear from thedisplay and theamountofrefrigerantrecovered will
remain displayed.
5.9.5 Closemanifold valves and cylinder valve.
5.10 PauseFeature
5.10.1 The9055 features aPAUSE/CANCEL Key thatmay beused during automaticdispensing to
interruptflow or when manually charging to hold thedisplayed value.This is usefulifit
becomes necessary to switch refrigerantcylinders or whenallowing system pressures to equal-
ize.Itis also usefulwhen programming a criticalcharge(less than 8 ounces) and itis desirable
to lock thescaledisplay to preventdrift or bounce,when purging ahose,opening and closing
avalveor heating/shifting thecylinder.
5.10.2WhileinAutomaticMode:Press thePAUSE/CANCEL Key.This will freezethedisplay,thevalve
will close,refrigerantflow will stop,the
"
Auto
"
enunciator will flash and theunitwill beep every
second.
5.10.3 To resumeautomaticdispensing,press theGO Key.Thevalvewill reopen,refrigerantwill flow,
the
"
Auto
"
enunciator will stop flashing and theunitwill stop beeping.
5.10.4WhileintheManualMode:Press thePAUSE/CANCEL Key.This will freezethedisplay,theunit
will beep every second and the
"
Auto
"
enunciator will flash.Caution:Only do this ifyou arenot,
or havestopped,charging.With theManualPauseactivated,any weightadded or removed
from theplatform will beignored.
5.10.5 To resumethe
"
metering
"
ofrefrigerant,press thePAUSE/CANCEL Key again.
The
"
Auto
"
enunciator will stop flashing and theunitwill stop beeping.
5.11 CancelFeature
5.11.1 To cancelan inputpress thePAUSE/CANCEL Key.
5.11.2 To Canceltheoperation whilein theautomaticMode,atany timeafter theGO Key has been
pressed,press thePAUSE/CANCEL Key twice.The
"
Auto
"
enunciator will disappear and the
amountofrefrigerantcharged/recovered will remain displayed.Caution:Once this is done,the
remaining programmed cyclecannotbecompleted.
OPERATING INSTRUCTIONSCONT-