17
1. Application
The VarioSafe EXM 25 is an intrinsically safe multime-
ter which can be used in areas with a potentionally
explosive atmosphere in accordance with ATEX (EN
60079-0 and EN 60079-11) and EN/IEC 61010 for vol-
tage, resistance, current, frequency measurements
with measurement accessories for measurement of
temperature and high current.
2. Safety Precautions
When used for its intended purpose, safety of the
operator, as well as that of the instrument, is assured.
The voltage measurement tips contain moulded
multipliers within both test prods. They are extreme-
ly safe and comply with overvoltage category CAT IV.
In order to maintain awless technical safety condi-
tions, and to assure safe use, it is imperative that you
read these operating instructions thoroughly and
carefully before placing your instrument into service,
and that you follow all instructions contained
therein.
†
The device may only be used in the designated
Ex zones (section 5.1) and inside of the safety-
related limit values (section 5.2).
†
Before starting resistance measurements make
sure that the test object is at zero-potential.
†
Perform zero-potential tests at most up to the
following voltages:
Ex-class II up to 690 V, Ex-class I up to 1000 V and
with measurement tip 72011 up to 2000V.
†
Before putting into operation for the rst time and
at every day of use the VarioSafe EXM 25 has to be
checked at a known voltage source - e.g. a 230 V
socket to ensure that it works properly and
faultlessly (see display test and self test). If
indication of one or several systems fail in the
course of performing the self test according to
section 3, if function standby is not indicated or if
the device is damaged, the VarioSafe EXM 25 must
not be used again.
†
Maintainance is only allowed by the manufacturer
or explicitly authorised repair shops
(see section 7).
†
The measurement tips with extended (12 mm)
not insulated test electrodes are designed for use
at untouchable machinery materials. You may use
the extentions in EX-areas only when an uninten-
tional connection between conductive parts
(e.g. by short-circuit between test electrodes) can
denitely be excluded.