Symbol5.3.7(ISO7000-0632)
of ISO15223-1:2012
Temperature limits within which the device
can be safely exposed
Symbol5.3.8(ISO7000-2620)
of ISO15223-1:2012
Humidity limits within which the device
can be safely exposed
Symbol5.3.9(ISO7000-2621)
of ISO15223-1:2012
Pressure limits within with the device
can be safely exposed
“CE marking” Product compliant with EC Directive93/42/EEC
and subsequent amendments
Table3: Symbols on this device
5.1.3 Disclaimer
The warranty is valid eighteen months from the manufacturing date.The warranty covers any fault, material damage or manufacturing
defect in products used in accordance with the instructions in this leaflet.
This warranty does not preclude the application of current legal guarantees under national legislation governing the sale of consumer
goods.
Laboratoires THEA is not responsible for:
• Any damage resulting from disregarding what is described in this manual
• Any damage resulting from malfunctioning caused by a combination of connected devices
• Any damage resulting from transport, improper use or negligence, incorrect handling, modification of the system, poor
maintenance, use of wrong voltage, lightning, infiltration of sand or water, use of parts or accessories not provided or
recommended in this manual by Laboratoires THEA
5.2 WARNINGS & CAUTION
USAGE NOTES
BLEPHASTEAM®must only be used for the intended use described in this manual.
The device must be used in the specified ambient conditions.
WARNINGS
• Do not immerse the whole device in water or clean it in running water
• Do not use BLEPHASTEAM®near a mobile phone or radio emitter
• Do not touch the batteries location with a wet hand. Otherwise, it may cause electrical shock
• Do not short-circuit the batteries location. Otherwise it may cause fire or electrical shock
• Do not disassemble, modify or repair the instrument by yourself. Otherwise, it may cause fire, electrical shock, bodily injury, or
instrument malfunction. Refer all servicing to Laboratoires THEA or your authorized THEA dealer.The instrument disassembled,
modified or repaired by anyone other than a Laboratoires THEA designated repair facility will void the warranty
• Only use the specic rechargeable batteries provided by Laboratoires THEA. Use of batteries not specified by Laboratoires
THEA may cause fire or instrument malfunction
• If there is any abnormal odour, sound, heat, or smoke when power “ON” the device, turn the BLEPHASTEAM®switch
“OFF” immediately. Continued use may cause fire or instrument malfunction. Contact Laboratoires THEA or your authorized
Laboratoires THEA dealer for inspection
• BLEPHASTEAM®is vulnerable to damage caused by ESD (Electro Static Discharge), so avoid contact between the device and
your clothing
• Do not bend, crush or excessively strain the cable
• Do not attach or tighten cable to or around the head or neck. Cable can cause strangulation
• Do not expose the device to moisture. BLEPHASTEAM®is not water-proof and does not protect against the ingress of water or
moisture
• Do not place BLEPHASTEAM®in a microwave
• Do not autoclave BLEPHASTEAM®or its components
• Do not connect the device up to another appliance or to another source of power not provided by Laboratoires THEA. Otherwise,
it may lead to malfunctioning of the electromedical device or expose the user to higher electrical risks
• Do not operate the device in a hazardous environment that has a risk of explosion or contains volatile solvents (alcohol, etc.)
or flammable materials (anaesthetics, etc.) in the vicinity of the device
• Do not expose the device to very high temperatures or flames
• Do not use the device with a damage battery or charger (broken case, poor contact, broken feed cable). If the batteries are
damaged (cracked case, leakage of electrolyte, off-shape etc.) they should be replaced by the authorized personnel. Usage of
modified batteries may cause an explosion and/or damage the device
• Never attempt to access the internal hardware for any purpose including maintenance while the device is in operation. When
a function check is not successful, the user is prompted to contact technical service or the manufacturer
• Small Children: Do not leave your BLEPHASTEAM®and its accessories within the reach of small children or allow them to
play with it. They could hurt themselves or others or could accidentally damage the BLEPHASTEAM®. Your BLEPHASTEAM®
contains small parts with sharp edges that may cause an injury or may become detached and create a choking hazard
• Animals or insects: Do not leave your BLEPHASTEAM®and its accessories within the reach of animals or insects. After
treatment, your device must be put back in its original packaging
• The use of accessories other than those specied for the device is not recommended. They may result in increased
emissions or decreased immunity of the device
CAUTION
• This user manual must always be accessible and close to the device.The instructions manual should be accessible to the user
at all times and the user should consider all instructions carefully before using the device
• Only use the batteries supplied with the BLEPHASTEAM®. Use of batteries from other sources may cause damage, fire, injury
or instrument malfunction. For more information, see “Precautions for batteries” described in §6.
• Do not wipe the exterior of the mask with chemical products or solvent (e.g. acetone or ethanol). It may lead to dscolouration
or deterioration
• Do not allow terminal to come into contact with water as this may cause failure of the device. If BLEPHASTEAM®is dirty, wipe it
with a soft cloth containing no water
• Do not install the instrument on unstable location such as on a shaky base or a tilting surface. Otherwise, the instrument may
drop or fall over, causing a bodily injury
• If the device experienced an external mechanical impact (knocking, bumping, dropping, etc.), this type of shock may cause
malfunctioning of the device. In case of malfunctioning, please contact Laboratoires THEA or your authorized Laboratoires
THEA dealer for technical support
• Remove and store the batteries if the main unit will not be used for extended periods. Failure to do so may result in dead
batteries
• Insert the batteries in the battery compartment or the device may not work properly
• When using or carrying the instrument, hold it or attach it firmly. Dropping the instrument may lead to injury
• Periodic maintenance of BLEPHASTEAM®is not required by the user, only daily inspection is recommended. Maintenance of
the BLEPHASTEAM®is required when the device does not pass its function check. When the function check is not successful,
the user is prompted to contact technical support or the manufacturer (see last page)
5.3 ELECTRICAL SAFETY
BLEPHASTEAM®is supplied with a power plug cable. Do not use any other cable. Use the cable supplied with the device to connect the
device to the main power socket.
Position BLEPHASTEAM®so that the power cable is always accessible.
5.4 COMPLIANCE WITH STANDARDS / REGULATION AND CLASSIFICATIONS
Compliant standards Description Classication
IEC60601-1
• According to the type of protection against electric
shock
• According to the degree of protection against
electrical shock
• According to the mode of operation
ClassII (double isolation)
TypeBF
Continuous operation
IEC60601-1-2 • Electromagnetic compatibility * Refer EMC section10
IEC60529 • According to the type of protection against ingress
of water as detailed in the current edition of IEC
IP22 for the mask & IP21
for the base station
93/42EEC-2007 • According to the EU Medical Device Regulation Class IIa
Table4: Mains standards & Classification