TECNOLOGIC spa - TLK 43 - OPERATING INSTRUCTIONS - Vr.02 - ISTR 06267 - PAG. 4
This parameter can be programmed as :
=SE : The active Set Point can be modified while the alarm
thresholds cannot be modified.
=AE : The active Set Point cannot be modified while the alarm
thresholds can be modified
=SAE : Both the active Set Point and the alarm thresholds can be
modified
=SAnE : Both the active Set Point and the alarm thresholds cannot
be modified
2.4 – CONTROL STATE
The controller can act in 3 different ways : automatic control (rEG),
control off (OFF) and manual control (OPLO).
The instrument is able to pass from one state to the other :
- by selecting the desired state from the main selection menu suing
the keyboard.
- By using the key “U” on the keyboard; suitably programming par.
“USrb” (“USrb” = tunE; “USrb” = OPLO; “USrb” = OFF) it is possible
to pass from “rEG” state to the state programmed on the parameter
and vice versa.
- Automatically (the instrument swaps into "rEG" state at the and of
the auto-tuning execution)
When switched on, the instrument automatically reassumes the
state it was in when it was last switched off.
AUTOMATIC CONTROL (rEG) – Automatic control is the normal
functioning state of the controller.
During automatic control, on the SV display, it is possible to
visualize the control power on the display by pushing key “UP”.
The range of the power values goes from H100 (100% of the output
power with reverse action) to C100 (100% of the output power with
direct action).
CONTROL OFF (OFF) – The instrument can be swapped into the
“OFF” state, i.e. the control and the relative outputs are deactivated.
The alarm outputs are instead working normally.
BUMPLESS MANUAL CONTROL (OPLO) – By means of this
option it is possible to manually program the power percentage
given as output by the controller by deactivating automatic control.
When the instrument is swapped to manual control, the power
percentage, visualised on the SV display, is the same as the last
one supplied and can be modified using the “UP” and “DOWN”
keys.
In case of ON/OFF control, 0% corresponds to the deactivated
output while any value different from 0 corresponds to the activated
output.
As in the case of visualization, the programmable values range from
H100 (100% output power with reverse action) to C100 (100%
output power with direct action).
In case of motorized actuators with time positioning, the manual
control of the output is obtainable in the following way :
- By pushing the UP key, the opening of the actuator is driven
- By pushing DOWN key, the closing of the actuator is driven
For all time during which the manual control is active, “3 Pt” or
“OPEn” is visible on the lower display if the UP key is pressed or
“CLOS” if the DOWN key is pressed.
To return to automatic control, select "rEG" in the selection menu.
2.5 – ACTIVE SET POINT SELECTION
This instrument permits pre-programming of up to 4 different Set
points (“SP1”, “SP2”, “SP3”, “SP4”) and then selection of which
one must be active. The maximum number of Set points is
determined by the par. "nSP" located in the group of parameters “
]SP “.
The active Set point can be selected :
- by parameter "SPAt" in the group of parameters “ ]SP “.
- by key “U” if par. "USrb" = CHSP
- by digital input if par. “diF” = CHSP or = SP1.2 or =HECo)
- Automatically between SP1 and SP2 if a time “dur.t” (see par.
4.12) has been programmed.
Set Points “SP1”, “SP2”, “SP3”, “SP4” will be visible depending on
the maximum number of Set Points selected on par. “nSP” and they
can be programmed with a value that is between the value
programmed on par. “SPLL” and the one programmed on par.
“SPHL”.
Note : in all the following examples the Set point is indicated as
"SP", however the instrument will act according to the Set point
selected as active.
3 - INFORMATION ON INSTALLATION AND USE
3.1 - PERMITTED USE
The instrument has been projected and
manufactured as a measuring and control
device to be used according to EN61010-1 for
the altitudes operation until 2000 ms.
The use of the instrument for applications not expressly permitted
by the above mentioned rule must adopt all the necessary protective
measures.
The instrument CANNOT be used in dangerous environments
(flammable or explosive) without adequate protection.
The installer must ensure that EMC rules are respected, also after
the instrument installation, if necessary using proper filters.
Whenever a failure or a malfunction of the device may cause
dangerous situations for persons, thing or animals, please
remember that the plant has to be equipped with additional devices
which will guarantee safety.
3.2 – MECHANICAL MOUNTING
The instrument, in DIN case 48 x 48 mm, is designed for flush-in
panel mounting.
Make a hole 45 x 45 mm and insert the instrument, fixing it with the
provided special bracket.
We recommend that the gasket is mounted in order to obtain the
front protection degree as declared. Avoid placing the instrument in
environments with very high humidity levels or dirt that may create
condensation or introduction of conductive substances into the
instrument.
Ensure adequate ventilation to the instrument and avoid installation
in containers that house devices which may overheat or which may
cause the instrument to function at a higher temperature than the
one permitted and declared.
Connect the instrument as far away as possible from sources of
electromagnetic disturbances such as motors, power relays, relays,
solenoid valves, etc.
The instrument can be removed from its housing from the front side
: it is recommended that the instrument be disconnected from the
power supply when it is necessary to carry out this operation.
3.3 - ELECTRICAL CONNECTION
Carry out the electrical wiring by connecting only one wire to each
terminal, according to the following diagram, checking that the
power supply is the same as that indicated on the instrument and
that the load current absorption is no higher than the maximum
electricity current permitted.
As the instrument is built-in equipment with permanent connection
inside housing, it is not equipped with either switches or internal
devices to protect against overload of current: the installation will
include a two-phase circuit-breaker, placed as near as possible to
the instrument, and located in a position that can easily be reached
by the user and marked as instrument disconnecting device which
interrupts the power supply to the equipment.
It is also recommended that all the electrical circuits connected to
the instrument must be protect properly, using devices (ex. fuses)
proportionate to the circulating currents.
It is strongly recommended that cables with proper insulation,
according to the working voltages and temperatures, be used.
Furthermore, the input cable of the probe has to be kept separate
from line voltage wiring. If the input cable of the probe is screened, it
has to be connected to the ground with only one side.
We recommend that a check should be made that the parameters
are those desired and that the application functions correctly before
connecting the outputs to the actuators so as to avoid
malfunctioning that may cause irregularities in the plant that could
cause damage to people, things or animals.