Tech West ACO2S Manual

2625 N. Argyle Ave. • Fresno, CA 93727
(559) 291-1650 • (800) 428-7139 • FAX (559) 348-9677
T
ECH
W
EST
I
NC.
Manufacturers of Dental Vacuum
and Air Systems
ULTRA CLEAN
OILLESS COMPRESSOR
INSTALLATION
AND SERVICE
MANUAL
Revised 4-15

.

1
ULTRA CLEAN
OILLESS COMPRESSOR
INSTALLATION AND SERVICE MANUAL
This manual is for the installation and service of Tech West’s Ultra Clean Oilless Compressors.
CONTENTS
Installation
Location Requirements 2
Figure 1: Wire and Breaker Sizes 2
Ultra Clean Oilless Installation Steps 3
Ultra Clean Oilless Compressor Connections 4
Ultra Clean Oilless Start Up 5
Ultra Clean Oilless Maintenance 6
Ultra Clean Oilless Compressor Motor Assembly Breakdown 7-8
Coalescing Filter Parts Breakdown 9
Desiccant Dryer Parts Breakdown 10-11
Gauge and Moisture Indicator Assembly Breakdown 12
Single Ultra Clean Oilless Compressor Assembly 13
Dual Ultra Clean Oilless Compressor Assembly 14
Triple Ultra Clean Oilless Compressor Assembly 15
TROUBLE SHOOTING
Single Compressor Trouble Shooting 16-17
Dual Compressor Trouble Shooting 18-19
Triple Compressor Trouble Shooting 20-21
Air Line Sizing Chart 22
Maintenance & Service 23-24
Notes Sheet

ULTRA CLEAN OILLESS COMPRESSOR
INSTALLATION
1. OILLESS COMPRESSOR LOCATION REQUIREMENTS
The Ultra Clean Oilless Compressor location should be level, accessible and well ventilated.
If the Ultra Clean Oilless Compressor will be located in a confined space, provide adequate ventilation.
Electrical
(1) Line voltage must be within the limits of Figure 1 below. (Install a “buck-boost
transformer” if line voltage is not between these values.) Provide a separate line for
each motor. Circuit breaker switches must be 20 - 30 amp depending on model
and voltage necessary.
(2) Local code may require you to provide one quick disconnect (safety switch) for each
compressor motor.
(3) See Figure 1 below for breaker size and line voltage.
CAUTION - Voltage must be 208/240 V or motor damage may occur.
CAUTION - Voltage must be 110/130 V or motor damage may occur.
2
Figure 1: Recommended Wire and Breaker Size
Single Head Compressors
Model Voltage Amperage Wire Size Recommended
(Gauge) Breaker Size
ACO2S1 110/115 14.5 12 20
ACO2S2 208/230 7.2 12 20
ACO4D2 208/230 14.4 10 30
ACO6T2 208/230 21.6 10 30
Dual Head Compressors
Triple Head Compressors

ULTRA CLEAN OILLESS COMPRESSOR
INSTALLATION
2
2. INSTALLATION STEPS
This dental compressor should only be installed by qualified personnel. Should any questions arise
during the installation, call Tech West Technical Support between the hours of 7:00 a.m. to 4:00 p.m.
(Pacific Standard Time).
Place the compressor in a clean, dry, well ventilated area, on a solid, level surface. Consider sound
level and insulate as needed. Be sure that adequate ventilation is available as the compressor is air
cooled. Ambient temperature in the equipment room should be within the temperature range of 40
degrees Fahrenheit minimum to 100 degrees Fahrenheit maximum.
(a) Check the shipping carton for damage. This could detect damage to the unit which might
otherwise be overlooked. Remove cardboard shipping carton.
(b) Remove the Oilless Compressor from its shipping skid. Inspect the unit for damage.
Oilless Compressors are shipped bolted to a pallet. This pallet is intended for shipping only
and should be discarded.
(c) Remove installation kit attached to pallet. It should contain the following:
(4) Isolation Feet
(1) Alternate Air Hookup Hose
(1) 5’ Flexible Air Hose
(d) Install isolation feet on tank legs.
(e) Move compressor into place and level by observing bubble level on compressor platform.
(f) Wiring instructions:
(1) Have all electrical connections made by qualified personnel only. All connections
should be in accordance with local codes.
(2) Use the chart on page 1 to help determine the proper line and breaker size for
the unit that is being installed.
(g) Install the air line from the compressor tank to the building supply.
(h) Install the 1” flex alternate air hose from the compressor to a fresh air supply.
3

Electrical Connection
to disconnect and
electrical panel
220 v
(110 v / 220 v ) on the
single models only
Air Out
Connection to building
supply line
4
ULTRA CLEAN OILLESS COMPRESSOR
INSTALLATION
2
3. CONNECTIONS
Alternate Air
Connection
to fresh air supply
Figure 2
Dryer Purge
Connection

5
ULTRA CLEAN OILLESS COMPRESSOR
INSTALLATION
2
4. SAFETY PRECAUTIONS
• Keep fingers, foreign objects and clothing free from rotating parts and do not touch hot surfaces.
• Never attempt to service an operating unit.
• Isolate unit from system pressure and relieve backpressure before servicing.
• Disconnect all power before servicing. The thermal protector in single phase motors
automatically starts motor when device resets.
USE OF THIS PRODUCT IN OR NEAR EXPLOSIVE ATMOSPHERES, OR FOR
PUMPING MIXTURES OTHER THAN ATMOSPHERIC AIR MAY CAUSE AN
EXPLOSION OR FIRE, RESULTING IN PERSONAL INJURY OR DEATH.
2
5. START-UP STEPS
(a) Make sure the shut-off valve from the compressor tank is closed.
(b) Turn the breaker from the panel to the “ON” position.
(c) Turn power “ON” from the toggle switch on the compressor. Compressor should run quietly
and vibration free. The storage tank should start to build pressure.
(d) The compressor will run until the pressure gauge reads 100 psi. The compressor then will
automatically shut off and the dryer will purge with a quick blast of air.
(e) Using soapy water, check the compressor plumbing hook ups for leaks. Repair leaks if needed.
(f) Pressure test the entire plumbing system for leaks. Use the storage tank pressure gauge to
monitor a pressure drop. After the plumbing system has been pressurized for 30 minutes,
re-check the gauge for pressure drop. If there is a drop in pressure, find and repair all leaks
in the office plumbing.
(g) Complete and mail in the warranty card for the compressor within ten days of installation.
AIR LEAKS ARE THE MAIN CAUSE OF COMPRESSOR FAILURES.

6
7. PERIODIC SERVICING
MONTHLY SERVICING
(a) Remove the purge bucket and empty as needed
(b) Crack the drain petcock on the bottom of the storage tank to check for water and drain if needed.
(c) Check moisture indicator to see that it is still “BLUE”. If it is pink, it is time to service the air dryer.
(d) Check pump-up times of the compressor.
(e) Check intake air supply filters. Replace filters as needed.
YEARLY SERVICING
(a) Repeat all of the monthly servicing.
(b) Replace the dessicant air dryer.(Part #: RDC-100)
(c) Replace the alternate air filter. (Part #: TIF-100)
(d) Replace the intake filters on the Oilless compressor head. (Part #: TIF-100))
(e) Replace the moisture indicator on the compressor. (Part #: MI-100)
(f) Check the coalescing filter and change if needed. (Part #: CFEO-375)
ULTRA CLEAN OILLESS COMPRESSOR
INSTALLATION
2
6. GENERAL SERVICE INFORMATION
For parts and service on VX series products contact the nearest authorized Tech West distributor. To
expedite appropriate service, be prepared to provide the unit model number, identification number,
and serial number found on the nameplate located on the front of the unit motor.
Component life operating at continuous duty & maximum pressure
Life of the rings and skirts is difficult to predict due to many conditions which directly influence wear.
Some of these conditions may include ambient air temperature, air cleanliness, operating pressure,
piston stroke on the particular model being utilized, duty cycle, maintenance of filters, etc.
Because of these various factors it is appropriate to generalize on component wear life and choose
some conservative estimates for most standard applications.
With these conditions in mind, we recommend the following preventative maintenance schedule.
RECOMMENDED MAINTENANCE
FOR VX MODELS
Minor Service Kits, Piston Rings & Springs, Skirts, Etc.
Major Replacement Kit, Piston & Rod Assemblies
Replace inlet Filter
HOURS
Cont. Duty
Maximum Pressure
8,000 Hours
12,000 Hours
4,000 Hours
TIME
Based on 100%
Duty Cycle
12 months
18 months
6 months

7
KEY PART NO. DESCRIPTION UNIT
1 TOM-1-115 OILLESS COMPRESSOR MOTOR 1HP 115V 1
2 TOM-1-230 OILLESS COMPRESSOR MOTOR 1HP 230V 1
3 TFF-100 FRONT FAN ASSY. 1 HP 1
4 TFC-100 FRONT FAN COVER 1 HP 1
5 TPA-100 PISTON ASSEMBLY COMPLETE 1 HP 1
6 TCA-100 CYLINDER ASSEMBLY COMPLETE 1 HP 1
7 TAF-100 AIR FILTER ASSEMBLY COMPLETE 1 HP 1
8 TIF-100 AIR FILTER ELEMENT ONLY H HP 1
9 POVT-100 POP OFF SAFETY VALVE 1
OIL-FREE COMPRESSOR MOTOR 1HP
See next page for numbered part identification
ULTRA CLEAN OILLESS COMPRESSOR
PARTS LIST

8
8
7
69
1
2
3
4
5
ULTRA CLEAN OILLESS COMPRESSOR
PARTS LIST
Este manual sirve para los siguientes modelos
3
Tabla de contenidos
Otros manuales de Compresor de aire de Tech West