Ambient conditions
0 0
Functional temperature range -20 C... + 50 C
0 0
Storage temperature range -25 C... + 70 C
Climatic class 2z/-20/50/70/75 % with reference to
VDI / VDE 3540
Altitude up to 2000 m
Mechanical configuration
Protection type IP 50 for the connection sockets IP 20
according to DIN VDE 0470 Part 1 / EN
60529
Dimensions 84 mm x 195 mm x 35 mm
Weight 350 g approx., including battery
18. Maintenance
Caution :
Disconnect the meter from the measuring circuit before you open
it to replace the battery or the fuse !
18.1 Battery
Prior to initial start-up, or after storage of your meter, verify that the battery of
your meter does not leak. Repeat this check in regular short intervals. If the
battery leaks, completely remove the battery electrolyte carefully with a moist
cloth and install a new battery before you operate your meter again.
When the symbol “ ” (17) appears on the LCD (1), replace the battery as
soon as possible. You can continue to measure, but a reduced measuring
accuracy must be taken into account.
The meter operates with a 9 V flat cell battery according to IEC 6F 22 or IEC 6
LR 61 or with a suitable NiCd storage battery.
Replacing the battery
êplace the meter on its face, loosen the two screws on the rear and
remove the lower part of the case, lifting it from the bottom. The lower and
the upper part of the case are fixed together at the top on the front by
means of detent hooks.
êRemove the battery from the battery compartment and carefully
disconnect the contacts from the battery.
êSnap the connection contacts to a new 9V battery and insert the battery
into the battery compartment.
êReplace the lower part of the case. Start at the top on the front and take
care that the detent hooks are properly engaged at this point.
êTighten the lower part with the two screws.
êPlease destroy the batteries in an environmental friendly way.
18.2 Fuses
A blown fuse is signaled on the LCD display the instant a measured quantity
having a voltage of more than 4V is applied to the corresponding connection
sockets. Then, the digital display (9) shows “FUSE”. The 16 A fuse interrupts
the 3 A and 10 A ranges, the 1.6 A fuse all other current measuring ranges.
All other measuring ranges continues to function. When a fuse blows, first
eliminate the cause of the overload before using the meter again!
Meaning of the symbols on the device
Warning of a danger point
(Attention, refer to documentation)
Ground connector
Double or reinforced - insulation
CAT II/III/IV Instrument for a overvoltage catagory
II/III or IV
EU conformity mark
For US & Canada
Repair, replacement of parts and calibration
When opening the meter, live parts may be exposed. Therefore, the meter
must be disconnected from the measuring circuit prior to opening its case
for repair, replacement of parts or calibration. If repair or calibration cannot
be avoided unless the meter is open and live, this work must only be
performed by a qualified person who understands the danger involved.
Faults and extraordinary stress
When it must be assumed that safe operation is no longer possible, take
the meter out of service and secure it against accidental use. It is assumed
that safe operation is no longer possible.
—when the meter shows obvious signs of damage,
—When the meter no longer functions correctly,
—after prolonged storage under adverse conditions,
—Due to sever stress during transportation.
2. Switching the meter on Battery
Fit the meter with 9 volt flat cell battery provided along with the meter.
Before you use the meter for the first time or after storage,
absolutely refer to Section “18.1 Maintenance Battery”.
Switching the meter on
íPress the “ON/OFF” pushbutton (2).
Switch-on is acknowledged by a sound signal. As long as you keep the
push-button pressed, all segments of the liquid crystal display (LCD) will
appear. The LCD is shown on page 2.
After the pushbutton is released, the meter is ready for operation.
Note :
Electric discharge and high frequency interference may cause incorrect
information to be displayed and block the measuring process. Reset the
meter by switching it OFF and ON again. Otherwise, check the battery
connections.
Disconnect the meter from the measuring circuit before you open
it and see section “18. Maintenance”!
Automatic turn-off
Your meter turns off automatically, when the measured value remains
constant (variations of the measured value < ± 20 digits) for about 10
minutes and when neither a pushbutton nor the function selector switch is
operated during that time.
How to prevent automatic tun-off
Switch your meter to “CONTINUOUSLY ON” mode.
íTo do this, press the yellow multi-function pushbutton (5) and the
“ON/OFF” pushbutton (2) together. The function “CONTINUOUSLY ON”
is shown on the LCD (1) by the symbol (8).