Refine MaxCure5 Manual de usuario

Please read this manual before operating
RF-RCL-M014 Edition: V2.0 Modify: 20210529

content
1 Product introduction ……………………………………………………………………………………4
2 Model and technical parameters ……………………………………………………………………6
3 Installation and demounting …………………………………………………………………………8
4 Operation ………………………………………………………………………………………………9
5 Cleaning, disinfecting and sterilizing……………………………………………………………… 10
6 Transportation, storage and maintenance ……………………………………………………… 11
7 Troubleshooting …………………………………………………………………………………… 12
8 Environmental protection ………………………………………………………………………… 14
9 Statement …………………………………………………………………………………………… 14
10 EMC-Declaration of conformity…………………………………………………………………… 14
Attachment. Reprocessing instructions of cleaning, disinfecting and sterilizing ……………… 20

1
Safety Precautions
WARNING: If you neglect these safety precautions, you may cause personal injury such as electric
shock, re or damage to the product.
1. Use a separate, grounded power outlet. Never use wet hands to unplug the power cord.
2. Please do not use other than the specied voltage to recharge the battery.
3. Please recharge the battery at least 4 hours before rst time usage.
4. It is forbidden to touch the charging connector with metal or other conductor, to avoiding damage the
circuit of charge or the battery.
5. It is forbidden to extrude, shake or rock the battery. The Lithium battery is forbidden to be in short-circuit
situation and it is forbidden to put the battery with metal or other conductors.
6. Please recharge the battery in cool and ventilated room.
7. It is forbidden to self-taking-apart the battery, in order not to result in short-circuit or leakage.
8. Keep the Curing Light clean before and after operation. The optical ber must be sterilized before each
treatment.
9. During the operation, the light should be aimed straightly at the composite resin to ensure the eect of
solidication.
WARNING: Avoid aiming at eyes directly.
10. Don't knock or rub the Curing Light.
11. This product is intended for use in hospitals and dental clinics only. The user must be professionally
trained and qualied dentists.
12. Only the original pedestal charger, adapter and Lithium battery could be used, because other brand
pedestal charger, adapter and Lithium battery are likely to damage the circuit.
13. If you don't use this equipment for a long time, please take the battery out and preserve separately.

2
14. As a professional manufacturer of medical instruments, we are only responsible for the safety on the
following conditions:
• The maintenance, repair and modication are made by the manufacturer or the authorized dealer.
• The changed components are original of our company and operated correctly according to instruction
manual.
WARNING: No modication of this equipment is allowed.
WARNING: If this equipment is modied, appropriate inspection and testing must be conducted to
ensure continued safe use of equipment.
WARNING: If the curing light works for 40s continously,the temperature of the top of optical ber may
reach 56 °C.
WARNING: Do not modify this equipment without authorization of the manufacturer.
Contraindication :
The heart disease patient, pregnant woman, children and the person who are allergic to blue light should be
cautious to use this equipment.

3
Symbol instruction
Symbol Instruction Symbol Instruction
Warning, Caution and Important! Check
the Instruction Manual Consult the accompanying documents
Date of manufacture Manufacturer
According to the type of protection
against electric shock: CLASS II
EQUIPMENT
According to the degree of protection
against electric shock:Type B applied
part
Screw inside/ outside Used indoor only
Recovery Keep dry
Handle with care Atmospheric pressure for storage
Temperature limitation for storage Humidity limitation for storage
Appliance compliance WEEE directive,
Dispose as required by the law. CE Mark
Authorised Representative in the EUROPEAN COMMUNITY

4
1 Product introduction
The product suppose to be used in hospital and dental clinic, should be used by trained qualied dentist.
This Curing Light is used for the principle of ray radiation to solidify the light-sensitive resin by shooting at it
in a short time, it is used to restore teeth and solidify material for whitening teeth.
Figure 1-A MaxCure5 Figure 1-B MaxCure9

5
Figure 1-C MaxCureG Figure 1-D MaxCureH Figure 2 Adapter
Note: The button labeled 'Time' indicates the time setting, and the button labeled 'Mode' indicates
the mode setting.
Safety classication Working condition
1. Protection type against electrical shock: Class II
2. Protection degree against electrical shock: Type B
3. Protection against harmful ingress of water or particular matter:
ordinary equipment (IPX0), can’t be waterproof.
4. operation mode: short time run equipment.
5. Safety in the presence of ammable anesthetic mixture with air,
oxygen or nitrous oxide: not suitable under this condition.
a) Environment temperature:
5°C-+40°C
b) Relative humidity: 30%-75%
c) Atmosphere pressure:
70kPa-106kPa

6
2 Model and technical parameters
Table 1: Technical Parameters of table-top models
Parameters MaxCure5 MaxCure9
Power supply
(Rechargeable Lithium
battery)
Battery model: ICR18650
Battery voltage and capacity:
3.6V/2600mAh
Battery model: ICR18650
Battery voltage and capacity: 3.6V/2600mAh
Power supply
(Adapter)
Input : 100-240VAC, 50Hz/60Hz
Output: 5VDC/1A
Input : 100-240VAC, 50Hz/60Hz
Output: 5VDC/1A
Applied part optical ber Light guide
Light Intensity 1000mW/cm2- 1800mW/cm21000mW/cm2- 2500mW/cm2
Modes setting
TURBO mode: Display P1,
Illumination 1600mW/cm2-
1800mW/cm2
NORMAL mode: Display P2,
Illumination 1000mW/cm2-
1200mW/cm2
TURBO mode: Display P1, Light Intensity
2300mW/cm2- 2500mW/cm2
NORMAL mode: Display P2, Illumination
1000mW/cm2- 1200mW/cm2
Time setting
TURBO mode: 3S, 5S
NORMAL mode: 5S, 10S, 15S,
20S
Lightly press the time button to
choose the solidication time.
TURBO mode: 1S, 3S
NORMAL mode: 5S, 10S, 15S, 20S
Lightly press the time button to choose the
solidication time.
Dimensions Φ25mm*250mm Φ26mm×255mm

7
Parameters MaxCure5 MaxCure9
Net weight 152g 144g
Light source a) 5W high power blue LED
b) Wave length: 420nm-480nm
a) 5W high power blue LED
b) Wave length: 385nm-515nm
Consumption power ≤5W ≤5W
Composed mainly
main unit, charger, optical ber,
light hood, LED lamp, battery and
power adapter.
main unit, charger, light hood, LED lamp,
battery and power adapter.
Table 2: Technical Parameters of built-in models
Parameters MaxCureG MaxCureH
Power supply Input : 24VAC, 50Hz/60Hz Input : 24VAC, 50Hz/60Hz
Light Intensity 1000mW/cm2 - 2500mW/cm21000mW/cm2 - 1800mW/cm2
Applied part Light guide Optical ber
Modes setting
TURBO mode: Display P1, Light Intensity
2300mW/cm2 - 2500mW/cm2
NORMAL mode: Display P2, Illumination
1000mW/cm2 - 1200mW/cm2
TURBO mode: Display P1, Illumination
1600mW/cm2 - 1800mW/cm2
NORMAL mode: Display P2, Illumination
1000mW/cm2 - 1200mW/cm2

8
Parameters MaxCureG MaxCureH
Time setting
TURBO mode: 1S, 3S
NORMAL mode: 5S, 10S, 15S, 20S
Lightly press the time button to choose the
solidication time.
TURBO mode: 3S, 5S
NORMAL mode: 5S, 10S, 15S, 20S
Lightly press the time button to choose the
solidication time.
Dimensions Φ25mm*275mm Φ25mm*265mm
Net weight 96g 159g
Light source a) 5W high power blue LED
b) Wave length: 385nm-515nm
a) 5W high power blue LED
b) Wave length: 420nm-480nm
Consumption
power ≤5W ≤5W
Composed mainly main unit, LED lamp, light hood, cable. main unit, optical fiber, light hood, LED
lamp, cable.
3 Installation and demounting
3.1 Installation of the table-top models
a) Take o the red cap from the optical ber and then insert the metal part into the front of Curing Light,
make sure to screw the ber to the end. To install the light hood on the top of the optical ber. (The models
with optical ber)
b) To install the light hood on the top of the optical ber.
c) For dismantlement, taking the instruction of istallment above reversely.
d) When finishing using or when the battery needs to be charged, connect the plug of the adapter into
Este manual sirve para los siguientes modelos
3
Tabla de contenidos
Otros manuales de Equipo dental de Refine
Manuales populares de Equipo dental de otras marcas

Vatech
Vatech EzRay Air VEX-P300 Manual de usuario

KaVo
KaVo GENTLEpower LUX Contra-angle 25 LP Manual de usuario

DENTSPLY
DENTSPLY SmartLite Focus Manual de usuario

LM
LM ProPower CombiLED Manual de usuario

Owandy Radiology
Owandy Radiology RX-AC Manual de usuario

mectron
mectron Piezosurgery Manual de usuario