If the device is calibrated, You can cancel the “ZERO” function by pressing the (OK) button.
WARNING! While zeroing, the calibration sheet should be placed on a flat, stable non-metalic surface (e.g.
do not place the sheet on the bodywork). Also, please avoid holding the sheet in Your hands as this may affect
the operation.
Please note that zeroing is worth to be made when the measurement conditions are changing (e.g. the
temperature or humidity rises or falls).
ATTENTION!
1. Use only alkaline batteries as normal batteries provide unsufficient power.
2. Instead of a alkaline battery You can use a compatible, fully charged accu.
3. If the device seems to work improperly it can be caused by an empty battery or accu.
To take a measurement please put the device’s sensor against the examined surface. The sensor should
adhere as flat as possible. The examined surface should be clean and smooth, otherwise the measurement may be
incorrect. It is advised to hold the gauge with both hands and press gently towards the examined surface to
eliminate vibration. Usually it takes 1-2 seconds to stabilize the measurement result.
3. Gauge manual
The device is operated with 2 buttons:
- (OK / MENU): turns the device on, accepts the selected functions and returns to the main menu, when
pressed during the measurement it memorizes the displayed value;
- (FUNCTION): switches between the functions in the main menu, switches between the stored
measurement values.
To turn the gauge on please press the (OK) button for about 1 second. After displaying the company’s logo and
the device’s type, the device switches automatically to the main menu. In the first line You will see a
„>MEASUR” caption and in the second „SUBST” and the battery monitor. The „>” symbol indicates the
currently selected function. Please press the (FUNCTION) button to switch between the following functions:
-MEASUR –measure function: takes measurements;
-SUBST –substance function: allows to choose the measured material;
-MEMORY –memory function: recalls the stored measurement values;
-REF PT –reference point function: sets the reference point;
-OFF –off function: switches the device off;
-DELETE –delete function: deletes the stores measurement values;
-ZERO –zeroing function: calibrates the gauge;
-ASSIST - Enables / disables the measuring assistant.
To select a function please press the (OK) button.
Measure function
After You select this function the display will show the currently selected material, e.g. “Zn/Al” and the
battery indicator in the first line and „----um” in the second. The gauge is ready to take measurement.
After You put the sensor against the bodywork the result of the measurement will be shown on the display in the
second line.
To store the result please press the (OK) button when taking the measurement. A „*” symbol appears next to the
shown value. Attention! The capacity of the device’s memory is 100 results. After that the latest stored values
are deleting the oldest ones.