CENTERONE/CENTERTWO/CENTERTHREE
Controller, Contrôleur, Centralina, Unidad de control, Bedieningsapparaat,
Styrenhet, 控制器 , 제어 장치

2
Installation
1 Installation
1.1 Range of application
The CenterOne, CenterTwo or CenterThree devices are to be installed and operated in
the following ambient conditions:
1.2 Installation, Setup
The unit can be operated in three different ways:
– mounted in a 19" rack cabinet
– mounted in an operator panel
– as a desktop unit
Rack mounting The unit is designed for installation into a 19" rack chassis adapter according to DIN 41
494. For this purpose, four collar screws and plastic sleeves are supplied with it.
Installation location sheltered (interior)
Degree of protection IP 30
Installation altitude max. 2000 m
Ambient temperature +5°C to +50°C
Relative humidity ≤80% up to +31°C, decreasing to 50% at +40°C
Air pressure 860 hPa - 1060 hPa
DANGER
Commissioning of a defective product
Putting a product which presents a visible damage into operation can be extremely haz-
ardous.
If the product presents a visible damage do not put it into operation and make sure it
is not inadvertently put into operation.
NOTICE
Loss of the protection class of the rack
If the product is installed in a rack, it is likely to lower the protection class of the rack
(protection against foreign bodies and water) e.g. according to the EN 60204-1 regula-
tions for switching cabinets.
Take appropriate measures for the rack to meet the specifications of the protection
class.
NOTICE
Thermal overload!
Set up the unit so that the admissible maximum ambient temperature (e. g. due to
installation or sunlight) is not exceeded.
Ensure adequate air circulation!

Installation
3
Guide rail
Fig. 1: Guide rail on the rack module adapter
It is advisable to fit a guide rail to the rack module adapter to ease the pressure on the
front panel of the device.
Mounting height of the rack module adapter
Ascertain the mounting height of the device on the basis of the following diagram.
Fig. 2: Mounting height (1 or 2)
Rack mounting at mounting height 1 (only applies to CenterOne)
Secure the rack module adapter in place in the switchgear cabinet
Slide the device into the rack module adapter.
Secure the device in place using the screws supplied.
?
12
NOTICE
Thermal overload!
When setting up the device, ensure that the maximum permissible ambient temper-
ature (due to installation or solar radiation, for instance) is not exceeded.
Ensure that there is good air circulation!

4
Installation
Rack mounting at mounting height 2 (applies to CenterOne/CenterTwo/Cen-
terThree)
Secure the rack module adapter in place in the switchgear cabinet
only for CenterOne: Fasten the adapter plate in place as an upper extension of the
device front panel using the screws supplied with the adapter plate.
Slide the device into the rack module adapter.
Secure the device in place using the screws supplied with the device.
Fig. 3: CenterOne: Insert the adapter plate, slide in the device and fasten it in place
Fig. 4: CenterTwo / CenterThree: Slide in the device and fasten it in place
NOTICE
Thermal overload!
When setting up the device, ensure that the maximum permissible ambient temper-
ature (due to installation or solar radiation, for instance) is not exceeded.
Ensure that there is good air circulation!
NOTICE
Accessories for mounting the CenterOne
An adapter plate inclusive of two neck screws and a plastic nipple are available for the
installation of the CenterOne(see accessories).

Installation
5
Installation in a Con-
trol Panel
The following operator panel cutout is required for mounting in an operator panel:
Fig. 5: Operator panel cutout and mounting CenterOne
Fig. 6: Operator panel cutout and mounting CenterTwo / CenterThree
Guide the device into the operator panel cutout.
– It is advisable to support the device in order to ease the load on the front panel.
Fasten the device in place with four M3 or comparable screws.
Desk-Top Unit The CenterOne/CenterTwo/CenterThree is also suited for use as desk-top unit. For this
purpose, two self-adhesive rubber feet as well as a slip-on rubber bar are supplied with it.
NOTICE
Loss of the protection class of the rack
If the product is installed in a rack, it is likely to lower the protection class of the rack
(protection against foreign bodies and water) e.g. according to the EN 60204-1 regula-
tions for switching cabinets.
Take appropriate measures for the rack to meet the specifications of the protection
class.
NOTICE
Thermal overload!
Set up the unit so that the admissible maximum ambient temperature (e. g. due to
installation or sunlight) is not exceeded.
Ensure adequate air circulation!
105
91.4
M3 or Ø3.5
78
69
105
91.4
M3 or Ø3.5
122.5
112

6
Installation
Stick the two supplied rubber feet to the rear part of the bottom plate.
Slip the supplied rubber bar onto the bottom edge of the front panel.
1.3 Connecting to the mains power supply
The CenterOne/CenterTwo/CenterThree is equipped with an universal power supply
with an input voltage range of 100 ... 240 V AC.
NOTICE
Thermal overload!
Set up the unit so that the admissible maximum ambient temperature (e. g. due to
installation or sunlight) is not exceeded.
Ensure adequate air circulation!
CAUTION
Excess voltage!
Danger of destroying the unit.
Power connections must comply with local regulations. Voltage information given on
the rating plate must correspond to the mains voltage values.
Use only cable with protective earth conductor.
To protect the electronics and the power supply cable in case of failure provide an
appropriate network safety device.
WARNING
Danger of electric shock
In case of defect, the parts connected to the mains supply are under voltage.
Always keep the mains connection freely accessible so you can disconnect it at any
time.

Installation
7
Fig. 7: Mains cable with protective earth conductor
Switch off switch S1 on the power pack (position "0").
Connect the mains cable to the mains.
Always ensure a safe connection to the protective earthing conductor (PE, protection
class I).
To protect the unit in case of malfunction, carry out fuse protection in accordance with
the regional regulations.
– Select a fuse with slow characteristics.
If the unit is installed in a rack, the mains voltage must be supplied via a switched power
distributor.
Earthing On the rear side there is a screw to connect the CenterOne/CenterTwo/CenterThree via
a protective ground if necessary, for example, with the protective earthing of the pumping
station.
Fig. 8: Protective ground
1.4 Connections
Transmitter terminals
CH 1, CH 2 and CH3 There are two terminals wired in parallel available for each measuring channel.
●an 8-pin RJ45 socket (CH A)
●a 15-pin sub-D socket (CH B)
Connect the transmitter using a preassembled measuring cable or with a self-manu-
factured, shielded cable (electromagnetic compatibility) to terminals CH 1, CH 2 or CH
3 on the back of the device.
S1
10 A
max
X: Do not unfasten this screw
(internal ground protection)
X

8
Installation
Pay attention to the list of usable measuring tubes!
Fig. 9: Only one transmitter per measuring channel
Pin assignment
CH 1, CH 2 and CH 3 RJ45 socket
The pins on the 8-pin RJ45 socket (FCC68) are assigned as follows:
Fig. 10: RJ45 socket
Sub-D socket
The pins on the 15-pin sub-D socket are assigned as follows:
DANGER
Dangerous contact voltage
Voltages above 30 V AC or 60 V DC are deemed to be dangerous in the event of contact
in accordance with EN 61010.
Only apply a grounded protective extra-low voltage (PELV).
NOTICE
Multiple assignment
Possibility of damage if transmitters are connected simultaneously.
Only connect one transmitter to each measuring channel (terminal CH A or CH B).
i
Pin Signal
1 Infeed (+24 V DC)
2 Infeed ground (GND)
3 Signal input (measuring signal 0 ... +10 V DC)
4 Identification
5 Signal ground
6Status
7HV_L
8HV_H/HV_EMI
Pin Signal
1 EMI status
2 Signal input (measuring signal 0 ... +10 V DC)

Installation
9
Fig. 11: Sub-D socket
CONTROL terminal
(CenterOne) This terminal can be used to:
●read out the measuring signal,
●perform an isolated evaluation of the state of the switching function and error monitor-
ing,
●switch the high-vacuum measuring circuit on and off (only with the cold-cathode trans-
mitter PTR 225).
Connect peripheral components with a shielded connecting cable (electromagnetic
compatibility) to the CONTROL terminal on the back of the device.
Pin assignment and contact positions
The pins on the 15-pin sub-D connector are assigned as follows:
3Status
4HV_H/HV_EMI
5 Infeed ground (GND)
6n.c.
7 Degas
8 Infeed (+24 V DC)
9n.c.
10 Identification
11 Infeed (+24 V DC)
12 Signal ground
13 RxD
14 TxD
15 Housing
Pin Signal
DANGER
Dangerous contact voltage
Voltages above 30 V AC or 60 V DC are deemed to be dangerous in the event of contact
in accordance with EN 61010.
Only apply a grounded protective extra-low voltage (PELV).
Pin Signal
1 Analog output -5 ... +13 V DC 2)
2 Analog output GND
Switch function 1
3 Pressure higher than
threshold or device
switched off
Pressure lower than
threshold
4
5
6 HV_V on (+ 24 V)
HV_H off (0 V)
7 +24 V DC, 200 mA 1)
8 GND = housing 1)

10
Installation
1) 300 mA fuse protection with a PTC element, self-resetting after the CenterOne is switched off or
the CONTROL connector is removed. Meets the requirements for grounded protective extra-low
voltage.
2) The variation between the analog output (pin 1) and the display on the device is a maximum of +/-
20 mV.
Fig. 12: Sub-D connector
CONTROL terminal
(CenterTwo / Center-
Three)
This terminal contains:
●analog outputs for the signals from the individual measuring channels,
●logger output. This is a programmable analog output which can be assigned to any
measuring channel.
●HV-EMI. The high-vacuum measuring circuit for the PTR 225 transmitter can be
switched on and off here (On = +24 V / Off = 0 V).
Connect peripheral components with a shielded connecting cable (electromagnetic
compatibility) to the CONTROL terminal on the back of the device.
Connector pin arrangement
The pins on the 9-pin sub-D connector are assigned as follows:
1) The variation between the analog outputs (pins 1, 2, 6) and the display on the device is a maximum
of +/- 20 mV.
Error signal
9 No error Error or device
switched off
10
11
Switch function 2
12 Pressure higher than
threshold or device
switched off
Pressure lower than
threshold
13
14
15 GND = housing
Pin Signal
DANGER
Dangerous contact voltage
Voltages above 30 V AC or 60 V DC are deemed to be dangerous in the event of contact
in accordance with EN 61010.
Only apply a grounded protective extra-low voltage (PELV).
Pin Signal
1 Analog output 1 (-5 ... +13 V DC) 1)
2 Analog output 3 (-5 ... +13 V DC) 1)
3 Shield GND
4 HV_EMI 3
5 HV_EMI 1
6 Analog output 2 (-5 ... +13 V DC) 1)
7 Logger output (0 ... +10 V DC)
8 Shield GND
9 HV_EMI 2
Este manual sirve para los siguientes modelos
3
Tabla de contenidos
Idiomas:
Otros manuales de Controladores de Pfeiffer Vacuum

Pfeiffer Vacuum
Pfeiffer Vacuum OPS 100 Manual de usuario

Pfeiffer Vacuum
Pfeiffer Vacuum MaxiGauge TPG 256 A PT G28 760 Manual de usuario

Pfeiffer Vacuum
Pfeiffer Vacuum IMG 400 Manual de usuario

Pfeiffer Vacuum
Pfeiffer Vacuum DPG 202 Manual de usuario

Pfeiffer Vacuum
Pfeiffer Vacuum DPG 109 Manual de usuario

Pfeiffer Vacuum
Pfeiffer Vacuum TCP 120 Manual de usuario

Pfeiffer Vacuum
Pfeiffer Vacuum Tele TC cable 110 Manual de usuario

Pfeiffer Vacuum
Pfeiffer Vacuum TCM 1601 Manual de usuario

Pfeiffer Vacuum
Pfeiffer Vacuum TC 400 PB Manual de usuario

Pfeiffer Vacuum
Pfeiffer Vacuum TPS 110 Manual de usuario