Optima OP-W Series Manual de usuario

OP-W Series
User Manual

Warnings
•Avoid using the balance in any harsh environment.
•Only trained professionals are allowed to do repair work.
•Turn power off before connecting/disconnecting the power source.
•Avoid overloading or dropping material onto the platform as that could damage
the balance.
•Material that has a static charge could influence the performance of the balance.
•Electromagnetic interference can affect the reading of the balance.
•Unplug the power before cleaning and use a piece of soft cloth (do not use any
aggressive cleaning agent); Make sure no liquid enters the balance.

............................................................................................................Getting Started!4
...........................................................................................................Overview!4
............................................................................................................Contents!5
........................................................................................................Preparation!5
................................................................................................................Scale Settings!6
....................................................................................................Settings Menu!6
........................................................................................................Units Menu!8
.............................................................................................Si.CAL-01 Menu!11
....................................................................................................CALIB Menu!15
...................................................................................................FILTER Menu!17
.....................................................................................................OIML Menu!17
.....................................................................................................QUIT Menu!20
....................................................................................................RESET Menu!20
....................................................................................................C1/C2 Menu!21
......................................................................................................ADC Menu!21
........................................................................................................RST Menu!22
....................................................................................................ZERO Menu!22
...................................................................................................G1SET Menu!23
...................................................................................................G2SET Menu!24
....................................................................................................................More Info!25
..............................................................................................Troubleshooting!25
...........................................................................................Technical Support!25
!OP-W Series
!User Manual / Table of Contents
!3/25

Getting Started
Overview
Thank you for using Optima scales for your important application. Here is the list of
models this manual covers:
Model #
Max Weight (grams)
Accuracy (grams)
OPW-P3
3,000
0.2
OPW-P6
6,000
0.5
OPW-P15
15,000
1
OPW-P30
30,000
2
!OP-W Series
!User Manual / Getting Started
!4/25

Contents
•Scale main unit
•Scale pan
•User Manual
•AC adapter : 12V
Preparation
1. Carefully remove contents from the package which may include fragile parts.
2. Carefully assemble the scale by placing the scale pan.
3. Place the scale in a firm and flat surface, and avoid location where the scale
maybe subject to vibration, sunlight, airflow, contact with water, and strong
electromagnetic wave.
4. Level the scale by using the bubble level indicator (if available) and adjusting the
feet of the scale.
5. Locate scale in the following condition:
1. Temperature : 0oC ~ 40oC
6. Plug in the scale to a surge protected power outlet if using the AC adapter. You
may use the internal rechargeable batteries also once it has charged.
7. Press the ON/OFF key in the front panel to turn the scale on and wait for it to
finish initialization (i.e. when the display shows a weight of 0).
8. Let the scale warm up for 1 minute before calibration and use.
9. Make sure any weight placed on the scale is not beyond the rated maximum.
10.If scale is not used for an extended period of time, turn off the scale and unplug
the AC adapter (if it’s connected).
!OP-W Series
!User Manual / Getting Started
!5/25

Scale Settings
Settings Menu
After the scale has been turned on and in the normal weighing mode, to get to the
settings menu, follow the procedures below:
•Press the “PLU” key, holding it for more than 2 seconds.
•Press “ENTER” key
!OP-W Series
!User Manual / Scale Settings
1
!6/25

•Enter the password “505388” by using the up/down arrow keys to increase/
decrease the number, and the left/right keys to traverse to the digits of interest.
•Press “Enter” key, and now you are ready to change the settings of the scale. Units
will be the first configuration the user can change.
•For each settings, press “Enter” key to change the value of that particular
parameters by using the up/down keys to increase/decrease the parameters.
•Press “Enter” key to confirm.
•Use the left/right arrows to traverse among the menu settings.
•To exit without saving settings, press the “U” key (esc).
•To exit with saving settings, select “Quit” from the menu and press the “Enter” key
as shown below:
!OP-W Series
!User Manual / Scale Settings
1
!7/25

•Press “Enter” key after “Quit” (as shown above), and the indicator will go back to
normal weighing mode.
Units Menu
Select the weighing unit, the place of the decimal, the maximum capcacity, and the
first rang division and the second rang division, if it exists.
Select the menu “Units”
!OP-W Series
!User Manual / Scale Settings
1
!8/25

Press “Enter” key
Change the parameter to 0 (kg) or 1 (g).
Press “Enter” key.
Select 0 to 5 to choose the number of decimals after the decimal point
Press “Enter” key
On the following screen, enter the maximum capacity of the scale. For the example
below a 5kg max is set. Note the decimal places that was set in the previous step.
Press “Enter” key.
!OP-W Series
!User Manual / Scale Settings
1
!9/25

Press “Enter” key.
The screen above lets user enter the division of the scale. The options are 01, 02,
05, 10, 20, 50. If we follow the previous example, if we choose 01 we will obtain the
division/accuracy of 0.001 kg.
Press “Enter” key to configure the second rang or to exit from the “Units” menu. If
the unit does not have the second rang option, the message will not show (the
configuration of the second rang is the same as the configuration of the first rang).
To exit with saving, select the menu “Quit”, and press the “Enter” key.
To continue the rest of the scale configurations, use the left/right arrow keys to
select the setting of interest and press the “Enter” key.
!OP-W Series
!User Manual / Scale Settings
1
!10/25
Tabla de contenidos
Otros manuales de Escala de Optima