SAFETY PRECAUTIONS
For safe operation of the weighing indicator, please follow these
instructions:
●Calibration inspection and maintenance of the indicator are prohibited by non-pro-
fessional staff
●Please ensure that the indicator rests on a stable surface
●The indicator is a piece of static sensitive equipment; Please cut off power during
electrical connections
●Touching the internal components by hand is prohibited
●DO NOT exceed the rated load limit of the unit
●DO NOT step on the unit
●DO NOT jump on the scale
●DO NOT use this product if any of the components are cracked
●DO NOT use for purposes other then weight taking
●To avoid damaging the battery do not keep charger plugged in once battery is fully
charged
●Make sure the weight is not over the Max capacity as it could damage the load cell
inside
● Material that has a static electric charge could inuence the weighing. Discharge
the static electricity of the samples, if possible. Another
solution to the problem is to wipe both sides of the pan and the top of the case with
an anti-static agent
Please take anti-static prevention measures
Any accumulated charge on the body of the human operator should be discharged rst
before opening the protective container with ESDS devices inside. The discharge can
be accomplished by:
● Putting a hand on a grounded surface or, ideally, by wearing a grounded Anti-static
Wrist Strap and an Anti-static Mat
PREPARATION & SET UP
●Plug into a wall outlet to avoid interference with other wirings
●Turn on the balance while there is no load
●We suggest to warm-up the balance by powering on 5 minutes before use for
accurate weighing
●Calibration may be required before weighing when the balance is initially installed
or moved from a location

2
FEATURES
●Smooth indicator shape with steady performance PCB board, is suitable for
industrial or commercial use
●Multiple weighing units: (lb/oz/kg)
●Gross/Tare/Zero
●Low power design, longer battery usage time
●Hold function, locks the weight
●Idle mode and automatic power off function
●Low-battery detection and warning
● 6 digits 0.7” tall LCD display with backlight
●Splash proof keyboard and display
●RS232 Output
●Print to an optima scale ticket printer
●Stainless steel wall mount bracket included
TECHNICAL PARAMETERS
● Accuracy class: OIML III, n=3000e
● Tare range: 2%~100% max. capacity
● Initial zero range: ±10% max. capacity
● Manual zero range: ±2% max. capacity
●Resolution - Display: Up to 1/15,000 Display Resolution
●Operation temperature: -10 °C ~ +40 °C
● Operation humidity: ≤90%RH
● Storage temperature: -40 °C ~ +70 °C (32-104°F)
POWER SUPPLY
● The indicator uses 3 AAA batteries which will work for about 30 hours with the
backlight turned on, if you turn off the backlight the battery can last up to 100
hours
●This low battery icon will display when the indicator detects low battery, if not
replaced in time the indicator will power off automatically
●If you are using the included DC power supply, the indicator will not use battery
power
● Plug the adapter directly into the “DC” pin located at the back of the indicator. We
recommend to plug into a wall outlet to avoid interference with other wirings

3
DISPLAY AND KEY DESCRIPTION
KEY Weighing Mode Parameter Setting
ON/OFF Powers the Indicator On or Off if held for 2 seconds Exit
HOLD 1. Peak hold - Grabs the highest weight
2. Data hold - Holds the current weight value
3. Auto hold
Conrm
ZERO Zero’s the scale
Tare’s the scale
Move Left
UNITS Shifts between weighing units (kg/lb) Move Down
DISPLAY Meaning
The scale is at zero
The scale is stable
Gross Shows you are in Gross weight mode; default mode
Net Shows you are in Net weight mode (without tared weight)
hold Shows you are in Hold mode
lb The weight is shown in pounds
oz The weight is shown in ounces
kg The weight is shown in kilograms
Low battery

4
OPERATING INSTRUCTIONS
Powering On or Off
●Press ON/OFF key for 1 second to power on. Machine will self inspect and count
from 1 - 9 before entering weighing mode and displaying 0
● Press ON/OFF key for 1 second to power off
●Indicator will auto power off when it has low battery
● Auto power-off in 10/20/60 minutes with no operation can be set in the parameters
Zero/Tare
●The zero function is used only when the scale is empty and is not at gross zero due
to material build up
● If the weight on the scale is within the zero range (2% of the max. capacity) and
stable and ZERO is pressed, the indicator will zero the scale and stay in gross
weighing mode
●The Tare function is used when you only wish to see the current change in weight,
not the entire amount of weight that is on the scale
● If the scale is out of the zero range (above 2% of the max. capacity) and
stable the indicator will Tare the scale and go into Net weighing mode when ZERO
is pressed
Print
●To print press and hold the UNITS key when weighing is stable
●Communication Setting must be set to #2 for Print mode

5
Hold
There are ve different hold functions, you can choose the best option for you and set it
the parameter settings
1. Peak Hold: Grabs the highest weight (for materials testing, ie. tension and pulling
force)
●Press the HOLD key then add weight to the scale
●The indicator will show the highest weight it recorded and hold it on the screen until
a higher weight is placed on the scale
2. Manual Hold: Grabs the current weight and holds it so it will not change/uctuate
●While weighing, press HOLD and the indicator will hold the current weight on the
screen until HOLD is pressed again
3. Auto Hold: If the weight on the scale is above 20d and is stable, the indicator will
hold that weight on the screen for 3 seconds then go back to general weighing
●Pressing the hold key is unnecessary, holding is done automatically when the scale
is stable
4. Average Hold: Used for animal weighing, the indicator will display the average
weight sampled from 1-9 seconds (Set in the HAt parameter setting)
● Add animal to scale and press HOLD
●Indicator screen will show “LOC” for 1-9 seconds, then display the average weight
from those seconds
●Press HOLD again to exit holding mode
5. Auto Average Hold: Used for animal weighing, the indicator will display the aver-
age weight sampled from 1-9 seconds without the need to press the HOLD key.
If the weight on the scale is above 20d (scales division x 20) and is stable, the indica-
tor will start grabbing the average weight sampled from 1-9 seconds (Set in the HAt:
parameter setting). In parameter Hadt you can set a determined loading time from
1-9 seconds. This is the time you have to load the animal before the scale begins to
take the average weight.
● Add animal to scale. Once weight is added to the scale, the scale will wait the set
loading time and then the Indicator screen will show “LOC” for 1-9 seconds
● It will then display the average weight from those 1-9 seconds for 3 seconds and
then repeat the process

6
CALIBRATION
To Enter Calibration Settings
●Turn indicator off. Then press and hold the HOLD and UNIT key simultaneously
while the indicator is turning on (Do this while indicator is counting down)
●The display will show [CAL kg], meaning the unit is kg
●Press UNIT if you want to change the unit to lb; Display will change to [CAL lb].
● Press HOLD to conrm your unit choice
●The display will most likely show [FS 5], meaning the max. capacity is 200kg,
(see tables below for other variances)
●Press UNIT to change the capacity range to your scales capacity
Calibration chart for kilogram usage:
Display Max. Capacity Division Calibration Loading Weight
FS 0 15.00kg 0.005kg 10kg
FS 1 30.00kg 0.01kg 25kg
FS 2 60.00kg 0.02kg 50kg
FS 3 100.00kg 0.02kg 75kg
FS 4 150.00kg 0.05kg 100kg
FS 5 200.00kg 0.05kg 150kg
FS 6 custom custom custom
Calibration chart for pound usage:
Display Max. Capacity Division Calibration Loading Weight
FS 0 33.00lb 0.01lb 22.05lb
FS 1 66.00lb 0.02lb 55.12lb
FS 2 132.00lb 0.05lb 110.25lb
FS 3 220.00lb 0.05lb 165.35lb
FS 4 330.0lb 0.1lb 220.5lb
FS 5 440.0lb 0.1lb 330.7lb
FS 6 custom custom custom

7
CALIBRATION cont.
If FS = 6
Please follow the procedure below:
●The display will show [dP 1], to indicate how many decimal places you want;
press UNIT to change, and press HOLD to conrm (ex. DP 2 = ####.##)
●The display will show [d 2], to indicate the division; press UNIT to change, and
press HOLD to conrm
●The display will show [CAP ], to indicate the max. capacity; press HOLD to
display the current max. capacity; press ZERO to left shift and UNIT to change the
capacity; Then press HOLD to conrm
●The display will show [LoAd ], to indicate the calibration loading weight; press
HOLD to display the current calibration loading weight; press ZERO to left shift and
UNIT to change the calibration loading weight; Then press HOLD to conrm
●The display will show [CALon] please follow the steps for FS values of 0-6 to
complete calibration
For FS values of 0 - 6
Please follow the procedure below:
●The display will show [CALon], press HOLD to enter into weight calibraion, or
press ON/OFF to exit calibration
●The display will show [noLoAd] for zero calibration; make sure there is no load
on scale, then press HOLD to conrm and zero calibrate
●The display will show [AdLoAd] for loading calibration; place calibration weight
as dened by the tables above on the scale
● Once the weighing platform is stable, press HOLD to conrm and calibrate
●The display will show [CALEnd] when the calibration is complete, press HOLD to
save the weight calibration and press ON/OFF to exit and go back to weighing mode

8
PARAMETER SETTINGS
●Press HOLD and UNIT together to enter the
parameter settings
●Use the following diagram to navigate through
the parameter settings:
Function Display Parameter Settings
Automatic
Power Off OFF 10 0 = turn off auto power off
10 = power off automatically if no change within 10 minutes
30 = power off automatically if no change within 30 minutes
60 = power off automatically if no change within 60 minutes
90 = power off automatically if no change within 90 minutes
Power Saving Mode BL 1 0 = turn off the backlight
1 = backlight only when the weight changes or keyboard is pressed
2 = constant backlight
Unit Conversion UnS 1 0 = turn off unit switch function
1 = kg/lb
2 = lb/oz
3 = kg/lb/oz
Hold Function Hold 1 0 = turn off hold function
1 = Peak hold - Grabs the highest weight
2 = Manual hold - Grabs the current weight
3 = Auto hold - Automatically holds data when stable
4 = Average hold - Averages the weight from a sample of 3 seconds
5 = Auto Average hold - Average hold without pressing HOLD
Hold Time Ht 1 0 = turn off hold time
1 = 15 second hold time
2 = 30 second hold time
3 = 45 second hold time
4 = 60 second hold time
5 = 90 second hold time
Average Hold
Sample Time HAt 1 1-9 Seconds = How many seconds sampled for the average weight
Loading Time HAdt 1 1-9 Seconds = How long the scale waits (So you can load the ani-
mal on the scale) until it begins to take the average weight
Zero Range
(% scale is allowed
to zero, anything
after will be tared)
2Ero 1 0 = turn off manually zero setting
1 = Initial zero range ±10% max capacity,
Manual zero range ±2%
2 = Initial zero range ±100% max capacity,
Manual zero range ±100%
Negative Range
(% it can go nega-
tive before an error)
nEgr 1 0 = -20d negative display
1 = 10% MAX negative display
2 = 100% MAX negative display
Communication
Setting trn 1 Set the serial interface data output method:
0 = Turn off serial interface data output
1 = Continuous sending mode, connect computer
2 = Print mode, connect printer
3 = Command request mode, connect computer
KEY Function
ON/OFF Exit
HOLD Conrm
ZERO Move Left
UNITS Move Down
Otros manuales para OP-902
1
Este manual sirve para los siguientes modelos
1
Tabla de contenidos
Otros manuales de Instrumento de medición de Optima Scale
Manuales populares de Instrumento de medición de otras marcas

Endress+Hauser
Endress+Hauser Proline Promag 50 Especificaciones técnicas

Siemens
Siemens SITRANS F Coriolis FCT030 Manual de lista de piezas

KLINGER
KLINGER CMF V Series Manual de usuario

EXFO
EXFO FTB-2 Manual de operación y mantenimiento

Keysight
Keysight M8290A Manual de usuario

ADTEK
ADTEK MW-5 Manual de usuario