ThankyouverymuchforpurchasingOPTEXFAproducts.
Thisdeviceisanon-contactthermometertoconverttheinfraredenergyemittedfromthe
surfaceofanobjectintotemperature.Thisthermometermeasuresthesurfacetemperature
ofsolidandliquidwithoutcontactingthem.Thetemperatureofgascannotbemeasuredby
thisthermometer.
Contents
Introduction................................................................................................2
SafeUsage................................................................................................3
Warnings&CautiononEnvironmentandUsage......................................3
Specications............................................................................................4
ExternalDimensions/PartName...............................................................5
WiringDiagram..........................................................................................6
Mounting/Installation..................................................................................8
FieldofView..............................................................................................9
HowtoUse..............................................................................................10
FunctionList............................................................................................ 11
SetupofFunctions...................................................................................12
BANK Setting.....................................................................................12
ResponseTimeSetting......................................................................12
AdjustmentSetting.............................................................................12
AnalogOutputSetting........................................................................15
AlarmOutputSetting.........................................................................16
TriggerInputSetting..........................................................................18
DisplayOffSetting.............................................................................20
SaveSettings.....................................................................................21
SettingInitialization............................................................................21
Troubleshooting.......................................................................................22
Maintenance............................................................................................22

3
Thisproductisnotaclinicalthermometerandtherefore,cannotbeusedformedical
purposes.
Useofcontrolsoradjustmentsorperformanceofproceduresotherthanthose
speciedhereinmayresultinhazardousradiationexposure.
Safe Usage
Thisinstructionmanualcontainsvariouswarningsforyoursafetyandproperusagetoavoid
possible personal injury. Please be sure to heed the warnings and strictly follow safety
instructions.
This symbol signifies that improper usage may result in injuries or
damage.
EnvironmentalWarnings-Warning-Caution
UsageWarnings-Warning-Caution
Caution
Caution
Caution
DONOTUSETHERMOMETERWHENITISWETOR
SOAKEDINLIQUID.
Althoughtheproductiswater-resistant,usingitwithwater
dropsonitslensorinwetconditionsmaycauseincorrect
measurement.
AVOIDMEASURINGSHINYOBJECTS.
Shinyobjectsreectsurroundingtemperatures.Incorrect
measurementmayoccuralthoughspecifyingtheemissivity
ratecancorrectit.
DONOTUSEWITHNON-STANDARDVOLTAGE.
Usingtheunitoutof12to24VDCrangemayresultin
damagetotheunit,shorts,resandinjuries.Insuchcases,
immediatelyswitchtheunitoff.
DONOTTOUCHTHELENS.
Donottouchthelenswithhardorsharpobjects.Donot
insertforeignobjectsintothelightreceivingpart.Otherwise
incorrectmeasurementwilloccur.
KEEPTHETHERMOMETERAWAYFROMDIRECT
SUNLIGHT,DUST,HIGHTEMPERATURESANDHIGH
HUMIDITYDURINGUSEANDSTORAGE.
Otherwiseitmaycauseirreparabledamageorincorrect
measurement.
DONOTDROPTHISTHERMOMETERNORGIVE
ASTRONGIMPACTTOIT,WHICHMAYCAUSE
IRREPARABLEDAMAGEORINCORRECTMEASUREMENT.
Otherwiseitmaycauseirreparabledamageorincorrect
measurement.
DONOTLETTHETHERMOMETERTOUCHTHEOBJECT
THATISBEINGMEASURED.
Thisproductisanon-contactthermometer.Touchinghigh-
temperatureobjectmaycausedeformationofthemeter,
irreparabledamageorincorrectmeasurement.
DONOTBRINGTHETHERMOMETERCLOSETO
ELECTRICALLYCHARGEDOBJECTS.
Otherwiseitmaycauseirreparabledamageorincorrect
measurement.
KEEPTHETHERMOMETERAWAYFROMSUDDEN
CHANGEINAMBIENTTEMPERATURE.
Suddentemperaturechangemaycauseincorrect
measurement.Startmeasurementwhentemperaturehas
becomestableafterleavingthemeterforawhile.
KEEPTHETHERMOMETERAWAYFROMPRODUCTS
WHICHPRODUCESTRONGELECTROMAGNETIC
WAVES.DONOTUSEINANATMOSPHERECONTAINING
CORROSIVEGASESOREXPLOSIVEGASES.
Otherwiseitmaycauseirreparabledamageorincorrect
measurement.

4
Accessories:MountingNut(M12×P1.0)×2
Options:BlackBodyTape HB-250, Mounting FittingCS-FB12, Changeable Laser Marker CS-LDP,
AirPurgeCollarCS-AP1,CFLensCS-CF01,MassiveHousingUnitCS-MH01
*Thespecicationsaresubjecttochangewithoutnoticeforproductimprovement.
*ForChinaRoHS,pleaserefertohttps://www.optex-fa.com/rohs_cn/
Specications
Model
CS-30TAC-HT CS-40TAC-HT
CS-30TAC CS-40TAC
Temperaturerange 0°C to 1000°C -40°C to 500°C
Areasize φ30/500 mm
22:1
φ40/500 mm
15:1
φ30/500 mm
22:1
φ40/500 mm
15:1
Optics Siliconlens(Water-repellentcoat,Oil-repellentcoat)
Detectionelement/
Wavelength Thermopile/8to14μm
Responsespeed 150msec/90%
Accuracy
0 to 200°C:±2°C
201 to 1000°C:Readingvalue±1%
-40 to 0°C:±3°C
1 to 200°C:±2°C
201 to 500°C:Readingvalue±1%
Measurementcondition:ε=1.0,23±5°C,Measurementdistance250mm,Targetsize
φ
100 mm
Repeatability Upto200°C:±1.0°C
201°Candmore:±0.5%
Emissivityrateadjustment
0.1to1.2
Powersupply 12to24VDC±10%
Consumption 120mA(Max.load),80mA(Ecomode)
Ambienttemperature Sensorhead:0to180°C,Amplier:0to65°C Sensorhead:0to100°C,Amplier:0to65°C
Ambienthumidity 35to85%(withoutcondensation)
Storagetemperature 0 to 70°C
Waterresistance Sensorhead:IP69K,Amplier:IP40
Vibrationresistance 10to55Hz,1.5mmamplitude,2hourseachforXYZdirections
Mainmaterial Sensorhead:SUS,Amplier:ABS
Dimensions Sensorhead:Approx. φ14(M12)×34mm,Amplier:Approx.35×52×39mm
Weight
Sensorhead:Approx.100g(includingacableof3m),
Amplier:Approx.200g(includingacableof2m)
Display LED
Resolution 1°C
Analogoutput 4 to 20 mA
Analogoutputresolution 0.5°C
Analogoutputaccuracy ±0.5%or±1.0°C
Analogoutputupdatingtime
10msec
Analogoutputallowableload
250Ω
Analogoutputimpedance 47Ω
Contactoutput PhotoMOSFET×2(ccontact×2),300mA/30VDCorless
Others Triggerinput
Bank×4
Applicableregulations
EMCDirective(2014/30/EU),RoHSDirective(2011/65/EU),ChinaRoHS(MIITOrderNo.32)
Applicablestandards EN61326-1:2013

5
External Dimensions/Part Name
Lenssurface
φ7
M12×P1.0
10
30
φ14
●Sensorhead
Sensorcable
φ4×3000(MAX)
Cablecover
SerialNo.
▲/▼buttons
ENT.button
Outputcable
φ6×2000(MAX)
CableNo.
Maindisplay
●Amplier
Subdisplay
64
52
35 39
31
[mm]
DINrailattachmentpart
CableNo.
φ10
4

6
Wiring Diagram
BANK (1) (2)
1OPEN OPEN
2CLOSE OPEN
3OPEN CLOSE
4CLOSE CLOSE
15pieces(AWG26)
Dotmark1
Dotmark2
No.
Outputcable
Descriptions
Line
color
Dotmark
Color
Quantity
1 Pink Red 1 Powersupply 12to24VDC
2Gray Black 1GND
3 White Red 1
Analogoutput
4-20 mA
+
4 White Black 1-
5 Pink Black 1
Externaltrigger
Input
6Gray Black 2Bankswitch (1)
7 White Red 2 (2)
8Yellow Red 1 Alarmoutput
H
N.C.
9Gray Red 1 COM
10 Yellow Black 1N.O.
11 Orange Red 1 Alarmoutput
L
N.O.
12 Gray Red 2 COM
13 Orange Black 1N.C.
14 Orange Red 2 Donotwire
15 Orange Black 2ConnecttoGND
16 *4(Shieldedcable) -
●Outputcable
*1 Connecttothe4-20mAinputofananalogdevice.
Analogoutputallowableload250Ωandanalogoutputimpedance47Ω
*2 Externaltrigger:Switcheson/offintherangefrom2to5.
*3 Bankswitch:SwitchesOPEN/CLOSEintherangefrom2to6orfrom2to7toselectabank.
*2 *3
c contact
300mA/30VDCorless
c contact
300mA/30VDCorless
Donotwire
ConnecttoGND
*1
*4 Whenyoucut theoutputcablesshorter, a shielded cableforreinforcementwillcomeout.Cutthe
shieldedcabletopreventitfromcontactingwithothercables.
*Cables not used should be cut so that they do not contact with other cables, and insulated with
adhesivetapeorbyothermethods.

7
Sensor
cable
Connector
No.
Green 1
Yellow 2
Brown 3
Shield 4
-5
Whenyoucutthesensorcables,ensuretoperform
endtreatmentandconnectionofthecables.Inthe
caseoftheshortcircuitofthesensorcables,the
sensorcannotmeasureprecisely.
(1)Cutthecablestoadesiredlengthandtreattheir
endasshownintheleftgure.
Aguidewireissheathedintheshieldedmesh
cable.Cuttheothercablesthantheguidewireat
theirbase.
(2)Passthesensorcablebundlethroughthehole
ofthecablecoverandtightenthecabletieatthe
pointshownintheleftgure.
*Theserialnumbersareprintedonthecablecover.
Makesuretoputeachsensorcablesbacktothe
sameholesthatyouremoved.
(3)Connectthecablesandshieldedcable(guide
wire)totheconnector.
(4)Inserttheconnectortothecablecover.
*Themetalplateofconnectorshouldbeplacedonto
theclaspinthecablecover.
●Sensorcable
5
15
5
Cabletie
(ofapprox.2
mmwidth)
Guidewire
(1)
(2)
(3)
(4)
[mm]
○ Connectingthesensorcables
Keepingtheconnectorpushed
in,tthecablecovertothe
groovesoftheamplier,and
insertthecablecover.
○ Removingthesensorcables
Pinchthetabofthecable
coverandpullthecable
coverupward.
Sensorheadandamplierhas
beencalibratedasaset.
Makesuretousesensorhead
andamplierwiththesame
serialnumber.
Amplierhasnotaprotectivestructure. Whenyouconnecttheoutputcableandthesensorcable,
makesurethatthewateroroildoesnotpenetratetoamplieralongthecable.Whenusedinsuch
anenvironmentwherewateroroilmightgetintotheamplier,pleaseuseoptionalprotectivecase.
(EquivalenttoIP65)
Pleasenotedthatcontinuoushotwaterwithhighpressuremaycausebreakageofthecableorcovering.

8
Mounting/Installation
●Sensorhead
●Amplier
MounttheamplierontheDINrailusingthehooksonitsbottom.
Avoidalocationwherewateroroilmayspillonit. ○ Mounting
MakethetwohooksontheSWsidecatch
theDINrailandpushintosettheamplier.
(1)
(2)
(1)
(2)
○ Removing
PushthetwohooksontheSWsidetothe
oppositedirectionandraisetheamplier.
TheexternalscrewisM12×P1.0.
Fixsecurelyintotheholeofφ12mmormoreusingtheattachedhexagonnut.
Theoptionalmountingttingcanhelpyouadjusttheangleeasily.
○Mounting
Mountthesensorheadperpendiculartothe
target.
Avoidalocationwherethesensorheadmay
beexposedtovibrationorimpact.
Theambienttemperatureshouldnotrapidly
changeandshouldbewithintheoperating
temperaturerange.
Donotxthecablewhenitisbentor
excessiveloadisappliedtoit.
Althoughthesensorheadiswater-resistant,
waterdropsonthelensmaycauseanerror.
90°
Target

9
Field of View
[CS-30TAC/CS-30TAC-HT]
[CS-40TAC/CS-40TAC-HT]
φ12/100
φ21/300
φ30/500
φ14/100
φ27/300
φ40/500
[CS-30TAC/CS-30TAC-HT + CF] [CS-40TAC/CS-40TAC-HT + CF]
φ0.6/12 φ1.0/12
φ25/50
φ26.5/50
[Forcorrectmeasurement]
Therangeofeldofviewisequivalentto90%ofopticalresponse(energy).
Thetargetshouldbesufcientlylargerthantheeldofviewshownabove.
Whenmeasuringahigh-temperaturetarget,keepasmuchdistancefromitaspossiblewithintherange
oftheeldofview.Otherwiseitmaycauseincorrectmeasurement.
●WhenusingtheoptionalCFlens
* WhentheCFlensisattached,correctionisnecessarybecauselightintensityreceivedfromthetarget
decreasesby20to30%.
Whenmeasuringaminutespot,therecommendedtargetsizeisapproximately1.5timesoftheeld
ofviewshownabove.
[mm]
[mm]
Areasize/Distance
Areasize/Distance
D(distance):S(area)=22:1
D:S=15:1
φ7/0
φ7/0

10
How to Use
(1) Conrmthewiringsandthenturnthepoweron.
Thedisplayashesandthentemperaturemeasurement
starts(NormalOperation).
(2) Conrmthatthemeasurementvaluechangeswhenyouput
yourhandintotheeldofview.
*LeavetheUnitforawhileafterthesensorheadismounted.
Temperaturechangemaycauseincorrectmeasurement.
25
E1.00
MODE
8888
NormalOperation
Press3sec.ormore
SettingOperation
*MakesuretoperformSaveSettings.
IfyoucancelinSaveSettings,allsettingsenteredarenotsaved.
8.8.8.8
8.8.8.8
(3) ToconrmorchangethesettingspresstheENTERbutton
forthreesecondsormore.
ThedisplayashesandthenSettingOperationstarts.
(4) Selectthefunctionwith▲/▼buttons.
ENT.
END
SET
SaveSettings
BANK
1
➊Bank Setting
DELY
1
➋Response
Time Setting
TADJ
E1.00
➌ Adjustment
Setting
ASCL
-40
➍ AnalogOutput
Setting
ALAM
400
➎ AlarmOutput
Setting
TRIG
NoN
➏ TriggerInput
Setting
ECO
OFF
➐DisplayOff
Setting
▲button:
▼button: