2) Part Functions
① Indicator
This LED indicator shows the measurement value of one of tissue blood flow
parameters selected by "Parameter selection switch ②". Also, it shows the
output mode selected by Mode switch ③.
FLO-C1 measures three blood flow parameters, FLOW, MASS and VEL.
FLOW, MASS and VEL. represent tissue blood flow, tissue blood volume and
tissue blood velocity respectively. The ranges of the three parameters are 0-
100, 0-2000 and 0-10.0. All of the parameters can be always monitored
from "Output 1" regardless of the indicated parameter.
Output mode has three modes. "MEA" mode is for measurement, "0" mode
shows 0 (zero) on this indicator and sets the output to 0 (V) and "CAL" mode
shows calibration value and outputs its voltage. The calibration values of
FLOW, MASS and VEL. on Indicator are 10.0, 200 and 1.0 respectively. The
voltage of the all calibration value is 1.0 (V).
② Parameter selection switch
One of the three parameters, FLOW, MASS and VEL., showed on
"Indicator" is selected by this switch.
③ Mode switch
Output mode of "MEA"、"0" and "CAL"is selected by this switch.