4
Model 2
30 mm x 22 mm x 6 mm
4 grams
12 Months
Storage Conditions: Temperature -4 to 95 °F (-20 to 35 °C), Relative Humidity 10–100%, Atmospheric Pressure 50-106
kPa
Transport Conditions: Temperature -40 to 122 °F (-40 to 50 °C), Relative Humidity 10–100%, Atmospheric Pressure 50-
106 kPa
Use Conditions: 77 to 113 °F (25 to 45 °C), Relative Humidity (non-condensing) 30-100%, Atmospheric Pressure: 70 =
106 kPa pressure range of 700 hPA to 1,060 hPA
Axillary (under the arm)
77 to 113°F (25 to 45°C ) Relative Humidity 30–100% Atmospheric Pressure 70-106 kPa
Green: Ready to Use Temperature is in normal range.
Yellow : Caution – prompt response by user is needed. Indicates a possible fever of 100.4°F (38.0°C)
Red : Warning – Immediate response by user is required. Indicates a temperature equal to or higher than the user defined
limit.
Accuracy and regular ability to measure and communicate results are considered as essential performance for the
purposes of risk and electrical safety analysis. Also ability to communicate battery voltage.
Non-rechargeable
Nominal voltage Range: 2.5-3.2 V
±0.3°C (±0.5°F) in the range of less than 35.8°C (96.4°F)
±0.2°C (±0.3°F) in the range of 35.8°C (96.4°F) to less than 37°C (98.0°F)
±0.1°C (±0.2°F) in the range of 37.0°C (98.0°F) to 39.0°C (102.0°F)
±0.2°C (±0.3°F) in the range of Greater than 39.0°C (102.0°F) to 41.0°C (106.0°F)
±0.3°C (±0.5°F) in the range of Greater than 41.0°C (106.0°F)
90 feet
IOS, Android.
Designed to comply with the following standards:
ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization
ASTM E112:2018 Standard Specification for Electronic Thermometer for Intermittent Determination of Patient
Temperature
ASTM E1104:2016 Standard Specification for Clinical Thermometer Probe Covers and Sheaths
AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R) Medical
Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential
Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
ISO 80601-2-56 First Edition 2009-10-01Medical Electrical Equipment - Part 2-56: Particular Requirements For
Basic Safety And Essential Performance Of Clinical Thermometers For Body Temperature Measurement.
IEC 60601-1-11:2010 Medical electrical equipment- Part 1-11: General requirements for basic safety and essential
performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical system used in
the home healthcare environment.
IEC 60601-1-2 ed4.0 (2014-02) Medical electrical equipment- Part 5.2.1.1(b): the performance of the ME
EQUIPMENT or ME SYSTEM that was determined to be ESSENTIAL PERFORMANCE and a description of what the
OPERATOR can expect if the ESSENTIAL PERFORMANCE is lost or degraded due to EM DISTURBANCES (the defined term
“ESSENTIAL PERFORMANCE” need not be used).
IEC 60601-1-2 ed4.0 (2014-02) Medical electrical equipment and systems- Part 5.2.2.1(a): the compliance for
each EMISSIONS and IMMUNITY standard or test specified by this collateral standard, e.g. EMISSIONS class and group
and IMMUNITY TEST LEVEL.
FCC Part 15 Rule 247 and/or 249 for intentional radiation of radio frequency
ASTM D4169-14 Standard Practice for Performance Testing of Shipping Containers and Systems
The NIRA Temp shall comply with the medical device and radiofrequency emitting device regulations for the territories
it will be marketed to. In particular, it shall comply with Class II medical device regulations of US Food and Drug
Administration (FDA) in the United States of America and FCC regulations for intentional emitters of RF radiation.
IP67
Read instructions carefully before use.
Model Number
Device Dimensions
Device Weight
Device Operating Life
Environmental Conditions
•Storage
•Shipping
•Use
Temperature Measurement Location
Temperature Range
Temperature Indications
Accuracy or Essential Performance
Battery
Accuracy
Transmission Distance
Compatible Smart Devices
Standards
Regulatory Requirements
Water Ingress Protection
IMPORTANT:
Specifications