Netech Delta 1600 Manual de usuario

Innovative Instruments Since 1987
Delta 1600
AED TESTER
User Manual
(Automated External Defibrillator Tester)
Intended Use: Delta 1600 AED Tester is a precision instrument that
simulates various cardiac waveforms and measures discharged energy.
It assures AEDs detect the waveforms, audibly advise shock, and
discharge energy as per manufacturer’s specications.

2 | Delta 1600 - Quick-Reference Manual
Table of Contents
Notices.............................................................................. 3
Unpacking and Inspection of the unit .................... 3
Contact Information............................................... 3
Warnings............................................................... 3
Intended User........................................................ 4
Environment.......................................................... 4
Introduction..................................................................... 5
Delta 1600 Features.............................................. 5
Specications................................................................. 6
Energy Measurement............................................ 6
GeneralSpecications.......................................... 7
Part Numbers and Ordering Information............... 8
Preparation for Use............................................... 9
InitialSetup...................................................................... 12
Introduction........................................................... 12
Shock Advisory Test.............................................. 13
AUTO Function Test.............................................. 15
ECGSimulation............................................................... 16
Rate Selection...................................................... 16
Amplitude Selection.............................................. 16
Warranty........................................................................... 17

Delta 1600 - Quick-Reference Manual | 3
Notices
Unpacking and Inspection of the unit
Before unpacking the Delta 1600, inspect the shipping box for any
visual damage. If damage is found, do not unpack the unit and
immediately notify the shipping carrier. If no damage is found to the
shipping box, open the box and perform a visual inspection of the
Delta 1600. If any damage to the unit is observed please contact
Netech Customer Service.
The Service Return Form may be obtained at our website http://www.
netechcorp.us/RMARequests or contact our Netech Customer
ServiceDepartment[email protected]
Contact Information
NetechCorporation,110ToledoSt.Farmingdale,NY11735
Phone: 631-531-0100, 1-800-547-6557
Calibration
The Delta 1600 is factory calibrated, thoroughly tested, and meets
Netech’s ISO 9001-2015 quality standards.
Any attempt to remove or tamper with the calibration seal will void
the warranty.
Warnings
Caution Important Safety Information
Hazardous Voltage
Conforms to European Union Directive
SYMBOL DESCRIPTION

4 | Delta 1600 - Quick-Reference Manual
Intended User
Environment
The Delta 1600 is intended to be used by a
trained technician.
The Delta 1600 is intended for testing and
verifying the performance of AEDs.
The Delta 1600 should never be used in a
patient care location where it might come in
contact with a patient.
The Delta 1600 is intended to be used
within the published specifications. Any
unauthorized alteration or modifications will
result in a hazardous condition or improper
operation.
The Delta 1600 should never be used in
harsh environmental conditions outside
the specifications as this will affect the
performance of the unit.
The Delta 1600 should never be used in wet
condition or in a flammable environment.

Delta 1600 - Quick-Reference Manual | 5
Introduction
The Delta 1600 is a precision, compact, and battery operated
instrument for testing automated external debrillators (AEDs). The
Delta 1600 simulates various cardiac waveforms to ensure that the
AED can detect, analyze and audibly advise shock. In addition to
the shock advisory test, Delta 1600 measures discharged energy
of the AED, thereby ensuring that the AED complies with specied
requirements. The dedicated function keys and menu driven display
on the unit makes it simple for operation.
The Delta 1600 has a built-in load resistance of 50 ohms to simulate
the human body impedance. To establish electrical connection
between the AED and the Delta 1600, manufacturer specic
interface cables
(contact Netech Sales for custom cables)
are required. The
interface cable must be connected between the AED and input jacks
of Delta 1600. Once the AED is discharged, Delta 1600 samples the
pulse, calculates and displays the delivered energy.
Delta 1600 Features
lUnique, one-of-a kind instrument for testing all AEDs including
Pulsed (Schiller AEDs)
lCompact, lightweight and easy-to-use
lAuto Function – Eliminates multiple discharges
lLarge backlit Graphics Display
lMultiple Arrhythmia simulations
l7.4 Volt Li-Ion Rechargeable battery

6|Delta 1600 - Quick-Reference Manual
Specifications
Energy Measurement
EnergyOutput:
Maximum Energy : 500 J
Display Resolution : Based on Auto Range
Resolution : 1 J
Accuracy : ±1 % of reading ± 1
LoadResistance:
50 ohms ±1 %; non-inductive (< 1 μH)
Sampling Time : 20 uS
Measurement Time Window : 60 ms
Pulse Width : 60 ms
PatientSimulator
Waveform Simulation: ECG (NSR),
ECG Rate: 60, 70, 80, 90, 100, 110, 120, 130, 140, 150, 160, 170,
180, 190, 200, 220, 240. BPM
Accuracy: ± 1%
Waveform Amplitude: ECG (AMP)
ECG Amplitude:1.0, 2.0, 3.0, 4.0 mV
Accuracy: ± 5% peak to peak

Delta 1600 - Quick-Reference Manual | 7
ArrhythmiaSelections:(Shockadvisorywaveforms)
Ventricular Fibrillation Coarse (VFBC)
Ventricular Fibrillation Fine (VFBF)
Ventricular Tachycardia (VTAC)
Atrial Fibrillation (AFIB)
General Specifications
TemperatureRange:
Operating : +15°C to +35°C
Storage : 0°C to +50°C.
Display:
Type: LCD with LED backlit Graphics display
Power: 7.4 Volt Li-Ion Rechargeable Battery
100-230 VAC input 8.5V Battery Charger
Housing:
High impact ABS plastic enclosure
Dimensions:
D x W x H :10.25 x 6.25 x 2.5 Inch
26 X 16 x 6.4 Cm

8 | Delta 1600 - Quick-Reference Manual
Part Numbers and Ordering Information
PartNo : 630
Delta 1600 AED Tester
(All Standard Accessories are included).
StandardAccessories: (Included with the unit)
630-Li-Battery: Li-ion rechargeable battery
630-LI-Charger: Li-ion Battery Charger
630-HARD-CASE: Hard carrying case
630-6383: Open-ended interface cable
OptionalAccessories:
PartNo. InterfaceCable
630-651-HP HP/Philips
630-651-ZOLL ZOLL
630-651-SURVIVAL Cardiac Science / Survivalink/Burdick
Cardiovive
630-651-PHYSIO Physio-Control / Life-Pak / Medtronic
630-612 Philips HeartStart FRx
630-615 Philips (and Laerdal) HeartStart FR3
630-617 Philips (and Laerdal) HeartStart FR/FR2/
FR2+
630-618 Philips HeartStart OnSite and HeartStart
Home
630-619 Debtech Lifeline or Lifeline AUTO AED
630-620 Debtech Lifeline VIEW/ECG/PRO
630-621 CU Medical Systems
630-622 ZOLL CPR-D-padz
630-623 ZOLL AED Plus or ZOLL AED Pro
Note:
Various Manufacturers’ model specic interface cables are available.
Please call for the part numbers or visit www.NetechCorp.us

Delta 1600 - Quick-Reference Manual | 9
Peparation for Use
The Delta 1600 is shipped with the following items included in the
customized hard case:
• Delta 1600 AED Tester.
• Li-Ion Rechargeable Battery Charger (110/220 V input)
• User manual / USB Jump Drive
Before getting started it is important to get
familiar with the control and operation of the
unit.
Instrument Familiarization
Charge Indicator LED
Li-Ion
Charger Input Power \
ON/OFF Keys
Menu
Selection Keys
Graphics Display
AED Interface Jack

10|Delta 1600 - Quick-Reference Manual
Follow AED’s shock advisory audio
response.
Display
The Delta 1600 incorporates a high contrast LCD graphics display
with an LED backlight for easy viewing.
Key Pad
The keypad utilizes soft touch tactile keys, providing a uid resistant
barrier to the internal circuitry. The dedicated key functions and
interactive menu make it easier to use.
The following are the description of the Key Functions:
F1,F2,F3,F4,<,and>:These are interactive keys associated with
respective function selections.
“ESC”:Reverts to the previous menu
“AUTO”: Pressing “AUTO” key changes the Arrhythmia (previously
selected) to VFBC. Once the discharged energy is displayed,
arrhythmia changes automatically to NSR (Normal ECG @ 90 BPM)
This will eliminate multiple discharges by the AED thus saving the
AED batteries.
AED Interface Jack:
2 RED Shrouded banana jacks.
Power ON/OFF Switch:
This switch powers the unit on/off. The power ON/OFF switch is
located on the lower side panel.
Tabla de contenidos