
Apex Probe Module 2 (PM2) - Setup Guide Page 7
CONDUCTIVITY PROBE CALIBRATION
NOTE: Set the Conductivity Range and Temperature Compensation settings before
calibrating the Conductivity Probe. See the previous section for instructions on setting
these parameters.
To calibrate a Conductivity probe:
Apex Display: Go to Setup –> Cond Setup –> Cond Calibrate
1. If you have more than one Conductivity probe enabled on your system, use
the up/down arrow keys to choose the probe you wish to calibrate, press
Select.
2. Remove the conductivity probe from the tank and rinse with RO/DI water,
then dry it off. It is not necessary to let the probe dry out – just gently
shake off excess water and blot the tip with the corner of a paper towel.
3. Wait for the numbers on the bottom of the LCD screen to stop changing. It
does not matter what value is displayed only that it is not changing. When
the display stops changing press the select button.
4. Use the Up and Down buttons to select the calibration solution. Press the
select button when the correct value is displayed. See Table 1 -
Conductivity Range Options for the recommend calibration solution.
5. Place the Conductivity probe into the calibration solution. Wait for the
numbers on the bottom of the LCD screen to stop changing. When the
display stops changing press the select button.
6. The conductivity probe should now be properly calibrated.
NOTE: Many conductivity calibration solutions do not have the same ionic
composition as salt water so using a refractometer to measure the salinity of the
calibration solution is not valid. For example the Neptune 53.0 mS/cm
conductivity calibration standard will not have a salinity of 35 PPT or a specific
gravity of 1.0259.
CONDUCTIVITY PROBE MAINTENANCE
Conductivity probes should be periodically cleaned for best performance. The period
between cleanings is up to the user but typically every 1-3 months is normal. To clean
the probe, gently brush the sides and tip of the probe with a soft brush to remove any
debris. Rinse the probe in tap water. Soaking the probe in acid (vinegar) or alkali (borax)
solution to remove mineral deposits is not typically needed but can be used to try and
bring back an old probe that has stopped working.