Naneos Partector 2 Manual de usuario

Partector 2
Aerosol Dosimeter
Light Version
Operation Manual
English
Jan 2022
A nanoparticle dosimeter for easy workplace exposure monitoring

www.naneos.ch
2 | 24
Document information
© naneos particle solutions gmbh, 2022
Related device versions: v 1.0 standard pump & silent pump
Related firmware versions: ≥212
Older firmware may not offer all features
described in this manual
Document Date: 2022-01-11
Document Revision: X

www.naneos.ch
3 | 24
Abbreviations
DC Diffusion charging, diffusion charger
HV High voltage
LED Light-emitting diode
LDSA Lung deposited surface area
N Number concentration
d Average particle diameter
S Total surface area
PM0.3 Mass of particles smaller than 0.3 m
SD Micro secure digital memory card
Safety information
General notes and warnings
This operation manual must be read completely before using the
Partector 2. Incorrect use or handling and any consequences arising from
this may damage the device or endanger personnel.
The manufacturer declines all liability resulting from incorrect use and
handling.
Safety notes
Never operate the instrument at condensing conditions.
Do not blow into the instrument.
Do not open the Partector 2 as you may damage it.
The Partector 2 is a sensitive instrument made for measuring
nanoparticles. Sampling too much coarse dust will eventually lead to
deteriorating instrument performance. Use a cyclone when measuring in
environments where large numbers of microparticles might be present.
Do not operate the Partector 2 in an explosive atmosphere or in the
presence of flammable gases or fumes.
Have your instrument recalibrated once a year.
Use only the original USB power supply and USB cable.

www.naneos.ch
4 | 24
Instrument specifications
Measurements and accuracy Partector 2 (Light Version)
Measured values: LDSA Lung deposited surface area
Concentration range: LDSA 0 - 12’000 m2/cm3
Typical accuracy: LDSA ± 30 % or ± 3 m2/cm3
Particle size range: 10 nm to 10 m
Displayed LDSA value is only accurate in the size range of 20-400 nm;
however, the instrument can be used to measure micron-sized particles too.
Noise floor: ~ 0.5 m2/cm3in particle-free air (LDSA)
Time resolution: 1 s
Response Time: 2/4/8/16 s, user-adjustable
Technical specifications Partector 2 (Light Version)
Inlet flow: 0.5 l/min
Operating Temperature: 0 –40 °C
Storage Temperature: -10 –50 °C
Relative Humidity: 10 % to 90%, non-condensing
Operating height: 0 - 3000m above sea level
Battery: Rechargeable Li-Ion, 12Wh
Battery lifetime: ~ 24h (with a new battery; aging will occur;
low power mode enabled, ~20h with silent
pump)
Battery Charging: via USB, 5V / 1.5A
Data storage: SD-card
PC Interface: USB/serial by virtual COM-Port
Serial port adapter for SD card available
BT wireless range: 3 - 30 m, depending on obstacles
Noise: ~ 35dB in 1m distance (standard pump)
inaudible (silent pump)
Dimensions: 8.8 x 14.2 x 3.4 cm
Weight: 400 g (390 g with silent pump)
Pollution degree: 2, or with cyclone at inlet 3.

www.naneos.ch
5 | 24
Checking the shipment contents
When unpacking the device, verify that you have received the following
items. If any of the components are missing or damaged, contact naneos.
Type
Items
Standard
Partector 2 device
USB cable, 1.2m
SD-card 8 GB
SD/SD-card adapter or SD/USB-adapter
Carrying sleeve
USB charger
12 cm Tygon tubing
Peli case
Storage and transport
To avoid damage during storage and transport, please use original
packaging.
Avoid locations with high/low temperatures and high humid conditions or
places that can get wet.
Do not store the device with an empty battery.
Do not store the device in direct sunlight.
For longer storage, mount the protection cap.

www.naneos.ch
6 | 24
Getting started
Instrument description
The Partector 2 (Light Version) is a hand-held, battery powered instrument
to measure the lung deposited surface area (LDSA). It displays the
measured data on a graphical display and stores it on a SD card. A user-
defined LDSA alarm level can be set. If the alarm level is exceeded, the
instrument’s internal buzzer is turned on to produce an audible warning.
The device can communicate with Android smartphones that support BT
low energy.
Figure 1: Partector 2 main parts
1 aerosol inlet 5 display 9 charge LED
2 USB Port 6 right soft key 10 serial plate
3 SD card slot 7 middle soft key
4 power button 8 left soft key
10
, 12Wh

www.naneos.ch
7 | 24
Aerosol inlet and outlet
The aerosol inlet is located at the top of the instrument. The internal pump
provides a regulated air flow of about 0.5 l/min. The standard aerosol inlet
has a 6-mm barb for tubing connection and contains a coarse wire mesh
to keep coarse dirt and small insects out of the instrument.
The air exits the device via the slits in the right-hand side of the housing.
Note that the air is not filtered internally, so do not connect the instrument
to hazardous nanoparticle sources!
Tubing
When using the Partector 2 with tubing make sure not to use
conductive silicone tubing. It releases silanes which lead to rapid
fouling of the Partector's corona wire. We recommend using PVC tubing
instead –a nice flexible version is Tygon S3 E-3603, a sample of which
is included in the Partector 2 case. If in doubt about using tubing, please
contact us. In any case, tubing should only be as long as necessary to
prevent particle losses in the tubing.
Figure 2: different tubing options
Supplying power
To chargethe Partector 2 or to operate it forperiods longerthan the battery
lifetime, connect the Partector 2 USB port to a PC, Laptop or to a USB-
charger or to a mobile power bank to charge the device.
While charging, the blue charge LED on the front panel lights up. When
the Partector 2 is fully charged, the LED turns off.
The Partector 2 supports fast charging (1.5A charging current). For
optimum charging, use the original USB power supply and USB cable.
Some PC’s and USB power supplies only provide 0.5A charging current,
which will prolong the charging duration.
Turning the device on and off
To switch on the device, press the power button. To turn it off, press and
hold the power button for two seconds and release it when the goodbye
message appears.

www.naneos.ch
8 | 24
Insert / remove SD-card
To store data on the microSD-card, insert it in the Partector 2 as shown in
Figure 3 till it snaps in.
To remove the SD-card, press on the inserted
card which will force the release. To insert the
card in your SD-card slot on a PC you need to
insert the card first in the SD/SD-card adapter or
SD/USB adapter provided.
Recording measurements on the SD-card
If a SD-card is inserted in the Partector 2 before startup, the device
automatically starts recording on the SD-card when the device is turned
on. It automatically generates a filename depending on start date, time
and device serial. For more information about this, read the “Partector 2
data file” section.
If you insert the SD-card while the device is running, the data
logging will not start automatically. You need to turn the device
on after inserting the card. The Partector 2 will complain that
the SD card is missing if it is not in the slot at startup. When the
Partector 2 is recording data, you can see a blinking red dot in
the status bar of the main screen.
Do not remove the SD-card while the device is running. The data file
needs to be closed properly, otherwise it will be corrupt.
Figure 3: micro-SD
card slot

www.naneos.ch
9 | 24
Keypad functions & symbols
In the picture below, you can see the Partector 2 (Light Version) display
screens:

www.naneos.ch
10 | 24
Figure 4: Partector 2 (Light Version) menu screens
Home screen
On the home screen you can see the LDSA, the number concentration,
the average particle diameter, date and time and the charge state of the
battery. If no key is pressed for more than one minute and the low power
option is activated in the configuration, the display goes dark to prolong
battery life. By pressing any key, the display turns on again.
The lightning symbol indicates that the high
voltage in the charger is on and working properly.
The blinking red dot indicates that data is being
written on SD-Card.
The exclamation mark indicates that an error
occurred at some point during the current
measurement.
Color coding of LDSA values
The LDSA display is color-coded to provide a quick assessment of air
quality. LDSA values below 50 m2/cm3are displayed in green, values
above 250 m2/cm3in red, and values in between in yellow.
Figure 5: color coding of LDSA values
Otros manuales para Partector 2
2
Tabla de contenidos
Otros manuales de Instrumento de medición de Naneos
Manuales populares de Instrumento de medición de otras marcas

Endress+Hauser
Endress+Hauser Proline Promag 50 Especificaciones técnicas

Siemens
Siemens SITRANS F Coriolis FCT030 Manual de lista de piezas

KLINGER
KLINGER CMF V Series Manual de usuario

EXFO
EXFO FTB-2 Manual de operación y mantenimiento

Keysight
Keysight M8290A Manual de usuario

ADTEK
ADTEK MW-5 Manual de usuario