Microlife NEB200 Manual de usuario

NEB200
Compressor Nebulizer
Please read this instruction manual completely before operating this unit.
Caution– Federal (U.S.A.) law restricts this device to sale by, or on the order of a physician.


1
Important Product and Safety Information
Read all instructions carefully before use.
Improper care or use of the nebulizer may result in injury, damage to the unit or ineffective
treatment. Following these instructions will ensure the nebulizer’s efficacy and long life.
Nebulizer is non-sterile. Clean before use. The device is suitable for both adult and pediatric
as prescribed by a physician.
When using an electrical product, especially when children are present, basic safety precau-
tions should always be followed
DANGER:Urgent safety information for hazards that will cause serious injury or death.
WARNING:Important safety information for hazards that might cause serious injury.
CAUTION: Information for preventing damage to the product.
NOTE: Information to which you should pay special attention.
CAUTION Failure to read and observe all precautions could result in personal injury or
equipment damage.
DANGER To reduce the risk of electrocution
1. Always unplug this product immediately after using.
2. Do not use while bathing.
3. Do not place or store product where it can fall or be pulled into a tub or sink.
4. Do not place in or drop into water or other liquid.
5. Do not reach for a product that has fallen into water. Unplug immediately.

2
WARNING: To reduce the risk of burns, electrocution, fire or injury to persons:
1. The product should never be left unattended when plugged in.
2. Close supervision is necessary when this product is used by, on, or near children or disabled
individuals.
3. Use this product only for its intended use as described in this manual. Do not use attach-
ments not recommended by the manufacturer.
4. Never operate this product if:
a. It has a damaged power cord or plug.
b. It has been dropped or damaged.
c. It is not working properly.
d. It has been dropped into water.
5. Keep the power cord away from heated surfaces.
6. Never block the air openings of the product or place it on a soft surface, such as bed or
couch, where the air openings may be blocked.
7. Never use while drowsy or asleep.
8. Never drop or insert any object into any opening or hose.
9. Do not use outdoors. This product is for household use only.
10. Do not use this product with a DC to AC converter, or with any AC voltage and/or frequency
other than specified. SEE PRODUCT SPECIFICATION on this manual
11. Unplug the unit before filling the nebulizer.
12. DO NOT use this product in or near explosive environments, or where aerosol (spray)
products are being used.

3
13. DO NOT pump combustible liquids or vapors with the product or use in or near an area
where flammable or explosive liquids or vapors may exist.
14. DO NOT operate this product in an oxygen enriched environment, i.e. oxygen tent or oxygen
hood.
15. Use only in well ventilated areas. Do not expose and store the unit to direct sunlight or at
high temperature and humidity
16. Do not operate unit in a backpack or carrying bag.
17. Care should be taken when this product is used near other electronic equipment concern-
ing the possible interference between them. They may need to be moved apart.
18.
Use caution when plugging in and unplugging device. Do not allow jewelry
or other metal
items to contact plug.
19. Warning: This product contains small parts that may represent a choking hazard.
20. Do not position the equipment so that it is difficult to operate.
21. To avoid strangulation and entanglement, keep power cord and air tubes out of reach of
young children.
22. The air filter shall not be serviced or maintained while the compressor is in use.
23. Do not unplug the air tube during the treatment, otherwise may get injured because of the
effect of pressure.
24. Do not bend the air tubing of the unit where the air tube may be obstructed.
25. No modification of this equipment is allowed.
26. The device does not need to be calibrated

4
CLEANING WARNING
1. Clean all necessary parts before first use and after each treatment as instructed in this manual;
if the unit has not been used for an extended period, all parts should be cleaned prior to use
2. Disconnect the unit from the electrical outlet before cleaning
3. Never immerse the unit in water to clean as it may damage the unit
4. Do not use abrasive cleaners
5. All accessories must be cleaned and not exchanged with other patient before use to avoid
cross infect.
6. The mouthpiece should be replaced at least once per year

5
Compressor Nebulizer
Instruction Manual
Table of Contents
1. Introduction
2. Components of your compressor nebulizer
3. Using your nebulizer
3.1. Preparation
3.2. Use
4. Care and maintenance
4.1. Cleaning the nebulizer accessories
4.2. Cleaning the compressor
4.3. Changing the air filter
5. Troubleshooting
6. Limited warranty
7. Technical specifications
7.1. Warning and cautions: device symbols
7.2. Electromagnetic compatibility
7.3. Product specifications
8. How to contact us

6
1. Introduction
Please read this instruction manual completely before using your unit to ensure safe and
effective use, as well as long life of the product.
Statement of Intended Use
The aerosol generator is for the production of compressed air to operate a nebulizer for the
production of medical aerosol for respiratory disorders. The aerosol generator is used in con-
junction with a pneumatic nebulizer which nebulizes specific drugs for inhalation by a patient
and requires the order of a physician for medical use.
Compressor nebulizers are used to efficiently deliver medication in the form of a mist to the
lungs. Prescribed medicine is placed in the nebulizer and the compressor blasts air through the
liquid medication, converting it to an aerosol mist which is then inhaled. This method of de-
livery is beneficial in the treatment of asthma and other respiratory ailments, as the medicine
is administered directly to the bronchial passages of the lungs to provide quick and effective
relief. With proper use and care, your new nebulizer will provide you with many years of safe,
effective breathing therapy treatments.
CAUTION: Federal law restricts this device to sale by or on the order of a physician. Operate
this unit only as instructed by your physician and/or respiratory therapist.
Thoroughly read and understand the information contained in these instructions on the use
and care of this unit before operating. Save these instructions for future reference. Any use of
this product other than the purpose or which it is intended shall void the warranty.

7
2. Components of your compressor nebulizer
Compressor (piston compressor):
1. Compressor (piston compressor)
2. Power cord
3. ON/OFF switch
4. Air filter compartment
5. Air filter cap
6. Air filter
Nebulizer accessories:
7. Nebulizer
a. Nebulizer top
b. Nebulizer nozzle
c. Nebulizer cup
8. Air tubing
9. Mouthpiece
10. Adult face mask
11. Child face mask
12. Nose piece
Nebulizer assembly:
13. Nebulizer accessories,
assembled
5
6
7
7c
89
10
12
13
11
7a
7b
Note: All items, other than the Piston Compressor, are accessories.

8
3. Using your nebulizer
3.1.Preparation
NOTES:
• This unit is intended for single patient use and should not be shared with other people.
• Before the first use of the nebulizer or if the unit has not been used for an extended period
of time, it should be cleaned thoroughly as described in the Cleaning section of this manual.
WARNING: Before connecting the power cord, make sure the I/O (ON/OFF) switch is in the O
(OFF) position. The plug is also served to disconnect the device.
CAUTION: The Compressor is designed for intermittent use only. Do not operate it continu-
ously for more than 30 minutes for a single use without turning it off and following a cooling
period for least 30 minutes.
NOTE: the compressor pump motor has a thermal protector, which will turn the power off be-
fore the unit overheats. If the thermal protector shuts the power off, contact Customer Service
for assistance: 1-800-568-4147.