Metertech SP-870+ Guía de inicio rápido


1
Metertech Inc. provides this publication “as is”without warranty of any kind,
either express or implied, including, but not limited to the implied warranties
of merchantability or fitness for a particular purpose.
All rights are reserved, including those to reproduce this publication or parts
thereof in any form without permission in writing from Metertech Inc.
Information in this publication is subject to change without notice and does
not represent a commitment on the part of the vender.
Any errors or omissions which may have occurred in this publication despite
the utmost care taken in its production will be corrected as soon as possible,
but not necessarily immediately upon detection.
Copyright© Metertech Inc. 2005
Metertech Inc.‧63-2, ChengGong Rd., Sec. 1, NanGang, Taipei, Taiwan, ROC
Tel: 886-2-2783-2854 ‧Fax: 886-2-2783-1764

2
CONTENTS
Introduction .......................................................................................
3
Description ...................................................................................
3
Components .................................................................................
3
Keyboard and display layout .......................................................
4
Rear panel layout .........................................................................
5
Specifications ...............................................................................
6
Installation .........................................................................................
7
Standard equipment .....................................................................
7
Environmental and electrical conditions .....................................
7
Installation ...................................................................................
8
Operation ..........................................................................................
10
Powering up the equipment ........................................................
10
Absorbance and Transmittance measurements ...........................
10
Concentration measurement with a factor ...................................
11
Concentration measurement with a standard ...............................
12
System functions ..........................................................................
13
Data saving and retrieve ..............................................................
15
Printer ..........................................................................................
15
Analog output ..............................................................................
15
Maintenance ......................................................................................
16
Routine care .................................................................................
16
Replace the lamp .........................................................................
16
Messages .....................................................................................
18
Replacement parts .......................................................................
18

3
INTRODUCTION
Description
The model SP-870+ visible spectrophotometer is a digital microprocessor
controlled instrument, providing photometric absorbance, transmittance and
concentration measurements with the wavelength range of 330 to 1100 nm.
The spectrophotometer has the following unique features:
1. Two-line, 16-large character LCD display
2. Motorized optical filter wheel and grating
3. Multi-use sample compartment for square cuvettes, 0.5” & 1” round tubes
and 50 mm longpath cells
4. Printer and RS232 interface
5. Base-line data for entire wavelength range
6. 100 sets of data storage
Components
1. LCD display 4. Model name and company brand
2. Sample compartment door 5. AC adapter
3. Keyboard 6. Lamp compartment door
Figure 1: Components of SP-870+ spectrophotometer

4
Keyboard and display layout
1. A/T/C button:
Switch among absorbance, %transmittance and
concentration modes
2. λbutton:
For entering desired wavelength
3. Numerical buttons:
For input the wavelength, factor, and
concentration
4. ENTER button:
To confirm selections or entries of fields or
parameters, or to start measuring process
according to current setting
5. ESC button:
To quit from the current process, or to quit the
current screen and return to previous screen
6. BLANK button:
Set the instrument to 100% transmittance, or
zero absorbance or zero concentration
7. FUNCTION button:
Selecting and performing supporting functions
shown in the screen
8. PRINT button:
Print out the current reading on screen
9. SAVE button:
Save the current reading in memory
10. VIEW button:
Review the stored data sequentially
11. Display panel:
16 large characters by 2 lines LCD
Figure 2: Keyboard and display layout

5
Rear panel layout
1. DC fan
2. Power switch
3. DC voltage input
4. RS232 port
5. Printer port
Figure 3: Rear panel layout

6
Specifications
_______________________________________________________________
Wavelength 330 ~ 1100nm
Bandwidth pass 5nm
Wavelength accuracy +/- 1nm
Wavelength repeatability +/- 0.5nm
Photometric range -0.300 ~ 3.500Abs
Measurement mode A / %T / C
Factor 0.1 ~ 9,999.9
Concentration 0.000 ~ 99,999.99
Photometric accuracy 1% or 0.003A(from 0.000 to 2.500A),
whichever is greater by NIST SRM930e
Stability < 0.003Abs/hr at 500nm after one hour
warm-up
Baseline flatness +/- 0.005 Abs(at 400 ~ 1100nm)
Stray light < 0.1% at 340nm by NIST SRM2031
Light source 12V halogen lamp
Detector Si photodiode
Wavelength scanning speed 2500nm/min
Cuvettes 10mm square cuvette, 10 ~ 13mm round
tube, 16mm round tube(option), 50mm
rectangular cell, and 1 inch round tube
Light beam height 8.5mm at measurement position
Data storage 100 sets
Display 16 x 2 large characters LCD
Interface RS232 and parallel printer
Function Initial wavelength, lamp tuning, system
validation, resolution
PC software Data collection from saved memory; real
time data display and storage
Power 100 ~ 240V at 50/60Hz
Dimension 400(W) x 300(D) x 145(H) mm
Weight 7.5 kgs
Ambient temp/humidity room temperature 15 ~ 35℃,
humidity < 80%
_______________________________________________________________
Specifications subject to change without notice

7
INSTALLATION
Standard equipment
The package contains items listed below. Unpack the carton and inspect
carefully. If any part is damaged or missing, contact your dealer immediately.
Qty Description
1 SP-870+ spectrophotometer
1 AC adapter
1 Power cord
1 Operator’s manual
5 10mm plastic square cuvette
1 Dust cover
150mm cuvette bracket
1 1 inch tube bracket
1
Environmental and electrical conditions
The instrument has been designed to operate and meet the specifications on
previous pages under the electrical and environmental requirements listed
below.
Line voltage
100-240Vac, 50-60 Hz
Environment conditions
Ambient temperature from 15℃to 35℃and maintained at +/-2℃;
relative humidity from 20% to 80% and maintained at +/-5%.

8
Installation
1. Place the spectrophotometer on a solid and level surface in a dry, clean
environment.
2. Leave an open space at least 50mm distance from rear panel and side walls
to guarantee air flow around the instrument.
3. Connect the female end of power cord into the inlet of AC adapter, and then
the outlet plug of AC adapter into the DC input hole on the rear panel of
instrument.
4. Plug the power cord into a grounded wall outlet with the appropriate
voltage.
5. Connect the Centronics parallel printer to the instrument at the print port.
6. Turn on the power switch and allow instrument to warm up at least 30
minutes for electronic and thermal stabilization.
7. The standard cell holder is adaptable for 10 mm square cuvette and 1/2”
round tube. For 50 mm long-path cell measurement, please unscrewing
two white knobs (#1) and removing metal bracket #2 inside the sample
compartment as shown in figure 4a, then locate bracket #3 and tighten two
knobs again(figure 4b).
8. For 1” round tube measurement, orient bracket #4 and tighten two knobs
(figure 4c).
Figure 4a: Standard cell holder for 10 ~ 13 mm cuvettes

9
Figure 4b: 50 mm rectangular cell holder
Figure 4c: 1” cylindrical cell holder
Tabla de contenidos
Otros manuales de Instrumento de medición de Metertech
Manuales populares de Instrumento de medición de otras marcas

Endress+Hauser
Endress+Hauser Proline Promag 50 Especificaciones técnicas

Siemens
Siemens SITRANS F Coriolis FCT030 Manual de lista de piezas

KLINGER
KLINGER CMF V Series Manual de usuario

EXFO
EXFO FTB-2 Manual de operación y mantenimiento

Keysight
Keysight M8290A Manual de usuario

ADTEK
ADTEK MW-5 Manual de usuario