Measuring Gas Carrier Gas Basic range Smallest range Smallest suppressed zero range
H2 He 20% - 100% 20% - 40% 85% - 100%
H2 CH4 0% - 100% 0% - 0.5% 98% - 100%
H2 N2 or air 0% - 100% 0% - 0.5% 98% - 100%
H2 Ar 0% - 100% 0% - 0.4% 99% - 100%
H2 CO2 0% - 100% 0% - 0.5% 98% - 100%
He N2 or air 0% - 100% 0% - 0.8% 97% - 100%
He Ar 0% - 100% 0% - 0.5% 98% - 100%
CH4 N2 or air 0% - 100% 0% - 2% 96% - 100%
CH4 Ar 0% - 100% 0% - 1.5% 97% - 100%
O2 N2 0% - 100% 0% - 15% 85% - 100%
O2 Ar 0% - 100% 0% - 2% 97% - 100%
NH3 H2 0% - 100% 0% - 5% 95% - 100%
N2 Ar 0% - 100% 0% - 3% 97% - 100%
N2 CO2 0% - 100% 0% - 4% 96% - 100%
CO H2 0% - 100% 0% - 2% 99% - 100%
Ar N2 or air 0% - 100% 0% - 3% 96% - 100%
Ar CO2 0% - 100% - 80% - 100%
CO2 N2 or air 0% - 100% 0% - 3% 96% - 100%
CO2 Ar 0% - 100% 0% - 10% -
SF6 N2 or air 0% - 100% 0% - 2% 96% - 100%
Table 1: The measuring ranges of typical gas compositions for analysis with the FTC320-OEM.
2 FTC320-OEM Detector Unit
2.1 Description
The FTC320-OEM is a highly precise and stable Thermal Conductivity Detector (TCD) that is
designed for the use as an OEM detector. The unit consists of a hermetically sealed pressure
proof stainless steel block with a gas duct, which is suited for pressures up to 20 bar. Sample
gas entering through the gas inlet is guided to the micro-mechanical thermal conductivity
sensor and further downstream to the outlet port. In particular the pneumatics are designed to
minimize the influence of a changing gas flow. The operating temperature of 63◦C is stabilized
by a highly accurate PI control loop.
In order to avoid electrical interference on the measuring output, the high performance analog
adaption circuit is directly mounted on top of the stainless steel block. The Processor board
digitizes the signal in a 24bit A/D converter. The micro-controller performs all calculations,
such as linearization, calibrations and cross sensitivity compensation directly on the detector
unit.
For indication of the signal, a 0 to 10V analog output is provided. For internal calibration,
configuration and digital indication of the signal a TTL-level RS232 or USB communication
2is required. For these features and for testing purposes (e.g. plotting and logging the
signals), Messkonzept offers a Microsoft Windows software and/or a terminal program in order
to perform Offset and Gain calibration on a PC. For more information, see part V.
2depends, among others, on the place and way of installing the device