1.4 Safety
Preventing Misuse
Any operator of a Mentalab Explore device should read the instruction manual specific to the
device and API versions. The device is only to be operated according to the instructions.
Mentalab Explore is not intended for medical use, including any diagnostic or therapeutic
applications.
Mentalab Explore tolerates continuous exposure of [-2.7V, +2.7V] without electrical damage
and has a built-in ESD protection that protects against burst exposure:
• IEC 61000-4-2 Level 4 ESD Protection – ±12-kV Contact Discharge – ±15-kV Air Gap
Discharge
• IEC 61000-4-4 EFT Protection – 80 A (5/50 ns)
• IEC 61000-4-5 Surge Protection – 2.5 A (8/20 µs)
When using the Explore system with other hardware or implanted electrical cardiac or neural
stimulators (pacemaker, tDCS, tACS, DBS, etc), the above values should not be exceeded to
provide safe use. Using Explore with additional hardware can affect the accuracy of the
recorded signals.
Mentalab Explore is not protected against electrical defibrillation. The device and electrodes
must be detached from the subject’s skin beforehand.
Before use, inspect the device and accessories thoroughly for possible damage to
connectors, sockets, and cables. Especially, ensure that the electrode cables are intact. Any
part of the equipment must be replaced immediately if damaged or not working correctly.
The device and its accessories must not be exposed to increased mechanical stress.
Do not open or modify the device, it will void the warranty.
Using controls, adjustments, procedures, connections, or signal types other than those
specified in the Explore documentation may result in electrical and mechanical hazards.
Only use Mentalab approved accessories with this device. If an accessory is not listed, please
contact us for technical support.
Battery
While being charged or connected to a PC / tablet, the device should not be connected to
any subject.
The external power supply must fulfill the IEC/EN 60950-1 and SELV requirements to be used
safely with the device. Do not charge your device with anything other than the USB port of a