Page 3 of 8
Data Sheet CPC3000
Specications CPC 3000
Pressure ranges Gauge: 0 ... 5 up to 0 ... 1500 psig (0 ... 0.35 to 0 ... 103 bar)
Absolute: 0 ... 14.5 to 0 ... 1515 psia (0 ... 1 to 0 ... 104 bar)
Bi-directional: 5 ≤ span ≤ 1515
Pressure types Absolute, gauge or bi-directional ranges
Uncertainty 0.025% full span for all standard pressure ranges specied above. Uncertainty (k=2) includes
hysteresis, linearity, repeatability, reference standard, drift and temperature eects over the calibrated
range for the calibration interval specied, with periodic re-zeroing.
Optional Intelliscale Uncertainty
Gauge
Transducer Range (psig) Total Uncertainty (1) (cal interval)
0 ... 14.5 to 0 ... 1500 0.025% IS-50 (1) (365 days)
Absolute
Transducer Range (psia) Total Uncertainty (1) (cal interval)
0 ... 14.5 to 0 ... 1515 0.025% IS-50 (1) (365 days)
Bi-directional
Transducer Range (psi) Total Uncertainty (1) (cal interval)
-15 ... 145 to -15 ... 1500 0.025% IS-50 (1) (365 days)
5 ≤ span < 145 0.025% of Full Span (365 days)
(1) 0.025% IntelliScale-50 (0.025% IS-50): Uncertainty from Min to 50% of Max = (0.025% x 50%
x Max) or 0.025% of Reading from 50% to 100% of Max.
Optional barometer uncertainty 0.02% R from 8.5 to 17 psia for 365 days
Compensated temperature range 15 to 45 °C
Calibration interval 365 days
Pressure units: English psi, psf, oz/si, tons/sq in, tons/sq ft, atm, inHg @0C, inHg @60F, mTorr, Torr, inSW, ftSW, inH2O @4C,
inH2O @20C, inH2O @60F, ftH2O @4C, ftH2O @20C, ftH2O @60F
Pressure units: Metric mbar, bar, g/sq cm, kg/sq cm, dyn/sq cm, pascal, hPa, kPa, MPa,
mmHg @0C, cmHg @0C, mHg 0C, mSW, mmH2O @4C, cmH2O @4C, mH2O @4C, mmH2O @20C,
cmH2O @20C, mH2O @20C
Pressure units user dened 2 (multiplier from psi or Pascal)
Slew rate 3 seconds to stable ag (+/-0.025% full scale pressure) for a 10% pressure change typical into
150cc volume at pressures greater than 5 PSI. Larger volumes can lengthen this time. Controlling to
pressures less than atmosphere can lengthen this time.
Overshoot <1 % FS, in high speed mode
Control stability 0.003% of range
Pressure ports 7/16-20 Female SAE threaded ports for Measure/Control, Exhaust, Reference, and Supply.
Barometric Reference port is a hose barb.
Filter elements 40 micron lter element included in each pressure port (excluding the optional barometer and the
reference port on a gauge sensor)
Permissible pressure media Clean, dry, non-corrosive gases
Parts exposed to pressure media 6000 series aluminum, 316 SS, brass, Teon, Urethane, Silicone, RTV, Silicone grease, PVC, Epoxy, Ceramics
Overpressure protection Pressure relief valves
Instrument mounting Desk top with bezel and handle, or optional rack mount kit.
Display 7.0" color LCD
Resolution Six signicant digits
Warm-up time approx. 15 min
Digital interface IEEE-488, Ethernet, USB
Power supply 100 - 240 VAC, 50/60 Hz, 700 mA max
Maximum pressure, supply port 110 … 120 %FS
Pneumatic overpressure protection internal relief valves
Operating temperature 0 … 50 °C
Storage temperature 0 ... 70 °C
Air humidity 0 … 95 (% relative humidity without moisture condensation)
Operating position Negligible, can be removed with re-zeroing
Weight <20 lbs.(<9.1 kg) with all internal options
Dimensions 5¼ h x 8⅜ w x 12 d in (133 x 213 x 305 mm)
CE-mark Conformity certicate
Calibration NIST traceable calibration certicate included, A2LA certication