Me GSV-2MSD-DI Manual de usuario

GSV-2MSD-DI Data Logger
Instruction manual
Version: 26/08/2015
ME-Me systeme GmbH
16761 Hennigsdorf Fax: +49 3302 78620 69 Web: www.me-systeme.de


Contents
GSV-2MSD-DI Strain gage Data Logger.............................................................................................4
Connection and commissioning.....................................................................................................4
Operating the menu..........................................................................................................................4
Logger menu.................................................................................................................................4
Description of the logger menu:.................................................................................................5
Notes..........................................................................................................................................6
Measuring amplifier menu..........................................................................................................6
Description of the measuring amplifier menu....................................................................7
Notes..........................................................................................................................................8
Description of the buttons...............................................................................................................9
Number setting..................................................................................................................................9
Description of LEDs........................................................................................................................10
Maximum, minimum and mean value........................................................................................10
Recording measured data to file..................................................................................................11
USB operation..................................................................................................................................12
Power supply per USB connection...............................................................................................12
USB driver.........................................................................................................................................12
Driver installation........................................................................................................................13
Interface number / Determine COM-Port...............................................................................14
Dimensions.......................................................................................................................................17
Terminal assignment......................................................................................................................18
Connection plan for DMS bridges................................................................................................19
Technical data ................................................................................................................................20
ME-Me systeme GmbH
16761 Hennigsdorf Fax: +49 3302 78620 69 Web: www.me-systeme.de 3

GSV-2MSD-DI Strain gage Data Logger
Connection and commissioning
To switch on, press the MODE button until the display is illuminated.
The device contains a battery that is charged by connecting the power adapter supplied.
The “ON” LED flashes while charging.
To switch off, hold the MODE button (outside the menu) down and confirm the following
message “Power off? (OK)“ by pressing the OK button.
Please use SDHC Memory Cards, class 6 or class 10, but no UHS 2.
Operating the menu
The device has two main menus:
Logger menu.
The logger menu has the following functions:
•View and set data recording mode on the SD memory card,
•Select USB connection mode,
•Select display illumination characteristics,
•Display battery life,
•Set date and time,
•Select additional functions e.g. display and recording of maximum, minimum and average
measured values.
The Logger menu can be reached by pressing the MODE button from the measured value
display or via the last entry of the measuring amplifier main menu.
ME-Me systeme GmbH
16761 Hennigsdorf Fax: +49 3302 78620 69 Web: www.me-systeme.de 4

Description of the logger menu:
Menu entry Level
1
Menu entry Level 2 Menu entry Level 3
Logging
ON, OFF, onOK
➔Set Log.: ON Permanent recording of measured data
on the SD card
➔Set Log: onOK Recording of measured data on the SD
card as long as the OK button (at measured value
displayed) is pressed
➔Set Log: OFF No Recording
USBmode:
Comm, SDcard,
none
➔USBmode: SDcard The device is a Mass Storage
Device when connected to the USB port, which provides
access to the data on the SD card. Simultaneous
recording of measured value to file is not possible. If
this mode is activated, recording is switched off.
➔USBmode: Comm The device is in serial USB mode.
Our communication programs (e.g. GSVcontrol,
GSVmulti) can then be used. A standard driver is loaded
when connected for the first time, see below
➔USBmode: none USB is switched off.
Bat: level bar or
Charge (with
percentage
display)
➔USB power: ON, OFF
If power supply is via USB port is enabled, the GSV-
2MSD-DI can be supplied by the USB bus if it is
connected to a PC and configured as an USB device. If
the device is switched off, the battery can be charged.
➔Set: USBpwr ON
➔Set USBpwr OFF
Switch power supply
via USB on or off
Light
ON, Auto
➔Set: Light ON Display is permanently illuminated
➔Set: Light Auto Display is illuminated when buttons are
pressed and in the menu and goes out after 5 seconds -
this preserves the battery.
date / time ➔Date TT Mon JJJJ ➔Setting the date
➔Time HH:MM:SS ➔Setting the time
➔(see description of
numeric settings
below)
Val.mode
normal, MaxMin
➔SetMode: normal: Only displays the current measured
values
➔SetMode: MaxMin:The maximum, minimum or mean
value can be displayed in the display and all three are
ME-Me systeme GmbH
16761 Hennigsdorf Fax: +49 3302 78620 69 Web: www.me-systeme.de 5

Menu entry Level
1
Menu entry Level 2 Menu entry Level 3
recorded to file.
Menu entry Level 1 Menu entry Level 2 Menu entry Level 3 Menu entry Level 4
Logging options ➔Row elements ➔With date: Yes/No ➔With […] Yes/No
i e switching to the
setting not yet selected
➔With time: Yes/No
➔With unit: Yes/No
➔File options ➔heade: Yes/No
➔Number of lines ➔Setting the max
number of lines in a
file
➔Length of time ➔Setting the max
duration of the file
➔Directory ➔Every month
➔Every day
otes
•If a setting is in brackets in Level 1, this means that this setting has been selected
but is not currently active. This is the case if the conditions for this operating mode
are not met. Example:
•Logging (on): SD card is not inserted or write-protected or full or defective.
•USBmode: (SDcard): USB cable is not connected or no SD card is inserted
Measuring amplifier menu
The measuring amplifier menu is used to set the parameters of the measuring amplifier, for
example:
•Sensor scaling factor
•Unit
•User-definable offset
•Parameter memory
The measuring amplifier main menu can be reached by pressing the MENU button from the
measured value display.
ME-Me systeme GmbH
16761 Hennigsdorf Fax: +49 3302 78620 69 Web: www.me-systeme.de 6

Description of the measuring amplifier menu
Menu entry Level
1
Menu entry Level 2 Menu entry Level 3
Sensor config. ➔unit
➔Sensor capacity
➔Rated output
Select unit1
Numeric setting of the physical nominal value of
the sensor.2
Numeric setting of the electrical characteristic
value of the sensor.²
Strain analysis ➔Set gage factor
➔Set bridge type
➔Numeric setting of the K-factor between 0.2
and 2583
➔Full bridge: Full bridge circuit with 4 individual
DMS, all in longitudinal direction³
➔Half bridge: Half bridge circuit with 2 individual
DMS, both in longitudinal direction³
➔Quarter bridge: Quarter bridge circuit with one
DMS³
➔PR.full bridge: Full bridge circuit with 4
individual DMS, 2 in longitudinal direction and 2
in cross direction³
➔PR.half bridge: Half bridge circuit with 2
individual DMS, one in longitudinal direction and
one in cross direction³
Menu entry Level 4
Poissons ratio(only with selection of cross-
direction full bridge or cross-direction half bridge
in Level 3) Numerical value from 0 to 0.5.³
Load settings ➔default: Manufacturer’s
settings i.e. restoring
GSV-2 parameters to
default settings.
➔user 1: User-
configurable data set
no.1 i.e. loading the
parameters that were
1 Changing the unit does not change the measured value scaling!
2 Modification of the sensor measurement range or the characteristic value changes the scaling of the
measured value.
3 Modification of the DMS parameter of the strain analysis changes the scaling of the measured value and
the unit.
ME-Me systeme GmbH
16761 Hennigsdorf Fax: +49 3302 78620 69 Web: www.me-systeme.de 7

Menu entry Level
1
Menu entry Level 2 Menu entry Level 3
previously filed as user 1
with save settings.
➔...and so on, until
➔user 6: like user 1, but
No 6
Save settings Saves the current
configuration under user 1
to user 6
Set scaling Numerical value between
0.1 ...999999
Data acquisition1➔Data frequency
➔Data period
Numerical value: Num. Of values per second
Numerical value: Data period of acquired values
Note: the last confirmed with OK value is
adopted. Please confirm data period for slow
measurements, data rate for fast
measurements.
Options ➔Set channel (Set channel)
➔Set threshold (Set
threshold)
➔Offset value
➔Language 1
Numerical value 0 or 1
➔On-threshold Numerical value of the switch-on
threshold
➔Off-threshold Numerical value of the switch-off
threshold
➔Value that is added to each measured value.
➔German or English (Menu language setting)
Logger config. See logger menu.
- Press the Menu button to go back one level.
- Press the OK button to confirm an entry or to go to the next level.
- If a setting is selected with the OK button, the following message will appear “OK to
confirm“, which you can confirm by pressing OK to approve the setting - or cancel by
pressing the MENU button.
1 This menu entry is available from firmware version 1.5.08 on.
ME-Me systeme GmbH
16761 Hennigsdorf Fax: +49 3302 78620 69 Web: www.me-systeme.de 8

otes
•The symbol on the right edge of the display indicates that the menu entry is
activated.
•Access to the measuring amplifier menu will be blocked if communication is taking
place via the serial interface. In this case, the following is displayed:
“Menu blocked”
ME-Me systeme GmbH
16761 Hennigsdorf Fax: +49 3302 78620 69 Web: www.me-systeme.de 9

Description of the buttons
Button Function
MODE To switch on and off or access the Logger menu
MENU
(LEFT)
To access the measuring amplifier menu, to go one menu level higher or to cancel an
entry.
For numeric settings: to move the cursor left.
UP To move around the menu within the same level: switching to the next entry.
For measured value mode=MaxMin: to switch from one measured value display to
the next one up.
For numeric settings: to increase number and/or setting above the cursor.
DOWN To move around the menu within the same level: switching to the previous entry.
For measured value mode=MaxMin: To switch from one measured value display to
the next one down.
For numeric settings: to decrease number and/or setting above the cursor.
OK
(RIGHT)
To confirm the entry or move down one level.
For measured value display and recording mode “withOK”: to record measured values
to file.
For numeric settings: to move the cursor right.
SHORT To connect the +Ud and -Ud inputs (short-circuit of sensor signal)
ZERO To trigger an automatic zero adjustment.
umber setting
To set a numerical value and the date or time, move the cursor right by pressing OK and left
by pressing MENU. The digit (and/or the month) above the cursor flashes and can be
increased or decreased using the UP / DOWN buttons.
For the setting to take effect, the OK button must be held down until the whole number
flashes.
Then release the OK button, and the following message will appear “OK to confirm“.
Confirm this by pressing OK. To cancel the numeric setting, hold down the MENU button.
ME-Me systeme GmbH
16761 Hennigsdorf Fax: +49 3302 78620 69 Web: www.me-systeme.de 10