• Don’t immerse the On/Off or CAL buttons in water.
• Don’t replace the batteries or the Reference Electrode when the pH
meter is wet.
• Don’t press the POWER or the CAL Switches with a pointed object.
• Don’t use the pH Meter at temperatures out of the working range of
5ºC-40ºC (41ºF-104ºF). This will shorten the service life of the sen-
sor will be shortened.
• The following substances will damage the sensor :
Organic Solvent (thinner, benzene, etc.)
Strong acid (pH 0-2)
Strong alkali (pH 12-14)
Surface active agents, alcohol, oil, adhesive and cement.
• Keep the sensor away from direct sunlight or other strong light
during calibration and measurement.
• If the meter has not been used in several days, hydrate the sensor
by immersing it in water for about five minutes before calibrating.
• White crystals on the sensor are not a symptom of trouble. Simply
rinse the sensor before use.
• Although the meter is auto-temperature compensated, it is recom-
mended that, as much as possible, the sample and meter be ap-
proximately the same temperature.
• Avoid extremes of temperature. Don’t place the pH meter close to
heaters or where the meter might be exposed to direct sunlight.
• Don’t bend or drop the pH Meter.
• Standard solution is caustic acid. Wash the skin thoroughly with
water if the standard solution comes in contact with skin.
TECNOLOGÍAS EN INVERNADERO