Turning the Unit On
Press and hold the Right Key for 4 seconds, until
the LCD displays “On”, the buzzer beeps, and the
green LED is on. The unit enters a warm up and self-
test sequence, shows the firmware version, e.g.
"VER","1.0.0.0" and the programmed alarm limits,
STEL, TWA, etc.
Note: Some units may need several minutes to an
hour to stabilize and should be re-zeroed at that time.
Turning the Unit Off
In normal display mode, press and hold the Right
Key for a 5-sec. count down until the unit powers off.
Normal mode
In normal mode, the gas concentration displayed on
the LCD continuously and the unit alarms if a pre-set
limit is exceeded. The user can check various other
values like STEL, TWA, PEAK, and (for O2) MIN by
pressing the Right Key. Alarm events are also shown.
Enter Configuration Mode
Press and hold the Left and Right keys together for
3 seconds, until “PWD” is displayed, prompting
password entry. To input the password, use the Left
Key to increase the number, and the Right Key to
move the cursor. After all four digits are entered, the
cursor will move to “OK”. Use the Left Key to
accept and enter Config Mode.
Configuration Mode Menu
In Config Mode the user can do calibrations and
change various operating parameters:
⚫AiR, SPAN: Zero or Span calibration,
⚫BUMP: Bump test
⚫SET HIGH\LOW\STEL\TWA alarm limits
⚫SET Cal: Change span value
⚫CAL INTV, BUMP INTV span or bump interval
⚫SET UniT: Change concentration unit
⚫Vib SET: Vibrator enable/disable
⚫P-on Zero SET: Power on zero enable/disable
⚫FAST SET: Fast power on enable/disable
⚫RST cF9: Reset configuration to default values
⚫EXIT: Exit Config Mode
In Config Mode, in general, use the Left Key to
increase a number or confirm, and use the Right Key
to move the cursor or move to the next
programming item.
Exit Configuration Mode
Scroll to “EXIT?” and press the Left Key to exit
Config Mode and return to Normal Mode.
Daily Self-Test
Press the Left Key to confirm that the LEDs
flash, the buzzer beeps and the unit vibrates if
programmed to do so.
Alarm Limits
To adjust an alarm limit, enter Config Mode and
scroll to the desired alarm: SET HIGH\LOW\STEL\TWA?.
⚫Press Left to display the alarm value with the first
digit flashing
⚫Use Left to move to the desired flashing digit
⚫Use Right to increase the value and cycle 0-9.
⚫When finished, scroll Right to OK and press Left
to save and exit.
Zero (Fresh Air) Calibration
Zero calibration sets the base line for the sensor
and is done in fresh air or other clean air source. Enter
Config Mode and “AiR?”displays as the first menu
item. Press the Left key to start the15-second zero
calibration count-down, after which a “pass” or “fail”
result will be displayed.
To abort, press the Right key at any time during the
15-second count, and “AbRT”displays to confirm.
Span Calibration (continued)
Span calibration uses a known concentration gas
to determine the response of the sensor to the gas.
1. Attach the Calibration Adapter over the inlet
port on the front of unit by pressing it into place.
Use a fixed-flow regulator
with a flowrate of preferably
about 0.3 to 0.5 LPM. Higher flowrates can be
used if necessary, but not lower. For reactive gases,
e.g., hydrogen chloride, hydrogen fluoride, ozone,
chlorine dioxide, phosphine, ammonia and ethylene
oxide, be sure to use Teflon tubing connections.
2. Enter Config Mode and scroll to “SPAN?”
3. Start the gas flow, and press the Left key to
start the calibration countdown. The default count
time is 60 seconds, but some sensors will go longer.
4. When finished, a “pass” or “fail” result will be
displayed. Turn off the gas supply, remove the
calibration adapter, and exit to Normal Mode.
5. To abort at any time during the count-down,
press the Right Key and “AbRT”is displayed.