Lutron Electronics TM-917 Manual de usuario

PT 100 ohm, type K/J/R/E/T
PRECISION 0.01°
THERMOMETER
Model : TM-917
Your purchase of this PRECISION THERMOMETER marks a step
forward for you into the field of precision measurement.
Although this METER is a complex and delicate instrument, its
durable structure developed. Please read the following
instructions carefully and always keep this manual within easy
reach.
OPERATION MANUAL

TABLE OF CONTENTS
1. FEATURES............................................................... 1
2. SPECIFICATIONS.................................................... 1
2-1 General Specifications.......................................
.
1
2-2 Electrical Specifications......................................2
3. FRONT PANEL DESCRIPTION....................................3
3-1 Display.............................................................
.
3
3-2 Hold Button......................................................
.
3
3-3 Record Button...................................................3
3-4 Recall Button....................................................
.
3
3-5 Sensor Select Button.........................................
.
3
3-6 Rel. ( Relative ) Button...................................... 3
3-7 Power On/Off Switch......................................... 3
3-8 0.10.01Button................................................... 3
3-9 Offset VR..........................................................3
3-10 CFButton........................................................
.
3
3-11 Thermocouple Socket......................................
.
3
3-12 Pt 100 ohm Probe Socket.................................3
3-13 RS-232 Output Socket.....................................
.
3
3-14 Battery Cover/Compartment.............................3
4. MEASURING PROCEDURES....................................... 4
5. RS232 PC SERIAL INTERFACE..................................
.
5
6. REPLACEMENT OF BATTERY.....................................6
7. OPTIONAL ACCESSORIES......................................... 6

1. FEATURES
* Professional thermometer with high accuracy & 0.01
resolution.
* Accept multi type temp. probe input : platinum Pt 100
ohm, thermocouple type K/J/T/E/R.
* With 0.01 high resolution both for Platinum &
Thermocouple probe input.
* Pt 100 probe input cooperate with an 0.00385 alpha
coefficient, meet DIN IEC 751.
* Cooperate with 4 wires Pt-100 ohm probe, high precision.
* Wide range display from -100 to 1370 蚓( type k ).
* Build in 蚓& 蚌select button on the front panel.
* Build in 0.01& 0.1select switch on the front panel.
* Super large LCD with unit display, easy readout.
* Data hold function for stored the desired value on display.
* Sensor select button on the front panel, easy to change
different type probe.
* Memory function to record the maximum & minimum
reading with recall.
* Build the REL. button, useful for relative measurement.
* RS232 computer interface.
* Built-in low battery indicator.
* Optional heavy duty & compact housing case, designed
for easy carry out.
1

2
.
SPECIFICATIONS
2
-
1
G
enera
l
S
pec
ifi
ca
ti
ons
Display 62 mm x 34 mm supper large LCD
display, 13 mm ( 0.5" ) digit size.
Sensor type
6 types :
1. Platinum PT 100 ohm
( 0.00385 alpha coefficient, meet DIN
IEC 751 )
2. Thermocouple type K.
3. Thermocouple type J.
4. Thermocouple type T.
5. Thermocouple type E.
6. Thermocouple type R.
Functions 蚓, 蚌, 0.010.1 Data hold,
relative measurement.
Resolution 0.01蚓/0.1蚓, 0.01蚌/0.1蚌.
Circuit Use the software to make linearity correction
instead the traditional hardware circuit.
Probe input
Pt 100 ohm probe :
socket DIN 4 pin socket.
Thermocouple couple probe :
Standard 2 pin type K socket.
Sampling Time Pt 100 ohm probe - Approx. 1.5 sec,
Others - Approx. 2.5 sec.
Hold Function To freeze the display reading value.
Memory Recall Memorize the Maximum, Minimum
reading with recall.
Over Indication Show " - - - - ".
Data Output RS232 PC serial interface.
2

Power Supply
Heavy duty or Alkaline type,
DC 9V battery, 006P, MN1604 ( PP3 )
or equivalent.
Power Approx. DC 15 mA.
Consumption
Operating 0 to 50 蚓(32 to 122 蚌).
Temperature
Operating Less than 80% RH.
Humidity
Meter Size HWD 180x72x32 mm ( 7.1x2.8x1.3 inch ).
Meter Weight 285 g/0.62 lb.
Standard Operational manual..... 1 PC.
Accessory
Optional * Type K Thermocouple probe.....................
Accessories ...........
.
TP-01, TP-02A, TP-03, TP-04
* Pt 100 ohm probe ......................PT-100
Ref. page * Carryin
g
case....CA-06, CA-03, CA-05A
* Software .......................
.
SW-U101-WIN
* RS232 cable...........................
.
UPCB-01
3

2-2 Electrical Specifications ( 23 5 蚓)
Type Reso- Range Accuracy
lution
Platinum 0.01 蚓-199.99 to 199.99 蚓0.1% + 0.2蚓)
Pt 0.1 蚓200.0 to 850.0 蚓
100 ohm 0.01 蚌-199.99 to 392.00 蚌0.1% + 0.4蚌)
0.1 蚌392.0 to 1562.0 蚌
Thermo- 0.01 蚓-100.00 to 100.00 蚓0.1% + 0.8蚓)
couple 0.1 蚓100.0 to 1370.0 蚓
type K 0.01 蚌-148.00 to 212.00 蚌0.1% + 1.4蚌)
0.1 蚌212.0 to 2498.0 蚌
Thermo- 0.01 蚓-90.00 to 100.00 蚓0.1% + 0.8蚓)
couple 0.1 蚓100.0 to 900.0 蚓
type J 0.01 蚌-130.00 to 212.00 蚌0.1% + 1.4蚌)
0.1 蚌212.0 to 1652.0 蚌
Thermo- 0.01 蚓-100.00 to 100.00 蚓0.1% + 0.8蚓)
couple 0.1 蚓100.0 to 400.0 蚓
type T 0.01 蚌-148.00 to 212.00 蚌0.1% + 1.4蚌)
0.1 蚌212.0 to 752.0 蚌
Thermo- 0.01 蚓-80.00 to 90.00 蚓0.1% + 0.8蚓)
couple 0.1 蚓90.0 to 779.9 蚓
type E 0.01 蚌-112.00 to 194.00 蚌0.1% + 1.4蚌)
0.1 蚌194.0 to 1435.8 蚌
Thermo- 0.1 蚓0 to 600.0 蚓0.1% + 0.8蚓)
couple 1 蚓600 to 1770 蚓0.1% + 1蚓)
type R 0.1 蚌32.0 to 112.0 蚌0.1% + 1.4蚌)
1 蚌112 to 3218 蚌0.1% + 2蚌)
Remark :
* The accuracy is specified for the meter only.
* The above spec. accuracy are tested under the
environment RF Field Strength less than 3 V/M & frequency
less than the 30 MHz only.
4

3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display
3-2 Hold Button
3-3 Record Button
3-4 Recall Button
3-5 Sensor Select Button
3-6 Rel. ( Relative ) Button
3-7 Power On/Off Switch
3-8 0.10.01Button
3-9 Offset VR
3-10 CFButton
3-11 Thermocouple Socket
3-12 Pt 100 ohm Probe Socket
3-13 RS-232 Output Socket
3-14 Battery Cover/Compartment
5

4. MEASURING PROCEDURE
1)Connect the " Probe Plug " into the " Input Socket ".
* The input socket of PT-100 ohm is the " 3-12, Fig. 1 ".
* The input socket of Thermocouple type K/J/T/E/R is the
" 3-11, Fig. 1 "
2)Select the " Power Off/On Button " ( 3-7, Fig. 1 ) to " On "
position.
3)Select the sensor type ( PT-100 ohm, type K, type J, type
T, type E, type R ) by push the " Sensor Select Switch "
(3-5, Fig. 1).
4)Determine the display resolution to 0.1or 0.01by pushing
" 0.10.01Button " ( 3-8, Fig. 1 ).
* The meter can be selected the different
resolution 0.01 蚓, 0.1 蚓, 0.01 蚌, 0.1 蚌, 1 蚓,
1 蚌.
The relation of resolution/range/different sensor
type, please refer the " 2-2 Electrical
Specifications ", page 2.
* Push the " 0.10.01Button " once again, the
display will get the alternation of high resolution
and the low resolution.
6

5)
Data Hold :
During the measurement, push the " Hold Button " ( 3-2,
Fig. 1 ) will freeze the display value & LCD will show the
" HOLD " marker. Push the " Hold Button " once again
will release the data hold function.
6)
Relative Measurement
@During the measurement, the circuit will memorize the
last measured values if push the "REL. Button " ( 3-9,
Fig. 1 ) at once, then LCD will show "0" & a " REL "
marker appear on the LCD display.
@The new measured temperature values will deduct
above memorized " last measured value "
automatically.
@It will release the Relative Measurement function if
push the REL. button at once again, at same time the
"REL" marker will disappear.
Considering:
When marking the " Data Hold " & " Data Record "
measurement, the Relative function is prohibited.
7

6)
Data Record ( Max. and Min. reading )
The Data Record function displays the maximum and
minimum readings. To start the Data Record function,
press the " Record Button " ( 3-3, Fig. 1 ) once.
" REC " symbol will appear on the LCD display.
With the " REC " symbol on the display :
a)Push the " Recall Button " ( 3-4, Fig. 1 ) once, the
" Max " symbol along with the maximum value will
appear on the display.
b)Push the " Recall Button " again, the " Min " symbol along
with the minimum value will appear on the display.
c) To exit the memory record function, push the " Record
Button " once again. The display will revert back to the
current reading.
7)
Offset Value adjustment
@Caused by the environment temperature change or other
reasons.... The measuring value of the thermocouple
sensor input (type K/J/T/E/R) may exist the drift of small
digits.
@If find that the measuring values exist deviation of small
digits especially when measure the low temperature, then
can adjust the " Offset Adjust VR " to make the
compensation until get the correct reading exactly.
The offset adjustment procedures are following:
a)Plug in a type k thermocouple probe to the input
socket & Set the " Sensor Select Button " ( 3-5, Fig. 1 )
to the " Type K". Now the meter is measured the
environment Temperature value.
8
Tabla de contenidos
Otros manuales de Termómetro de Lutron Electronics

Lutron Electronics
Lutron Electronics TM-906A Manual de usuario

Lutron Electronics
Lutron Electronics TM-939 Manual de usuario

Lutron Electronics
Lutron Electronics TM-958 Manual de usuario

Lutron Electronics
Lutron Electronics TM-925 Manual de usuario

Lutron Electronics
Lutron Electronics TM-946 Manual de usuario

Lutron Electronics
Lutron Electronics TM-9017SD Manual de usuario

Lutron Electronics
Lutron Electronics TM-947SD Manual de usuario