Lovibond Tintometer PCM Manual de usuario

www.lovibond.com
Power & Communications Module
Junction Box Manual
Lovibond®Water Testing
Tintometer®Group


Contents
• 1.0 Introduction....................................................4
• 2.0 Installation.....................................................6
Unpacking .........................................................6
Mounting ..........................................................6
High Voltage Installation ...............................................8
Low Voltage Installation ..............................................12
• 3.0 Operation ................................................................ 17
Power Indicator.....................................................17
24VDC Power Switch ................................................17
Analog Outputs ....................................................17
Relay Outputs ......................................................17
USB..............................................................17
Optional RS-485 MODBUS ............................................17
• 4.0 Maintenance ............................................................. 18
Fuse Replacement ...................................................18
Cleaning ..........................................................19
• 5.0 Replacement Parts........................................................ 20
• 6.0 Specifications............................................................. 20
• 7.0 Certification .............................................................. 21
FCC Class A Notice ..................................................21
DOC Class A Notice - Avis DOC, Classe A .................................21
• 8.0 Troubleshooting .......................................................... 22
To Preserve, Protect and Improve the Quality of the Environment
Disposal of Electrical Equipment in the European Union
Because of the European Directive 2002/96/EC your electrical instrument must not be disposed of with normal
household waste!
Tintometer GmbH will dispose of your electrical instrument in a professional and environmentally responsible manner.
This service, excluding the cost of transportation is free of charge. This service only applies to electrical instruments
purchased after 13th August 2005. Send your electrical Tintometer instruments for disposal freight prepaid to your
supplier

4PCM_V1 07.2017
GB
Introduction
1.0 Introduction
The Power & Communications Module (PCM) contains the high
voltage power supply, digital and analog communicated interfaces for
Tintometer Process Sensors. It provides 24VDC, as well as
connections for two (2) 0-20 / 4-20mA analog current outputs, three
(3) digital relay outputs, and one (1) optional RS-485 MODBUS
network.
A qualified electrician trained in the installation of electrical
equipment should read and understand this manual prior to working
with this component as potential lethal hazards are present.
The manufacture does not take responsibility for any issues caused by
use of the unit not in accordance with the instructions laid out in this
manual.
Preface
Do not exceed temperatures or time spans in any case.
All warning labels must NOT be removed and should be replaced if
they become damaged or faded.
Important Information
The PCM can be mounted to a panel or to a wall.
Notes on power connection
The PCM should be permanently connected to mains power.
See local building codes for permanently connecting equipment.
Safety instructions for operation
Never open the high voltage housing while the PCM is connected to
mains power. There is a danger of electric shock and other hazards.
The high voltage area may only be opened and serviced by qualified
professionals.
Ensure proper grounding of the metal bonding plate near the bottom
of the enclosure. It must be connected to GND for proper operation.
Cleaning
Use a damp cloth with a non flammable, non corrosive detergent.

5
PCM_V1 07.2017
GB
Introduction
ATTENTION! - Indicates that important information or specific instructions need to be strictly followed;
(Information or instructions that can be found in the manual).
For professional users in the European Union:
If you wish to discard electrical and electronic equipment (EEE), please contact your dealer or supplier for
further information.
For disposal in countries outside of the European Union:
This symbol is only valid in the European Union (EU). If you wish to discard this product please contact
your local authorities or dealer and ask for the correct method of disposal.
DANGER! - A risk of severe injury or death from ELECTRICAL shock.
DANGER! - A risk of CHEMICAL injury.
WARNING! - A risk of severe burn; HOT SURFACE.
WARNING! - A risk of severe injury; PROTECTIVE EYE WEAR required.
ATTENTION! - radio wave emissions.
Guide to Symbols
Labels attached to the instrument should be strictly observed to avoid personal injury or
damage to the instrument. Refer to this document 'Guide to Symbols' for information
regarding the nature of the danger or risk before taking any action where such label is
present.
Read all instructions before using the instrument.
DANGER
A hazard exists that will result in death or severe injury if not avoided.
WARNING
A hazard exists that may result in death or severe injury if not avoided.
CAUTION
A hazard exists that may result in minor or moderate injury.
NOTICE
Important information or specific instructions need to be strictly followed..

6PCM_V1 07.2017
GB
Installation
2.0 Installation
Unpacking
Carefully inspect all items to ensure that every part on the list below is
present and no visible damage has occurred during transportation.
Store the packing material to return the unit for repair or other kinds
of transport.
The table below shows the parts included in the packing
Part List
Mounting
1. Remove the front cover from the PCM
2. Align the PCM on the wall or panel, and mark and drill 4
mounting holes
3. Install (4) M4 x 20 or similar mounting screws
Part Description Quantity
1 PCM 1
2 Instruction Manual 1

7
PCM_V1 07.2017
Figure 1: Mounting screw locations
GB
Installation
Figure 2: Mounting and product dimensions
122mm
4.80in
152mm
5.98in
170mm
6.69in
140mm
5.51in
96mm
3.78in
Figure 1:
Mounting screw locations

8PCM_V1 07.2017
DANGER
DISCONNECT POWER BEFORE SERVICING! Only a qualified installation technician
should connect AC power to the PCM
1. Remove the red thumb screw, and connectors at position J3 and J9
2. Remove the low voltage circuit board and high voltage barrier
GB
Installation
High Voltage Installation
NOTICE
A local disconnect mounted in a suitable and easily reached location must be included
in the installation of this equipment. The local disconnect must be marked as the
disconnecting device for this equipment.
NOTICE
This unit was tested with metallic conduit as per UL 61010-1 with a bending moment of
34Nm (300 lb-in) at a horizontal distance of 300mm (12"). Do not exceed this bending
moment

9
PCM_V1 07.2017
Figure 3: Red thumb screw, J3 and J9 locations
Figure 4: Remove high voltage barrier
GB
Installation

10 PCM_V1 07.2017
3. Remove the appropriate hole plugs from the bonding plate. The two
near the back are intended for high voltage use.
4. Connect AC power to the connector labeled J1 (see Figure 5 for
power connector location). AC power and protective earth wires
must be 12 to 18 AWG. A local disconnect switch or circuit breaker
which meets local electrical code is also required on the AC power
input, and must be located near the PCM.
5. Connect Relay outputs to connectors labeled J2, J3, and J4. Note that
Normally Open (NO) and Normally Closed (NC) contacts are provided/
See Figure 6 for wiring information.
6. Reinstall the low voltage board, connectors at position J3 and J9, and
the red thumb screw
GB
Installation
High Voltage Installation
DANGER
ELECTRIC SHOCK HAZARD. The barrier must stay in place except when a qualified
installation technician is connecting power or relay outputs
CAUTION
ENSURE PROPER GROUNDING. Connect metal bonding plate to protective earth
for proper operation
WARNING
Potential fire hazard. The relay contacts are not fused and rated at 5A max. External
devices should have a current limiting device to keep the current below 5A.
Tabla de contenidos
Otros manuales de Unidad de control de Lovibond
Manuales populares de Unidad de control de otras marcas

Festo
Festo Compact Performance CP-FB6-E Manual de lista de piezas

Elo TouchSystems
Elo TouchSystems DMS-SA19P-EXTME Manual de usuario

JS Automation
JS Automation MPC3034A Manual de usuario

JAUDT
JAUDT SW GII 6406 Series Guía rápida

Spektrum
Spektrum Air Module System Manual de usuario

BOC Edwards
BOC Edwards Q Series Manual de usuario