Laserpoint s.r.l. -Plus 2 - Quick Start Guide - All Rights Reserved 3
2.2.
C
ONNECTING A SENSOR HEAD TO
P
LUS
2
M
ETER
Insert the sensor head IIS Connector
(Intelligent Identification System) into
the socket marked "Head Input" on
Plus 2 top side (Figure 2).
Figure 2
Top Side View
2.3.
S
WITCHING
P
LUS
2
M
ETER
ON
/
OFF:
To switch Plus 2 Meter on keep the white on-off button on the instrument top side (Figure 2)
pressed until the screen lights on as shown in Figure 1.
Note: If the instrument is powered by its battery the sensor head has to be plugged in before switching
the unit on to avoid the "NO HEAD" blinking notice appearing on the display and the consequent
display switch off occurring in few seconds.
To turn Plus 2 off press the on-off button for few seconds until the display goes off. The current
Plus 2 settings are automatically saved before the instrument switches off.
2.4.
S
ETTING A LASER WAVELENGTH
:
a) Touch the Wavelength screen button
g) in Figure 1 to open the “edit lambda”
window (Figure 3).
b) Select a wavelength from the list
shown at the left hand side of the window or
touch an “edit nm” button to input a specific
wavelength not shown in the list.
c) A keyboard pops up (Figure 4) allowing
the wavelength selection. The selected
wavelength value must fall in the range
shown at the top of the wavelength menu.
Figure 3 Figure 4
Note: The available wavelengths and editable wavelength range depend on the Sensor head
characteristics and calibration. There is no choice but the list for wavelengths beyond 2100nm.