Knick Protos II 4400 Manual de usuario

The Art of Measuring.
Aktuelle Produktinformation: www.knick.de
Analysenmesssystem
Protos 3400(X)
Betriebsanleitung Messmodul Protos CONDI 3400 (X)-051
zur Leitfähigkeitsmessung mit induktiven Sensoren
Deutsch
Protos COND 3400(X)-041 Measuring Module
For Conductivity Measurement with
2- or 4-Electrode Sensors
User Manual
Protos II 4400(X) / Protos 3400(X)
Process Analysis System
Latest Product Information: www.knick.de

Trademarks
The following trademarks are used in this document without further marking:
Calimatic®, Protos®, Sensocheck®, Sensoface®, Unical®, VariPower®, Ceramat®, SensoGate®
are registered trademarks of Knick Elektronische Messgeräte GmbH & Co. KG, Germany
Memosens®
is a registered trademark of
Endress+Hauser Conducta GmbH & Co. KG, Germany
Knick Elektronische Messgeräte GmbH & Co. KG, Germany
Returns
Please contact our Service Team before returning a defective device. Ship the cleaned device to the
address you have been given.
If the device has been in contact with process medium, it must be decontaminated/disinfected before
shipment. In this case, place a Declaration of Contamination in the consignment to prevent any risk to the
health and safety of our service personnel. The declaration is available at:
https://www.knick-international.com/en/service/repairs/
Disposal
Please observe the applicable local or national regulations concerning the disposal of “waste electrical and
electronic equipment”.

3
Table of Contents
Protos COND 3400(X)-041 Module
Returns........................................................................................................................................... 2
Disposal ......................................................................................................................................... 2
Trademarks ................................................................................................................................... 2
Intended Use.................................................................................................... 5
Safety Instructions .......................................................................................... 6
Operation in Explosive Atmospheres: COND 3400X-041 Module ........................... 6
Firmware Version ............................................................................................ 7
Terminal Plate COND 3400-041 Module........................................................ 9
Installing the Module.................................................................................... 10
Wiring Examples............................................................................................ 11
Calibration / Adjustment .............................................................................. 19
Temperature Compensation ................................................................................................21
HOLD Function During Calibration ...................................................................................22
Automatic with Standard Calibration Solution.............................................................24
Manual Entry of Calibration Solution ...............................................................................26
Product Calibration .................................................................................................................28
Data Entry of Premeasured Sensors..................................................................................30
Sensor Calibration....................................................................................................................31
Temp Probe Adjustment........................................................................................................32
Parameter Setting ......................................................................................... 33
Parameter Setting: Operating Levels ................................................................................34
Parameter Setting: Locking a Function ...........................................................................35
Activating Parameter Setting ..............................................................................................36
Default Settings and Selection Range..............................................................................37
TC Process Medium ................................................................................................................38
Parameter Setting: Concentration Curves ......................................................................39
Concentration Table................................................................................................................43
USP Function..............................................................................................................................44
Calculation Blocks....................................................................................................................45
pH Value Calculation...............................................................................................................48
Logbook.......................................................................................................................................50
Restore Factory Settings........................................................................................................50

4
Table of Contents
Protos COND 3400(X)-041 Module
Messages: Default settings and selection range..........................................................51
Configuring the Current Output.........................................................................................53
Current Outputs: Characteristics ........................................................................................54
Current Outputs: Output Filter............................................................................................56
Maintenance .................................................................................................. 57
Sensor Monitor..........................................................................................................................57
Temp Probe Adjustment........................................................................................................57
Diagnostic Functions .................................................................................... 58
Point of Meas Description.....................................................................................................58
Logbook.......................................................................................................................................58
Device Description ..................................................................................................................59
FRONT Module..........................................................................................................................59
BASE Module..............................................................................................................................59
Module Diagnostics.................................................................................................................60
Sensor Monitor..........................................................................................................................60
Cal Record...................................................................................................................................60
Setting Diagnostic Messages as Favorite ........................................................................62
Message list ...............................................................................................................................64
Specifications................................................................................................. 71
Appendix........................................................................................................ 75
Minimum Spans for Current Outputs...............................................................................75
Overview ........................................................................................................ 76
Overview of Parameter Setting...........................................................................................76
Index............................................................................................................... 80

5
Intended Use
The module is an input module for conductivity measurement with commer-
cially available 2- or 4-electrode sensors.
The COND 3400X-041 module is intended for operation in locations subject to
explosion hazards which require equipment of Group II, device category 2(1),
gas/dust.

6
Safety Instructions
Operation in Explosive Atmospheres: COND 3400X-041 Module
The module is approved for operation in explosive atmospheres.
When installing the product in a hazardous location, observe the information
in the supplements to the certificates and, if applicable, the relevant control
drawings.
Observe all applicable local and national codes and standards for the instal-
lation of electrical equipment in explosive atmospheres. For orientation,
please refer to IEC 60079-14, EU directives 2014/34/EU and 1999/92/EC (ATEX),
NFPA70 (NEC), ANSI/ISA-RP12.06.01.
WARNING! Possible impairment of explosion protection.
• Modules which have already been used shall be subjected to a professional
routine test before they may be operated in another type of protection.
• Prior to commissioning, the operating company must verify the intrinsic
safety in accordance with the installation regulations of IEC60079-14 for
the complete interconnection of all equipment involved, including the
connecting cables.
• The interconnection of Ex and non-Ex modules (mixed assembly) is not
permitted.
• In hazardous locations the device shall only be cleaned with a damp cloth to
prevent electrostatic charging.
Maintenance
The Protos modules cannot be repaired by the user. For inquiries regarding
module repair, please contact Knick Elektronische Messgeräte GmbH & Co. KG
at www.knick.de.

7
Menu Display Device description
Device hardware and firmware
version
Provides information on all modules
installed: Module type and function,
serial number, hardware and firmware
version, and device options.
Select the different modules (FRONT,
BASE, slots 1 - 3) using the arrow keys.
Device description
Return
22.7 °C
1.225 mS/cm
Operating panel Protos
Module FRONT 3400-011
Hardware: 1.1, Software: A.2
Serial number: 08150815
BASE
Module FRONT
Query Current Device Firmware/Module Firmware
When the device is in measuring mode:
Press menu key, open Diagnostics menu: Device Description
Options
Device description
Return
22.7 °C
1.225 mS/cm
Input for 2/4-electr. sensors and °C
Module COND 3400-041
Hardware: 1, Software: 2.0
Serial number: 471101147
BASE
Module FRONT
Options
Query module software
Module COND 3400-041,
hardware and firmware version,
serial number –
here installed in slot3.
Note: The display may vary depending on the device version.
Firmware Version
Module compatibility COND 3400-041 COND 3400X-041
Protos 3400 from FRONT firmware version 1.0 x
Protos 3400X from FRONT firmware version 4.0 x
Protos II 4400 from FRONT firmware version 1.0.0 x
Protos II 4400X from FRONT firmware version 1.0.0 x
COND 3400(X)-041 module firmware: Firmware version 2.x

8

9
Terminal Plate COND 3400-041 Module
Attaching the Terminal Plates
The terminal plates of the lower modules can
be sticked to the inner side of the door. This
facilitates maintenance and service.

10
Installing the Module
CAUTION! Electrostatic discharge (ESD).
The modules' signal inputs are sensitive to electrostatic discharge.
Take measures to protect against ESD before inserting the module and wiring
the inputs.
Note: Strip the insulation from the wires using a suitable tool to prevent dam-
age.
1) Switch off the power supply to the device.
2) Open the device (loosen the 4 screws on the front).
3) Plug the module into the slot (D-SUB connector),
see figure.
4) Tighten the module's fastening screws.
5) Connect the sensor and separate temperature probe if necessary, see
"Wiring Examples".
6) Check whether all connections are correctly wired.
7) Close the device by tightening the screws on the front.
8) Switch on the power supply.
CAUTION! Incorrect measurement results.
Incorrect parameter setting, calibration or adjustment may result in incorrect
measurements being recorded. Protos must therefore be commissioned by a
system specialist, all its parameters must be set, and it must be fully adjusted.
Otros manuales para Protos II 4400
7
Este manual sirve para los siguientes modelos
5
Tabla de contenidos
Otros manuales de Instrumento analítico de Knick
Manuales populares de Instrumento analítico de otras marcas

EXTOL PREMIUM
EXTOL PREMIUM 8820043 Instrucciones de instalación y funcionamiento

RIDGID
RIDGID SeeSnake Compact Manual de usuario

Teledyne
Teledyne 3000PA Manual de usuario

YDLIDAR
YDLIDAR TIA Manual de usuario

Olympus
Olympus IPLEX FX Series Manual de usuario

Bosch
Bosch GLM 50 C Professional Manual de usuario