Kaye LTR-150 Manual de usuario

Kaye LTR-150
Dry Block and Liquid Bath
All-in-One Temperature Calibrator
Operating Manual
Validation
M5101-EN Rev. B
May 2018

2| Kaye LTR-150 - Dry Block & Liquid Bath All-in-One Temperature Calibrator
Table of Contents
Chapter 1. Product Overview.................................................................................... 4
Key Components of the LTR-150 .............................................................................................5
1.1 Dry-Block Function........................................................................................................6
1.2 Micro Bath Function......................................................................................................7
1.3 Surface Temperature Function......................................................................................8
1.4 Calibration Modes .........................................................................................................8
1.5 Contact Kaye Customer Service...................................................................................9
1.6 Calibration and Repair Information ...............................................................................9
Chapter 2. Safety Information................................................................................. 10
General safety instructions:.....................................................................................................10
Special Safety Instructions:.....................................................................................................11
Micro Bath Calibration Safety Instructions..............................................................................11
Silicone oil ...............................................................................................................................11
Chapter 3. Specifications ........................................................................................ 12
Chapter 4. Getting Started....................................................................................... 13
4.1 Unpack and Inspect ....................................................................................................13
4.2 Placement....................................................................................................................13
4.3 Connect to Main Power...............................................................................................14
4.4 LTR-150 Startup.................................................................................................... 14-15
4.4.1 Set Screen Brightness.................................................................................................16
4.4.2 Set Language Selection ..............................................................................................16
4.4 3 Set Date and Time.......................................................................................................16
4.4.4 Set Temperature Units ................................................................................................16
4.4.5 Set Decimal Resolution ...............................................................................................16
4.4 6 Switching Modes (Remote Control / Standalone).......................................................17
4.4.7 Powering Off................................................................................................................17
Chapter 5. Preparing for Calibration...................................................................... 18
5.1 Selecting LTR-150 Function .......................................................................................18
5.1.1 Dry Block .....................................................................................................................19
5.1.2 Thermocouple Fixture..................................................................................................20
5.1.3 Micro Bath Tub Kit ................................................................................................ 21-23
5.1.4 Surface Temperature Insert.........................................................................................24
5.2 Setting Calibration Function .................................................................................. 24-25

3
M5101-EN Rev. B |
Chapter 6. Automatic Calibration Mode................................................................. 26
6.1 Starting Automatic Calibration ............................................................................. 26-28
Chapter 7. Standalone Manual Mode .................................................................... 29
7.1 Standalone Main Screen ....................................................................................... 29-30
7.2 Toolbars and Icons......................................................................................................31
7.3 Selecting Functions.....................................................................................................32
7.4 Configuration Range Settings .....................................................................................32
7.5 Test Tasks ............................................................................................................. 33-36
Chapter 8. Starting Standalone Manual Calibration ............................................ 36
Chapter 9.Maintenance and Troubleshooting....................................................... 40
9.1 Introduction .................................................................................................................40
9.2 Maintenance ................................................................................................................40
9.2.1 Dry Block .....................................................................................................................40
9.2.2 Micro Bath ...................................................................................................................40
9.2.3 Surface Temperature...................................................................................................40
9.2.4 Main Unit .....................................................................................................................41
9.3 Troubleshooting...........................................................................................................41
Appendix A. Warrenty and Return Policy .............................................................. 42

4| Kaye LTR-150 - Dry Block & Liquid Bath All-in-One Temperature Calibrator
Chapter 1. Product Overview
The Kaye LTR-150 is the next generation, multi-purpose calibrator designed to address the flexibility and
capacity needs for validation calibrations and verifications. As a multi-purpose calibrator, the LTR-150 can
be easily transformed from a standard dry block to a micro bath or surface temperature calibrator, which are
purchasable options. The LTR-150 offers a broad temperature range from -30°C to 150°C with temperature
stability of ±0.01°C and uniformity of ±0.1°C. Performance of calibration and calibration verification can either
be performed automatically via Remote mode or manual through Standard mode.
The Kaye LTR-150 has been designed to interface directly
with all Kaye products (Validator AVS, Validator 2000, IRTD,
ValProbe, and RF ValProbe). The firmware is backward
compatible with all version of Kaye software without need
for software upgrades. This functionality allows the unit to
be utilized for fully automatic calibrations/verifications of
thermocouples, ValProbes or RF ValProbes utilizing your
existing IRTD and applicable Kaye software.
The Kaye LTR-150 also has the flexibility to calibrate other
process or validation sensors in Standalone mode through
the intuitive 7” touchscreen, and a variety of optional inserts.
To maximum accuracy of the control and measurement
during calibration, the LTR-150 includes an external Control
Reference Probe. This probe is used for feedback to the
calibrator controller.
The calibrator is portable, rugged and can be used in a
variety of environments from 0°C to 50°C and humidity ≤80%
(non-condensing).
Chapter 1. Product Overview

5
M5101-EN Rev. B |
Key Components of the LTR-150
1. Tank (Dry Block, Micro Bath, Surface).
2. Carrying handle.
3. Steel housing.
4. Touch screen for display and operation.
5. Calibrator exhaust ventilation
6. Connection for external reference probe.
7. USB Port for connection to Kaye hardware.
8. Main switch with fuse and mains plug socket.
Calibrator, bottom view:
• Calibrator ventilation
Inlet air for housing cooling
Inlet air for tank / metal block cooling
b
c
Chapter 1. Product Overview
b
c

6| Kaye LTR-150 - Dry Block & Liquid Bath All-in-One Temperature Calibrator
1.1 Dry-Block Function
The LTR-150 comes standard with a dry block insert containing sixteen (16) wells for thermocouples as
well as two (2) wells for IRTD, and one (1) well for the Control Reference Probe. Each of the sixteen (16)
thermocouple wells is included with a sleeve which can accommodate upwards of three (3) thermocouples.
The sleeves are optimized to handle 22AWG thermocouples, which are similar to the ones currently utilized
in existing Kaye dry block models.
Kaye also offers additional dry blocks for calibrating specialty process or validation probes. These blocks are
easily interchangeable utilizing the supplied insert tool.
Thermocouple insert wells
IRTD insert wells
Control Reference Probe well
Chapter 1. Product Overview

7
M5101-EN Rev. B |
1.2 Micro Bath Function
The LTR-150 has the ability to be utilized as a micro bath for calibration/verification of specialty probes or
even ValProbe Loggers. Utilization of the micro bath functionality requires the purchase of the optional Micro
Bath Kit (PN: 425-3061).
The kit includes a micro bath tub as well as one (1) sensor cage, two (2) magnetic stirrers, one (1) magnetic
lifter, one (1) sleeve exchange tool, one (1) work cover, and one (1) transportation cover. Kaye also offers
Silicon Oil (type 200 -10cs) which can cover the complete temperature range of the calibrator and can be
directly poured into the micro bath tub.
The use of calibration liquids offers certain
advantages if temperatures of unusual shape
and/or size require calibration. The sensor cage
can accommodate up to two (2) ValProbe rigid
loggers. The test item is immersed directly into
the desired liquid medium without an insulating
air gap, resulting in direct temperature contact
between the unit and test item(s). The continuous
adjustment of the magnetic stirrer together with
the removable sensor basket agitates the medium
to create a broad and uniform temperature
measuring zone. Furthermore, the sensor basket
guarantees unhindered stirring, protecting the
tank floor.
The sensor cover reduces evaporation of the liquid to a
minimum and cooling of the liquid surface. The cover is
screwed on to the micro bath and has five (5) openings for
sensors under test.
Chapter 1. Product Overview

8| Kaye LTR-150 - Dry Block & Liquid Bath All-in-One Temperature Calibrator
1.3 Surface Temperature Function
The LTR-150 can also be used for surface temperature calibrations
by utilizing the optional surface temperature insert (PN: 412-3054).
The surface insert is hollow from the bottom and longer than the
adapter sleeve enabling fast and easy calibrations.
The surface temperature insert has a bored hole where the
control reference probe can be inserted which is directly beneath
the contact surface of the insert. This enables very accurate
measurement of the surface temperature. The insert is designed
for optimized homogeneity of the contact surface’s center while
enabling solid thermal contact.
The surface temperature insert is ideal for calibrating / verifying Flat Surface
ValProbes in the temperature range from -25°C to 150°C.
1.4 Calibration Modes
The LTR-150 was designed to offer flexibility and ease of use. To
accommodate these needs Kaye implemented two main modes or menu’s Remote Control mode and
Standalone mode.
Remote Control mode is intended for
Automatic Calibration utilizing Kaye
Validator equipment (Validator 2000,
Validator AVS, ValProbe and RF ValProbe).
In this mode, control of the LTR-150 and
the touchscreen interface is automatically
driven by the appropriate Kaye software.
In the Kaye software, the user configures
the set points, the stability criteria, and
deviation criteria. The user simply starts
the calibration from the Kaye software
and the control of the calibration is
automated. No additional interaction with
the touchscreen is required. The LTR-150
when powered on automatically boots
into the Remote Control screen.
The Standalone mode is intended again
to offer flexibility by allowing fully manual
configuration of the LTR-150 to define all
different kinds of parameters from different
types of sensors, ramp and soaks setups /
profiles to address other calibration needs.
For Standalone mode, refer to Chapter 7.
Manual Calibration and Chapter 8. Controls
for details the options and screen flows to
manually configure the LTR-150.
Chapter 1. Product Overview

9
M5101-EN Rev. B |
1.5 Contact Kaye Customer Service
You may contact our North American customer service team at (814) 834-9140 (FAX: (814) 781-7969). You
may also call our European headquarters in Germany at +49 (0) 7231-14335 0.
1.6 Calibration and Repair Information
To schedule and send the LTR-150 to Kaye for calibration or repair:
1. Contact the Kaye Service Center in your area to schedule the calibration or repair.
2. Utilize original packing box and carefully pack and secure the LTR-150.
3. Send the LTR-150 to the Service Center.
Chapter 1. Product Overview

10 | Kaye LTR-150 - Dry Block & Liquid Bath All-in-One Temperature Calibrator
Chapter 2. Safety Information
WARNING! A Warning identifies conditions and procedures that are dangerous to the user.
CAUTION! A Caution identifies conditions and procedures that can cause damage to the product or
equipment under test.
See Table 1 below for list of symbols used in this manual and on the LTR-150.
Table 1: Hazard Signs and Symbols
Before you install the LTR-150, read through this operating manual carefully. If the instructions
contained within it are not followed, the safety guidelines, this could result in danger for people,
the environment, and the device and the system it is connected to.
To guarantee that the device operates safely, the operator must act competently and be conscious of safety
issues.
General safety instructions:
In all work, the existing national regulations for accident prevention and safety in the workplace must be complied
with. Any internal regulations of the operator must also be complied with, even if these are not mentioned in this
manual.
Ensure that the complete operating instructions are always available in excellent condition at the calibrator installation
site.
Degree of protection per EN 60529: Ensure that the ambient conditions at the site of use does not exceed the
requirements for the stated protection rating.
Structural safety in accordance with EN 61010-1: The calibrator must be installed in such a way that the
requirements for structural safety are met.
Only use the LTR-150 if it is in perfect condition. Damaged or faulty devices must be checked without delay and, if
necessary, replaced. If problems cannot be cleared, immediately shut down the calibrator and ensure that it
cannot be started up accidentally.
Chapter 2. Safety Information
Tabla de contenidos
Manuales populares de Instrumento de medición de otras marcas

Endress+Hauser
Endress+Hauser Proline Promag 50 Especificaciones técnicas

Siemens
Siemens SITRANS F Coriolis FCT030 Manual de lista de piezas

KLINGER
KLINGER CMF V Series Manual de usuario

EXFO
EXFO FTB-2 Manual de operación y mantenimiento

Keysight
Keysight M8290A Manual de usuario

ADTEK
ADTEK MW-5 Manual de usuario