8
1. Connect the AC line to the rear of the instrument. The model
6308CT can be used with 115 or 230 VAC 50/60 Hz. Power
consumption is 6 watts. Make sure the EARTH connector is
connected to the earth lead of the AC power line.
2. Connect the proper load to the output relays. Make sure that the
load does not exceed the relay rating, 5 Amp at 115 VAC and 2.5
Amp at 230 VAC.
3. Set the proper load to the 4-20mA output connector. Make sure
that the load impedance is less than 500 Ohms.
4. If the user uses a two-wire conductivity probe, the user needs to
short terminal D1 with terminal S1 and terminal D2 with terminal
S2.
CAUTION: Make sure you connect the AC power cord to the
correct AC terminals. Connecting incorrectly may damage the unit
permanently.
D. TURNING ON/OFF THE INSTRUMENT
By just plugging the unit to a
correct AC voltage the unit will be
ready for use. There is no Power
key so unplugging and plugging
the unit will turn OFF or turn ON
the unit respectively.
After the unit is turned on, it will
perform some basic self diagnostics
and will display “OK” or “BAD”. If
you received any “BAD” messages turn OFF the unit and turn it ON
again.(see VI. ERROR DISPLAYS AND TROUBLESHOOTING ).
If the message persists then you might need to call your distributor.
(See X. WARRANTY).
After the self-diagnostic is complete the temperature will be
displayed on the lowest part of the screen and you are ready to
make conductivity or TDS measurements. Just immerse the probes