Danfoss AKS 41 Liquid Level Transmitter
Publication 9-45 Section 9
Page 10 of 18 Issue 1 : 10/11
6.2.1. Calibration for Minimum Liquid Level
(a) Bring the liquid refrigerant to the level corresponding to the lower
limit of the measuring range. i.e. the point on the rod where the
transmitter’s output signal is to be 4 mA. The full length of the rod
should be used if possible, so calibrate the lower limit with the
liquid level below the end of the rod.
(b) Press the calibration push-button and keep it depressed for
approximately 5 seconds, until the green LED stops ‘flashing’, then
release.
(c) Within the next 10 seconds, press the calibration push-button
once. After a few seconds the green LED illuminates begins to
‘flash’ slowly. The new minimum calibration has been saved to
memory.
NOTE: if the calibration push-button is not pressed
within 10 seconds, the AKS 41/AKS 41U will
automatically leave calibration mode and return to
normal operation.
The minimum output is now set to 4 mA and the AKS 41/AKS 41U
is operating normally.
6.2.2. Calibration for Maximum Liquid Level
(a) Bring the liquid refrigerant to the level corresponding to the
maximum limit of the measuring range. i.e. the point on the rod
where the transmitter’s output signal is to be 20 mA. The full
length of the rod should be used if possible, so calibrate the upper
limit with the liquid level 43 mm below the top of the rod (measured
from the gasket’s contact face).
(b) Press the calibration push-button and keep it depressed for
approximately 5 seconds, until the green LED stops ‘flashing’, then
release.
(c) Within the next 10 seconds, press the calibration push-button twice
with a 1 second interval between each press. After a few seconds
the green LED illuminates then begins to ‘flash’ slowly. The new
maximum calibration has been saved to memory.
NOTE: if the calibration push-button is not pressed
within 10 seconds, the AKS 41/AKS 41U will
automatically leave calibration mode and return to
normal operation.
The maximum output is now set to 20 mA and the AKS 41/AKS
41U is operating normally.
6.3. Calibration for Minimum and Maximum Points Other Than 4 mA and
20 mA
The usual minimum and maximum calibration points are 4 mA and
20 mA, however, the transmitter can be calibrated at other points within
the signal range. This facility is especially useful if it is impractical to
bring the liquid level to the maximum and minimum limits.
6.3.1. Minimum Calibration When Minimum Liquid Level Different From
4 mA
(a) Dismantle the cap on the DIN connecting plug and pull it back a
little on the cable to gain access to the four screw terminals inside
the plug where the four wires are connected. Connect a meter
between terminals 2 and 3 so that the 4 to 20 mA output signal can
be measured.
(b) Bring the refrigerant liquid level to the desired minimum level.
(c) Press the calibration push-button and keep it depressed for
approximately 5 seconds, until the green LED stops ‘flashing’, then
release.