2Instantel
Calibration
Note
: The Linear Microphone must be sent to a manufacturer certied calibration facility, or Instantel headquarters
for calibration. Please visit our calibration facility locator at www.instantel.com. Instantel recommends calibrating
all air overpressure microphones once a year and provides the Next Calibration date on the unit and sensors.
The calibration process is the only method to ensure the instrument is performing as it was designed and
accurately measuring the air overpressure it receives.
Using a manufacturer certied facility ensures the proper procedures are followed when calibrating your
microphone. This removes any uncertainty associated with the reliability and measurement accuracy of
your microphone. It also helps ensure the integrity of the measured recorded events if they are ever called
upon as evidence in a court of law. Using unauthorized facilities may have negative legal consequences if the
reference equipment is not traceable, the proper procedures are not followed or if the certication documents
are incomplete. Please visit our calibration facility locator at instantel.com.
Specications
Field Values
Response Standard ISEE - 2017
Frequency Weighting Linear
Frequency Range 2 Hz - 250 Hz
Amplitude Range Up to 500 Pa (0.0725 psi, or 148 dB)
Trigger Levels 2 to 500 Pa (0.00029 to 0.0725 psi, or 100 dB(L) to 148 dB(L))
Sample Rates 1,024 - 4,096 S/s
Resolution 0.0156 Pa (0.05 dB)
Cable Length 2 m (6.5 ft)
Maximum Cable Length 75 m (250 ft)
Temperature Range -10° to 50°C (14° to 122°F)
Minimate Pro Setup
The four available record modes are:
1. Waveform (Compliance Setup: records up to 9,000 seconds at 512, 1024, 2048 S/s, 26,100 seconds at 4096 S/s,
Advanced Setup: records up to 12,300 seconds at 8,192 S/s, 6,190 seconds at 16,384 S/s,
3,180 seconds at 32,768 S/s, and 125 seconds at 65,536 S/s once it has been triggered)
2. Waveform Manual (can record up to 24 hours long at 1,024 S/s with extended memory).
3. Histogram (2 sec. to 1 hour recording intervals).
4. Histogram-Combo records both Waveform and Histogram components of an event.