5
marked end should be inserted so it faces the top of the instrument. Apply upward force to seat
the battery holder properly, and when fully inserted, push to the rear to have the battery holder
sit on a ledge in the handle (Fig. 5). Replace the battery door (Fig.6). With the software currently
on the system, the unit will immediately boot, as indicated by the LCD illuminating and the LED
above the LCD turning red and then green when energized.
Note: placing the battery holder in with the incorrect orientation will not damage the system. It
simply will not start.
Figure 4 Figure 5 Figure 6
Taking a Measurement
1) When the system has completed its boot up sequence, the home screen will display an
Android application called Pb200i. To access the application:
a. Swipe the “lock”icon on the home screen to the right (Fig. 7).
b. Select the Pb200i application by touching the app on the LCD (Fig. 8).
c. Enter the password assigned by the Compliance or Safety officer using the
keyboard provided (Fig. 9).
d. Accept the acknowledgment that you understand that the system produces
radiation when the lights are on (Fig. 10).
e. Select “Test”(Fig.11).
Figure 7 Figure 8 Figure 9