SINGLE LOOP CONTROL
6 of 17
SETTING THE CONTROL
TYPE
Simultaneously press both the Up-arrow Sand
the Down-arrow Tkeys for approximately 3
seconds until the Input Menu prompt InPt appears
in the upper display and the Setup Menu prompt
SEt appears in the lower display:
InPt / SEt
•Press the Up-arrow Skey twice. 9LbL appears
in the upper display:
9LbL/ SEt
•Press the MODE key four times. Cntl will appear
in the lower display.
•Press the Up-arrow Skey until nor appears in the
upper display.
•Press the DSPY key to return to the main display
format.
SETTING THE UNITS Make sure that this step is performed prior to setting the
low or high temperature limits.
•Simultaneously press both the Up-arrow Sand
the Down-arrow Tkeys for approximately 3
seconds until the Input Menu prompt InPt appears
in the upper display and the Setup Menu prompt
SEt appears in the lower display:
InPt / SEt
•Press the Up-arrow Skey twice. 9LbL appears
in the upper display:
9LbL/ SEt
•Press the MODE key once. C_F will appear in the
lower display.
•Either Ffor Fahrenheit units or Cfor Centigrade
units appears in the upper display.
•If the incorrect units are shown, press the Up-
arrow Sonce to change to the temperature units.
•Press the DSPY key to return to the main display
format.
SETTING THE LOW
TEMPERATURE LIMIT For routine operation, the low temperature limit for
thermocouple 1 (TC1) should be set to -200ºC, as follows:
•Make sure that the units have been set to
Centigrade, as described above.
•Simultaneously press both the Up-arrow Sand
the Down-arrow Tkeys for approximately 3
seconds until the Input Menu prompt InPt appears
in the upper display and the Setup Menu prompt
SEt appears in the lower display:
InPt / SEt
•Press the MODE key twice. rL1 will appear in the
lower display.
•Use the Down-arrow Tkey to set the upper
display to -200.
•Press the DSPY key to return to the main display
format.