Hanwei E6000 Manual de usuario

E6000 Portable Multi-Gas Detector
Operation Manual
(Ver: HW1403180296)
Read this manual carefully before using this device.

E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
1
SAFETY INFORMATION
Incorrect operation or unsuitable using environment may attenuate
the instrument’s performance. Therefore, please first read the
below safety information carefully before using and then operate
this device.
Please don't use the defective device. Before using, please
check if there is crack or spare part missing. If yes, please
contact the seller immediately.
It's suggested that the user carry out the "Bump Test" by
following Clause 5.3 of this manual before using the device. If
the device reading is beyond the specified range, please
calibrate the device by following Clause 5.7 and 5.8 of this
manual.
Periodic Bump test will test the response feature of the sensor.
Please make sure that the visual, audible and vibrative alarm
signals are ok.
Only accessories which are specified for E6000 or permitted by
the seller are allowed to be used.
Only the charger which is specified for E6000 is allowed to be
used. It's forbidden to charge the device in the dangerous
environment.
Please don't expose the device to the exceeding-range gas
environment for long time. Otherwise, it will badly influence the
performance and even damage the device.
If exposed to the environment consisting of leaded compound,
sulfocompound, organic phosphorus compound or silicon, the
gas sensor will be poisoned. Please don't use the device in the
above environment.
Please don't expose the device to the environment which
consists of H2S, hydrocarbons gas or high corrosive gas for

E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
2
long time. Otherwise, it will restrain the response of the gas
sensor and reduce the sensitivity. If the device has to be used
in the above environment, please follow Clause 5.3 to carry out
the Bump Test before using it.
Please don't expose the device to the environment which has
electric shock, strong magnetic field or serious continuous
mechanic shocking.
There is a lithium battery inside the device. Please don't place
the useless battery together with the rubbish. The useless
battery should be discarded by qualified withdrawers.
It's forbidden to disassembly, adjust or repair the device
without permission.
Please avoid the device falling from high place or serious
shocking.
Any other operation beyond this manual, please contact the
seller.
1. Brief introduction
E6000 is a compact and lightweight multi gas detector that
continuously measures combustibles, O2, CO, H2S and other toxic
gases in ambient air. By using 6 sensors, it can detect at most 7
gases at the same time. Its functional and watertight design (IP 66)
incorporates an Bump proof, rubberized housing to meet the
toughest requirements of harsh environments.
2. Main features
Advanced 16 digit MCU with low consumption
Ultra-wide angle LCM screen
Adjustable 2-level alarm points

E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
3
Adjustable calibration point
Self protection design for combustible gas sensor
Battery low voltage alert function
With real-time clock
Interchangeable smart sensor module design
Self adjustment design
Audible, visual and vibrative alarm signals
Data communications function
STEL/TWA alarm for toxic gases
Design of self-test, self-diagnose and self-repairing
Password management
Intrinsically safe design
3. Technical specification
Detecting method: Natural diffusion
Target gas: Refer to the Annex-1 in the end of this manual
Response time:
Semi-conduct, catalytic, thermal conduct sensor ----------T90<30s
O2, CO, H2S sensor---------------------------------------------- T90<30s
Other sensors -----------------------------------------------------T90<120s
Indication error range:
Combustible gas----------------------------------------------------- ±5% F.S.
Toxic gas------------------------------------------------------------------±5ppm
Working condition:
Temperature: -20℃~50℃Humidity: <95%RH
Power source: Lithium battery (DC3.6V, 6600mAh)
Working time per charging: ≤30 hours continuously (no alarm)
Charging time: ≤ 6 hours

E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
4
Explosion –proof grade: Exia IIC T4 Ga
Ingress protection: IP66
Dimensions and weight: 168mm×91mm×45mm about 500g
4. Structure and functions
Appearance
Display information

E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
5
5. Operation instructions
5.1 Power on
When machine is power off, hold for more than 3s and the
device will be power on. The screen shows as below.
After power on, the detector will proceed self-test and display is as
below drawing.
The device will make self test on the buzzer, alarm lights and
vibrator automatically. After self-test, the device enters into normal
detection status and the screen displays as below.

E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
6
Note:
If self-test fails, related information will be displayed. For
details, please refer to Clause 10 Troubleshooting of this
manual.
If self test succeeds, the device enters into warm-up period of
3-30s, which depends on the sensor type.
5.2 Power off
In normal detection status, hold for 3 seconds and the
screen will show “Shutting down…”. Meanwhile, the buzzer will give
out intermittent beep twice. Then machine is power off.
5.3 Bump test
Every day before using the device, the user is suggested carry out
Bump test, so as to check if the device is working normally.
Test method:
When the device is power on, put it into high level gas environment,
which is higher than the preset high alarm point of the device. If all
the device’s function is ok, then device can be used in the working
area.
Note:
If any reading on the screen is beyond the prescribed display
error range, please follow Clause 5.7 and 5.8 of this manual to
re-calibrate it.
If detector does not response or showing errors, please contact
the seller for repairing.

E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
7
5.4 Menu explanation
5.4.1 Common menu
In the normal detection interface, press both and
simultaneously and the device enters into common menu setting
interface. Below is the button function:
Menu
Submenu
Remarks
DATE
TIME
When changing the date or
time, the selected bit will blink.
ALARM
MODE
When changing the mode, the
selected bit will blink.
RECORD
GAP
Set time gap for record saving.
CAUTION
MESSAGE
Black box means this item is
selected.
BATTERY
MESSAGE
Display battery voltage and
remaining work time

E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
8
BACKLIGHT
TIME
Backlight time will be increased
by 5s each time.
After entering into certain submenu, the user can change the item
value by pressing and save the setting by pressing .
The user can also exit without saving changes by pressing .
5.4.2 Advanced menu
In the interface of common menu interface, press both and
simultaneously twice. The screen shows password inputting
interface. Input the password by pressing to increase the
bit value and press to confirm the input password. After
inputting the correct password, press to enter into advanced
menu interface. Select the submenu item by pressing and
enter into selected submenu by pressing .

E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
9
Menu
Submenu
Remarks
SAFETY
DEPLOY
To set user’s ID.
ZERO
CAL.
Zero calibration
DATA
UPLOAD
Data upload
PERIPHE
CONFIG.
One mode must be selected
POWER
MANAGE
Activate the other function
LANGUAGE
SELECT
One mode must be selected.
Note: Initial password for entering advanced menu is “0000”.
Tabla de contenidos
Manuales populares de Instrumento de medición de otras marcas

Endress+Hauser
Endress+Hauser Proline Promag 50 Especificaciones técnicas

Siemens
Siemens SITRANS F Coriolis FCT030 Manual de lista de piezas

KLINGER
KLINGER CMF V Series Manual de usuario

EXFO
EXFO FTB-2 Manual de operación y mantenimiento

Keysight
Keysight M8290A Manual de usuario

ADTEK
ADTEK MW-5 Manual de usuario