GF Signet 9050CR Manual de usuario

+GF+SIGNET 9050CR
Conductivity/Resistivity Controller
GEORGE FISCHER +GF+
Instruction Manual

CAUTION: Remove AC power to unit prior to
wiring input and output connections.
Important Safety Information!
CAUTION: Remove AC power before opening
unit. Electrical shock hazard exists

Chapter Page
1 Introduction 1
1.1 General Description 1
1.2 Front Panel Description 2
1.3 Rear Panel Description 3
2 Sensor Selection and Installation 4
2.1 Sensor Selection 4
2.2 In-Line Sensor Installations 6
2.3 Submersible Sensor Installations 8
3 Installation and Wiring 9
3.1 Mounting Instructions 9
3.2 Power Connections 10
3.3 Input Connections 11
3.4 Output Relay Connections 12
3.5 Verifying Analog Outputs 13
3.6 Analog Output Connections 14
4 System Configuration 15
4.1 Introduction 15
4.2 CAL Menu Functions 16
4.3 RELAY Menu Functions 18
4.4 OUTPUT Menu Functions 19
4.5
Quick-Reference Keypad Sequence
20
4.6 CAL Menu 22
4.7 Wet Calibration Procedure 24
4.8 Relay Operation 25
4.9 RELAY Menu 27
4.10 OUTPUT Menu 29
4.11 View Only Menus 30
5 Technical Support 32
5.1 Accessing Internal Options 32
5.2 AC Power Configurations 33
5.3 Security Code Function 33
5.4 Installing Input/Output Options 34
5.5 Output Card Configurations 34
5.6 Options and Accessories 35
Table of
Contents

5.7 Parts Per Million Conversion Factor 37
5.8 Temperature Coefficient 38
5.9 Troubleshooting 40
Specifications 43
Warranty Information 45

Unpacking and Inspection
Your controller package includes the following
items:
•+GF+SIGNET 9050CR Conductivity/
Resistivity Controller
• Two stainless steel mounting brackets
• Mounting Instructions w/self-adhesive template
• Panel gasket
• Instruction manual w/warranty card
Please fill out and return warranty card as soon as
possible.
Warranty Record
For your protection, record your unit's purchase
date and serial number for future reference. The
serial number decal is located on the instrument's
rear panel.
Model: +GF+SIGNET 9050CR Conductivity/
Resistivity Controller
Purchase Date: _______________________
Serial Number: ________________________
Purchased From: _______________________
Purchase Order Number: _______________


Chapter 1
Introduction
1.1 General Description
The +GF+SIGNET 9050CR Conductivity/
Resistivity Controller is specifically designed to
monitor and control conductivity levels of aqueous
solutions. The controller's compact 1/4 DIN
enclosure (front) is NEMA 4X/IP65 rated and
ideal for installations into instrumentation panels
with limited space.
Optional "plug-in" output cards allow you to
customize your controller to satisfy your application
demands. The controller's unique "slide-out"
chassis design makes option installation fast and
simple. Smart self-configuring microprocessor
based circuitry automatically inventories installed
options during power-up, allowing you to upgrade
your system in seconds without the need for
additional equipment.
The unit's front panel features a highly visible
4.5 and 8-digit liquid crystal display with
adjustable contrast. Measured conductivity/
resistivity and relay status is accessed at a glance.
Channel selection and solution temperature is
accessed with a touch of a button. During
calibration the user is prompted with clear
step-by-step instructions on the front panel display.
The 9050CR is designed for use with
+GF+SIGNET 3-28XX-1 series sensors, covering
a wide range of conductivity/resistivity
measurement. Each sensor is equipped with a
PT1000 temperature compensation device for
accurate temperature sensing (sec. 2.1).
1
The technical data given in
this publication is for general
information purposes only. It
implies no warranty of any
kind.

COND
RES
1.2 Front Panel Description
• Accesses one of three calibration menus: CAL, RELAY, OUTPUT
• Selects a menu item for modification
• Restores a menu item to previous value during modification
1. Relay An-
nunciators:
3.
5.
6.
7.
• Indicate activation status of control relays 1-4 (optional)
• Shows conductivity, resistivity, temperature °C, relay activation status, and
user messages during setup and operation
• Stores calibration and setup values into memory after modification
• Displays available output options during operation
4.
• Displays temperature in °C for each channel during operation
• Decreases the value of a selected digit in CAL, RELAY, or OUTPUT menus
8.
9. • Displays percent rejection between channels
• Increases the value of a selected digit in CAL, RELAY, or OUTPUT menus
• Used in conjunction with MOD key to access the RELAY calibration menu
• Accesses the RELAY "view-only" menu
• Used in conjunction with MOD key to access the OUTPUT calibration menu
• Accesses the OUTPUT "view-only" menu
• Displays conductvity/resistivity for each channel during operation
• Selects a digit for modification while in CAL, RELAY, or OUTPUT menus
• Returns the unit to operation mode from menus
10.
2. LCD
Display:
• Used in conjunction with MOD key to access the CAL calibration menu
• Accesses the CAL "view-only" menu
FunctionItem
MOD
ENTER
CAL
RELAY
OUTPUT
TEMP
% REJ
2
RLY 1
ENTER MOD
CAL RELAY OUTPUT TEMP
RLY 2
RLY 3
RLY 4
9050CR Controller
COND
RES % REJ
+GF+SIGNET
7
8
9
103
4
5
6
2
1

3
1.3 Rear Panel Description
• 90 to 132 VAC or 180 to 264 VAC system power
connection 1. AC IN
• 17 to 30 VDC system power connection 2. DC IN
• Analog output #1 from option socket #1 (optional) 5. ANL1
• Sensor input connections 6. CH1/CH2
• Serial outputs (future availability) 7. SERIAL
• Analog output #2 from option socket #2 (optional) 8. ANL2
• Relay #1 (COM, NO, NC) contact set for external device
control (optional) 3. RLY1
• Relay #2 (COM, NO, NC) contact set for external device
control (optional) 4. RLY2
Note:
Rear terminals
accept 18 to 22 AWG
wire
• Relay #4 (COM, NO) contact set for external device
control (optional) 9. RLY4
• Relay #3 (COM, NO) contact set for external device
control (optional) 10. RLY3
Item
Function
15 16 17 18 19 20 21 22 23 24 25 26 27 28
L
1
N
AC IN
+ GND NO C NC
RLY1DC IN NO C NC
RLY2
R+ R-
ANL1
1 2 3 4 5 6 7 8 9 10 11 12 13 14
C NO Tx GND Rx
CH2
S2 T2 S1 T1 SGCNOR+ R-
RLY3 RLY4 SERIALANL2 CH1
10 9 8 7 6
12 3 4 5

Chapter 2
Sensor Selection
and Installation
2.1 Sensor Selection
The 9050CR controller is compatable with +GF+
SIGNET 3-28XX-1 series sensors. +GF+Signet
offers five sensor versions satisfying a wide range
of measurement, listed below.
For optimum sensor performance and life, the
following requirements MUST be met:
• Verify your system's conductivity/resistivity
range.
• Verify the maximum pressure/temperature
specification for the sensor under consideration
is within your process range.
• Verify wetted sensor parts are chemically
compatible with process fluids (e.g. Compass
Corrosion Guide or equivalent).
Choose the +GF+Signet sensor best suited for
your application from the following:
4
Dimensions:
• 3-2819-1 Sensor
• 3-2820-1 Sensor
• 3-2821-1 Sensor
Order NPT Fitting
Number Cell Range Fitting Material
3-2819-1 0.01 0.055 - 100 µS 3/4 in. PP
(10 kΩ- 18 MΩ)
3-2820-1 0.1 1 - 1000 µS 3/4 in. PP
3-2821-1 1.0 10 - 10,000 µS 3/4 in. PP
Specifications
Temperature Compensation: PT1000
Wetted materials:
O-rings: EPR
Insulator: Teflon®
Electrodes: 316 stainless steel
Standard fitting: Polypropylene
Max. pressure: 100 psi/6.9 bar
Max. temperature: 212 °F/100 °C
Optional fitting: 316 SS (1/2 in. NPT)
#3-2820.392
Max. pressure: 200 psi/13.8 bar
Max. temperature: 248 °F/120 °C
152 mm
(6.0 in.)
107 mm
(4.2 in.)
4.6 m (15 ft.)
cable (standard)
Reversible
fitting assy
for submersion
mounting
3/4 in. NPT
(standard)
12.7 mm
(0.5 in.) 3-2821-1
sensor tip
3-2820-1
sensor tip