GERATHERM DUOTEMP GEF-100 Manual de usuario

1 - GB
INSTRUCTION FOR USE
Geratherm Duotemp GEF-100
DIGITAL EAR AND
FOREHEAD THERMOMETER

GB
2 - GB
CONTENTS
2 - GB
3 - GB
4 - GB
5 - GB
5 - GB
6 - GB
6 - GB
8 - GB
9 - GB
10 - GB
10 - GB
11 - GB
12 - GB
14 - GB
14 - GB
15 - GB
15 - GB
16 - GB
finger puppet
(exemplary
representation)
The parts of the unit
Specifications
Intended use
Precautions
Operation of the thermometer
What is body temperature
Temperature measurement - Ear
Temperature measurement - Forehead
Measuring water and room temperature
Memory
Changing the temperature scale
Battery replacement and insertion
Fault indication
Cleaning and Disinfection
Disposal
Limited Warranty
Symbol index
Certificate of guarantee
THE PARTS OF THE UNIT
Probe
Memory button
Probe attachment
(fore forehead use)
Probe
Battery cover
ON button
LCD panel

GB
3 - GB
SPECIFICATIONS
Model
GEF-100
Point of mea-
surement
Infrared measurement at the eardrum or
forehead
Measuring range
Ear mode:
Forehead mode:
0 °C to 100 °C (32 °C to 212 F°);
32.2 °C to 43.3°C (90 °F to 110 °F)
Operating
conditions
16 °C to 40 °C (60.8 °F to 104 °F),
humidity: 15 to 95 % R.H.
Storage condi-
tions
-25 °C to 55 °C (-13 °F to 131 °F),
humidity: 15 to 95 % R.H.
Resolution
0,1 °C or 0,1°F
Accuracy
(laboratory)
32.2 °C to 43.3 °C: +/- 0,2 °C
(90 °F to 110 °F: +/- 0,4 °F);
other ranges +/- 2%
Clinically
repeatability
max. +/- 0,3 °C
Memory
stores up to 10 measurements
IP Classification
IP20 (protected against foreign bodies
< 12,5 mm)
Automatic
switch-off
after 60 seconds of no use
Reaction time
abt. 1 second
Battery
DC 3V lithium battery CR2032
Battery life
approx. 1000 measurements
Dimensions
150 mm (l) x 36,7 mm (b) x 44 mm (h)
Weight
47 g (including battery)
Quality guarantee
DIN EN 12470-5 Medical Thermome-ters, Part 5: Standards for
Infrared Ear Thermometers. Geratherm
®
is certified in accordance
with Council Directive 93/42/EWG and EN ISO 13485 and
is entitled to affix the CE-mark (Notified Body: TÜV
Rheinland LGA Products GmbH).
Elektromagnetic compatibility (EMC)
The present medical product is a device with an extremely high
degree of noise immunity. Therefore, the enclosure with the EMC
tables was not printed out.
Specifications are subject to change

GB
4 - GB
• This thermometer may only be used for the pur-
poses described in this booklet.
• It is intended only for the ear and forehead meas-
ure-ments and not for rectal, oral or axillary body
temperature measurement.
• The thermometer is intended for private domestic
use only.
• Use of this instrument is not intended as a substi-
tute for consultation with your physician.
• A dirty sensor lens or a dirty/ curved auditory canal
may affect the measurements.
• The unit is not waterproof. Never immerse this in-
strument in any liquids.
• For cleaning and disinfection, please follow the
instructions in the section entitled “Cleaning and
Disinfection”.
• Ensure that children do not use the instrument un-
supervised.
• Do not use the instrument if you think it is damaged
or if you notice anything unusual.
• Without the manufacturer´s consent, no changes
may be made on this unit (e.g. opening of the unit
than for battery change purposes).
• Observe the storage and operating conditions de-
scribed in the «Specifications» section.
• Protect it from:
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
• If the instrument is not going to be used for a pro-
longed period, the batteries should be removed.
• Do not apply a protective cap!
• The unit has an off-set function to compensate for
the influence of the ambient temperature.
• When it is switched on, the unit conducts an auto-
matic self-testing procedure. Any faults are shown
on the display.
• The finger puppet is not suitable for children under
3 years.
PRECAUTIONS

GB
5 - GB
This thermometer is intended for the periodic
measurement and monitoring of human body
tempera-ture for private use.
INTENDED USE
WHAT IS BODY TEMPERATURE?
Body temperature refers to the internal temperature of
the body (core temperature). ‘Normal’ body tempera-
ture fluctuates over the course of 24 hours. It tends
to be lowest in the early morning and highest in the
late afternoon. Body temperature also varies depend-
ing on the part of the body where the temperature is
measured and also differs with age. Hence, the body
temperature measurement obtained depends on
when and where the temperature is taken and on who
the patient is.
Normal temperature
ranges - by point of
measurement
1)
Axillary
34.7 - 37.3 °C
Oral
35.5 - 37.5 °C
Ear
35.8 - 38.0 °C
Rectal
36.6 - 38.0 °C
Normal temperature
ranges - by age
2)
0 - 2
years
36.4 - 38.0 °C
3 - 10
years
36.1 - 37.8 °C
11 - 65
years
35.9 - 37.6 °C
> 65
years
36.8 - 37.5 °C
1) Chamberlain, J. M. Dr. med., and Temdrup, T. E. Dr. med., “New Light
on Ear Thermometer Readings”, Contemporary Pediatrics, March
1994.
2) Chamberlain, J. M. Dr. med., and Temdrup, T. E. Dr. med., “Determi-
nation of Normal Ear Temperature with an Infrared Emission Detec-
tion Thermometer”, Annals Emergency Medicine, January 1995

GB
6 - GB
1. Unpack
2. Remove protective foil from the display
3. Remove contact protecting strip of battery case
OPERATION OF THE THERMOMETER
Press the ON button [ON].
Insert the probe into the
auditory canal. Straighten
the acoustic meatus by
pulling the ear upwards and
backwards so that the mid-
dle ear is clearly visible.
Children under 1 year:
Make the ear canal straight
by gently pulling the ear
backward.
Children from 1 year and
adults: Make the ear canal
straight by gently pulling the
ear backward and upward.
TEMPERATURE MEASUREMENT –
EAR
In order to ensure accurate measurement, the
patient should first rest for 30 minutes in the
same room as the thermometer. Immediately after
exercise or a bath, the temperature may be raised.

GB
7 - GB
For the temperature to
be measured accurately
in the ear canal, the tip of
the probe must be directed
towards the eardrum.
Press the ON button and
wait until you hear a beep.
If no beep is heard, press
and release the ON button
again.
As soon as you hear the
beep, remove the probe
from the auditory canal.
Read the temperature
displayed on the LCD
panel.
36,8 °C / 98,24 °F
34,8 °C / 94,64 °F
• It is advisable to take the temperature in the same
ear, since the values measured in the left and right
ear may be different.
• In the case of patients whose auditory canal is
curved, the temperature values measured may de-
viate owing to anatomical influences.
• The ear should not be obstructed, for example by
excessive amounts of ear wax.
To switch from forehead mode back to ear mode,
press MEM for 2 seconds.
NOTE:

GB
8 - GB
TEMPERATURE MEASUREMENT –
FOREHEAD
Always check that the
probe attachment is fixed
onto the thermometer.
Prior to measurement, the
forehead should be free of
perspiration, make-up, etc.
Press the ON button [ON].
Press MEM for 2 seconds,
it will switch to Forehead
Mode.
Place the probe tip with the
attachment on the patient’s
forehead. The recomended
area is on the left-hand side
above the left eye-brow (or
on the opposite side if you
choose the right eyebrow).
Press the ON button (you
will hear a beep) and keep it
pressed down as you slide
the thermometer towards
the left, moving it for at least
three seconds over the
temple. Then remove the
thermometer from the point
of measurement.
In order to ensure accurate measurement, the
patient should first rest for 30 minutes in the
same room as the thermometer. Immediately after
exercise or a bath, the temperature may be raised.

GB
9 - GB
• The precise sensing system records the heat emit-
ted naturally through the skin. If the thermometer
is not moved over the temple, the temperature re-
corded may be lower.
• It is advisable to take the temperature 2 - 3 times
and to record the highest value.
• In some patients the forehead temperature may
differ as a result of environmental factors and it is
advisable to take and record your forehead tem-
perature when healthy.
• The measurement of forehead temperature is an
excellent point of reference for recognising raised
body temperature.
• If you are in any doubt about the temperature
measured, please measure it again in the ear and,
for the sake of accuracy, record the highest value.
• Measurement should be conducted in ear mode
only.
• To measure water temperature, keep the probe
at least 1cm (± 0.5 cm) above the surface of the
water. The greater the distance from the surface of
the water, the less accurate the measured values
will be.
Read the temperature
value displayed on the
LCD panel.
MEASURING WATER AND ROOM
TEMPERATURE
NOTE:
Temperatures measured in forehead mode are dis-
played as comparable rectal temperatures.

GB
10 - GB
MEMORY
Press the MEM button.
The first memory cell
number (“0”) will then show
the stored temperature (the
last reading).
Press the MEM button
again. Now the reading
before last will be shown.
The other values stored in
the memory can be viewed
by pressing the MEM button
again.
CHANGING THE TEMPERATURE
SCALE
In order to switch over to
Fahrenheit (°F), press the
MEM button and then press
and release the ON button
while the memory cell number
is displayed on the LCD
panel.
ON
MEM
Otros manuales para DUOTEMP GEF-100
3
Tabla de contenidos
Otros manuales de Termómetro de GERATHERM

GERATHERM
GERATHERM easy temp GT-2052 Manual de usuario

GERATHERM
GERATHERM rapid GT-195-1 Manual de usuario

GERATHERM
GERATHERM DUOTEMP GEF-100 Manual de usuario

GERATHERM
GERATHERM GT-101 Manual de usuario

GERATHERM
GERATHERM DUOTEMP GEF-100 Manual de usuario

GERATHERM
GERATHERM rapid GT-195-1 Manual de usuario

GERATHERM
GERATHERM duotemp Manual de usuario

GERATHERM
GERATHERM GT-101 Manual de usuario