PN 153-231-B www.formfactor.com ACP40-GSG-xxx Probes • 4
Quick Reference Guide
ACP40-GSG-xxx Probes
Probe Cleaning
Clean the probe tips and connectors occasionally, or when you suspect contact problems. Follow your microwave cable or
network analyzer manufacturer instructions for cleaning the precision microwave connector.
For more information, visit www.formfactor.com.
Symptom Possible Causes Solution
Probe is pulled to the side when
trying to land on the device
Cable applies strain on the probe Use positioner cable clamp to strain-relieve the probe
Intermittent electrical contact Dirty/contaminated probe tip Clean the probe tip
Dirty RF connectors Clean the RF connectors
Probe overtravel is not sufficient Adjust the overtravel
Positioner arm is not planarized after
mounting the probe
Planarize the positioner arm
Poor calibration accuracy SOLT calibration is used above 10GHz Use LR(R)M calibration
Cal kit is not defined correctly Verify your cal kit definitions
Poor calibration repeatability Poor connection in the system Check and retighten your RF connections
Defective cable Check the cable using VNA transmission calibration
Large temperature variations in the lab Switch to controlled temperature lab
Corporate Headquarters
7005 Southfront Road
Livermore, CA 94551
Phone: 925-290-4000
www.formfactor.com
© Copyright 2018 FormFactor, Inc. All rights reserved. No part of this document may be reproduced,
transmitted or displayed in any form or by any means except as duly authorized by FormFactor, Inc.
FormFactor and the FormFactor logo are trademarks of FormFactor, Inc. All other trademarks are the
property of their respective owners.
Important Notice
While the information contained herein is believed to be accurate as of the date hereof, no express or
implied representations or warranties are made with respect to its accuracy or completeness. FormFactor,
Inc., and its subsidiaries disclaim liability for any inaccuracies or omissions. All information is subject to
change without notice.
Users are required to read and follow carefully all safety, compliance and use instructions. Users assume all
loss and liability arising from the use of products in any manner not expressly authorized. The conditions and
methods of use of products and information referred to herein are the entire responsibility of the user and,
to the maximum extent permitted by applicable law, FormFactor, Inc., and its subsidiaries shall not be liable
for any damages, losses, costs or expenses arising out of, or related to, the use thereof.
No license, express or implied, by estoppel or otherwise, under any intellectual property right is granted in
connection herewith. Users shall take all actions required to avoid intellectual property infringement.