AccuMax Quantum Control Units
AccuMax Quantum Control UnitsAccuMax Quantum Control Units
AccuMax Quantum Control Units
Usage Instructions
Usage InstructionsUsage Instructions
Usage Instructions
3
Front Panel of Premium Control Unit
1. NORMAL PRESSURE LED—When green LED lights up, the pressure
inside of mattress has reached the desired position
2. MAX INFLATE PUSH BUTTON—Pressing the button suspends the
alternating function and fills every air
sector to the firmest level. Pressing the button again will resume normal
alternating function. The Max Inflate mode automatically times out to
insure patient safety.
3. MAIN POWER SWITCH—When pushed to “on,” the control unit will
begin filling the mattress as well as beginning the alternating function. The
lamp inside the switch will also illuminate. Pushing to “off” will turn off
power and the power indicator will cease to illuminate.
4. COMFORT CONTROL KNOB—Allows the caregiver to adjust the
firmness of the mattress to maximize patient comfort.
5. LOW PRESSURE LED—When illuminated, the air pressure inside the
mattress is below the normal range. When the mattress is first being
inflated, this LED will remain illuminated until mattress reaches normal inflation range. If this LED remains consistently lit, make sure that
connections to the mattress are correct and/or replace any damaged air tubes or hoses. If problem persists, please call an authorized Encompass
TSS service department.
CPR
When there is an emergency need to perform CPR on the patient, immediately detach the air hoses from either the pump end or the mattress end.
After disconnecting, please follow normal facility code procedures for administering CPR.
Air Filter Replacement (for Premium Control Unit only)
1. Loosen the screw holding the air filter in the control unit.
2. Take out the old air filter and replace with a new one.
3. We suggest replacing the air filter every twelve months of continuous use.
4. For replacement air filters, please contact Encompass TSS Customer Satisfaction Department 800-822-8288.